The Use of Wearables in Clinical Trials During Cancer Treatment: Systematic Review

The Use of Wearables in Clinical Trials During Cancer Treatment: Systematic Review

The Use of Wearables in Clinical Trials During Cancer Treatment: Systematic Review

Review

1Department of Oncology, Rigshospitalet, University of Copenhagen, Copenhagen, Denmark

2Faculty of Health, University of Copenhagen, Copenhagen, Denmark

Corresponding Author:

Cecilie Holländer-Mieritz, MD

Department of Oncology

Rigshospitalet

University of Copenhagen

Blegdamsvej 9

Copenhagen, 2100

Denmark

Phone: 45 35453545

Email: cecilie.hollaender-mieritz@regionh.dk


Background: Interest in the use of wearables in medical care is increasing. Wearables can be used to monitor different variables, such as vital signs and physical activity. A crucial point for using wearables in oncology is if patients already under the burden of severe disease and oncological treatment can accept and adhere to the device. At present, there are no specific recommendations for the use of wearables in oncology, and little research has examined the purpose of using wearables in oncology.

Objective: The purpose of this review is to explore the use of wearables in clinical trials during cancer treatment, with a special focus on adherence.

Methods: PubMed and EMBASE databases were searched prior and up to October 3, 2019, with no limitation in the date of publication. The search strategy was aimed at studies using wearables for monitoring adult patients with cancer during active antineoplastic treatment. Studies were screened independently by 2 reviewers by title and abstract, selected for inclusion and exclusion, and the full-text was assessed for eligibility. Data on study design, type of wearable used, primary outcome, adherence, and device outcome were extracted. Results were presented descriptively.

Results: Our systematic search identified 1269 studies, of which 25 studies met our inclusion criteria. The types of cancer represented in the studies were breast (7/25), gastrointestinal (4/25), lung (4/25), and gynecologic (1/25); 9 studies had multiple types of cancer. Oncologic treatment was primarily chemotherapy (17/25). The study-type distribution was pilot/feasibility study (12/25), observational study (10/25), and randomized controlled trial (3/25). The median sample size was 40 patients (range 7-180). All studies used a wearable with an accelerometer. Adherence varied across studies, from 60%-100% for patients wearing the wearable/evaluable sensor data and 45%-94% for evaluable days, but was differently measured and reported. Of the 25 studies, the most frequent duration for planned monitoring with a wearable was 8-30 days (13/25). Topics for wearable outcomes were physical activity (19/25), circadian rhythm (8/25), sleep (6/25), and skin temperature (1/25). Patient-reported outcomes (PRO) were used in 17 studies; of the 17 PRO studies, only 9 studies reported correlations between the wearable outcome and the PRO.

Conclusions: We found that definitions of outcome measures and adherence varied across studies, and limited consensus among studies existed on which variables to monitor during treatment. Less heterogeneity, better consensus in terms of the use of wearables, and established standards for the definitions of wearable outcomes and adherence would improve comparisons of outcomes from studies using wearables. Adherence, and the definition of such, seems crucial to conclude on data from wearable studies in oncology. Additionally, research using advanced wearable devices and active use of the data are encouraged to further explore the potential of wearables in oncology during treatment. Particularly, randomized clinical studies are warranted to create consensus on when and how to implement in oncological practice.

JMIR Mhealth Uhealth 2020;8(11):e22006

doi:10.2196/22006

Keywords



Technology expansion over the past decade, along with the use of various sensors and electronic devices and the arrival of more advanced devices, has led to new possibilities [1-3]. The potential use of sensor technology in the health care setting is wide and covers aspects suitable for many purposes [4,5]. In this review, we focus on the use of wearables in clinical trials during cancer treatment.

A wearable is a device with a sensor that can collect health-related data remotely [6]. Depending on the device design, it can be worn in different ways, either on the wrist, upper arm, around the waist, fastened to the hip, or on another location of the body. Wearables can collect information on various biometric data points (eg, heart rate, respiratory rate, blood oxygen saturation, sleep pattern, or body temperature) [7]. This information can be used alone or in combination with other information [eg, patient-reported outcome (PRO) or other patient-generated health data] to evaluate or estimate a clinical outcome; hospitalization, adverse events, performance status, and physical activity can be used as a clinical outcome. The information can be used for home monitoring with feedback to the clinician or for self-management by the patient. The collection of data from the wearable can be either offline or in real time, depending on the type of feedback wanted for the user (eg, the patient or clinician).

In oncology, wearables may offer new vital information about patients, which can potentially lead to better management of cancer treatment [7]. This is important both in the aspect of precision care and economy, with cancer being the second leading cause of death globally [8]. Wearables make it possible to monitor patients at home in their own environment, compared to monitoring only at in-clinic visits [9]. Tracking data in the patient’s own environment might allow the patient to continue their normal routines while their health data is being transmitted to a database or directly to the clinician. The possibilities are appealing in oncology from a “work smarter” approach and, more importantly, for the general improvements in the quality of life they may offer to patients with cancer of [10,11]. At present, there is little clinical evidence showing how wearables might improve the cancer pathway for patients with cancer during treatment [12].

To understand the potential use of wearables, it is relevant to capture what the wearable’s objective outcome is, what effect or role it has on the clinical outcome, and what it can be used for in a medical setting [7]. Evaluating the patient’s adherence to the wearable and defining valid data helps to ensure that new technologies are introduced into clinical practice with a focus on the patient’s perspective. Limited consensus and guidelines exist for designing or reporting trials using wearables as part of the intervention—but this research area is getting increased attention [13,14]. One initiative is the Clinical Transformation Initiative (CTTI), which has issued recommendations regarding the appropriate use of mobile technology in clinical trials [13]. They have also initiated a database on feasibility studies in clinical trials [15]. The database is not limited to one specific disease group but only includes feasibility studies [15].

The purpose of this review is to explore the use of wearables in clinical trials during cancer treatment, with a focus on adherence and the setting.


Study Design and Search Strategy

Systematic searches were performed in PubMed and EMBASE. Both databases were searched prior and up to October 3, 2019, with no limitation in the date of publication. Searches consisted of cancer/neoplasm keywords and terms for wearable devices.

In PubMed, the search consisted of the medical-subject-heading (MeSH) terms “neoplasms,” “medical oncology,” “surgical oncology,” and “wearable electronic devices,” along with a combination of free-text words for the topics “oncology,” “cancer,” “wearable device,” “accelerometer,” and “actigraph.” In EMBASE, the search included categorized terms for neoplasms and electronic monitoring devices such as “neoplasm,” “patient monitoring,” and “electronic device.” Additional search terms “ambulatory monitoring” and “telemedicine” were added. The search was limited to articles published in English. The search strategy is shown in Multimedia Appendix 1. The review was registered at PROSPERO (International prospective register of systematic reviews) ID number CRD42020154386 before data extraction was initiated [16].

Criteria for Inclusion of Studies

Studies found with the selected search strategy were screened by title and abstract, which was performed independently by 2 reviewers who were blinded to each other’s decisions. Cases of disagreement about whether to include or exclude a study were decided through a consensus decision. The included studies had their full text assessed for eligibility; disagreements were resolved by consensus, which was achieved in all cases.

For a study to be included, it had to be written in English and be either a randomized controlled trial (RCT), observational study, or pilot study/ feasibility study. Patients had to be 18 years of age or older and diagnosed with a solid malignant tumor. It was mandatory that studies took place during active cancer treatment; treatment could be either radiation or antineoplastic treatment, such as chemotherapy or targeted therapy. Studies investigating all types of wearables were considered eligible if they had an objective measure. Studies had to include a description of adherence to the wearable to be eligible for inclusion.

Exclusion criteria were studies registered as protocol descriptions, study protocols, abstracts from conferences, editorials, letters, or case reports. Also excluded were studies in which patients were cancer survivors or had hematologic malignancies and were treated with endocrine therapy only or surgery only. If the wearable devices were worn only pretreatment, the study used hearing aids as the wearable device, or wearables were used as treatment or for diagnostics, then these studies were also excluded.

The search is graphically presented according to the PRISMA flow diagram (Figure 1).

Figure 1. PRISMA flow diagram of the screening and selection of studies.
View this figure

Data Extraction

The following study characteristics data were extracted: study title, author, year of publication, country, study design, number of patients included, and main objectives. Study population information that was extracted included age group, cancer type, treatment type, and intent of treatment. Study data regarding the adherence to the wearable were the type of wearable used (hardware, software), placement, device outcome, planned wear time, valid wear time, and adherence to the wearable. For the purposes of this review, the device outcome was defined as the objective measures used in the study (eg, step count); planned wear time was defined as the time period that patients were supposed to use the device; and valid wear time was defined as the minimum wear time for data inclusion, extracted if available (eg, ≥10 hours/day). Adherence could either be the percentage of patients wearing the wearable for the period, the percentages of patients with evaluable sensor data, or the percentages of total evaluable days. Study outcomes were extracted and thematically grouped into wearable outcome, PRO, and clinical outcome. Wearable outcome was defined as circadian rhythm, physical activity (PA), skin temperature, and sleep. PRO topics were quality of life, PA, mental health, symptom registration, and others. Clinical outcomes included adverse events, performance status (PS), and hospitalization.

When reading through the full text of included studies, synonyms for “wearable” were registered and extracted from each study.

Outcomes and Analysis

The review was conducted to give a descriptive presentation of the use of wearables in clinical trials. The primary outcome was adherence to the wearable. The secondary outcomes were the study outcomes: the wearable outcome, the PRO, and the clinical outcome. The wearable outcome was subtracted from the device outcome by the authors, and thematically grouped as a wearable outcome. We also investigated whether the studies reported a relationship between the wearable outcome and the PRO. All data were presented descriptively.

Ethical Considerations

This review did not require national or institutional approval.


The search strategies were performed in PubMed and EMBASE on October 3, 2019. A total of 1281 studies were identified through the searches. There were 12 duplicated records, which were excluded, leaving 1269 studies eligible for screening. Titles and abstracts were examined, which resulted in the exclusion of 1180 studies that did not meet our inclusion criteria. The remaining 89 studies were evaluated for eligibility. Through full-text access, 64 other irrelevant studies were excluded, leaving a total of 25 studies to be reviewed for the purpose of this review. This process of study screening and selection is illustrated in a PRISMA flow chart (Figure 1).

Study characteristics are presented in Table 1, which shows a wide heterogeneity with respect to primary cancer sites (range, 1-8 sites), treatment-type specifics (eg, whole-brain radiotherapy, to all treatments allowed), sample sizes (n=7-180), and age groups (22-94 years). In terms of cancer diagnosis, 13 of the 25 studies included patients with breast cancer, 7 having breast cancer exclusively. Gastrointestinal (GI) cancer was featured in 12 of the 25 studies, 4 of which studied patients with GI exclusively. Lung cancer was present among 10 of the 25 studies, 4 featuring this diagnosis exclusively. For 19 of the 25 studies, the patients were treated with curative intent. For most (17/25) of the included studies, treatment was chemotherapy. Of the 25 studies included, a total of 12 were pilot/feasibility studies, 10 were observational studies, and 3 were randomized controlled trials (Table 1).

In 5 of the 25 studies, the planned wear time was ≤7 days, 13 were between 8-30 days, 5 were between 31-90 days, and in 2 studies, the planned wear time was over 90 days (Table 2).

Table 1. Characteristics of included studies, n=25.
Study (year)CountryPrimary cancer siteTreatment typeSample size, nAge group (range or mean [SD])Study type
Broderick et al (2019) [17]United StatesMixedChemotherapy4224-72Pilot / feasibility study
Champ et al (2018) [18]United StatesBreastRadiotherapy1052-79Pilot / feasibility study
Chevalier et al (2003) [19]FranceGastrointestinalChemotherapy1043-73Pilot / feasibility study
Dean et al (2013) [20]United StatesLungChemotherapy3548-94Observational study
Dreher et al (2019) [21]United StatesBreastChemotherapy6529-72Observational study
Edbrooke et al (2019) [22]AustraliaLungMixed9263 (12.3)Randomized controlled trial
Gupta et al (2018) [23]United StatesMixedSystemic therapy2454 (12.5)Pilot / feasibility study
Innominato et al (2016) [24]United KingdomMixedChemotherapy3135-91Pilot / feasibility study
Li et al (2019) [25]ChinaBreastAdjuvant chemotherapy18022-74Observational study
Low et al (2017) [11]United StatesGastrointestinalChemotherapy1440-74Pilot / feasibility study
Lowe et al (2014) [26]CanadaMixedRadiotherapy (whole brain)3163.5 (10.4)Observational study
Mouri et al (2018) [27]JapanMixedChemotherapy3070-84Pilot / feasibility study
Nyrop et al (2018) [28]United StatesBreastChemotherapy10024-64Observational study
Ohri et al (2019) [29]United StatesLungChemo-radiotherapy5038-90Observational study
Ohri et al (2017) [30]United StatesMixedChemo-radiotherapy3833-82Pilot / feasibility study
Ortiz-Tudela et al (2014) [31]FranceMixedChemotherapy4935-90Observational study
Parker et al (2019) [32]United StatesPancreasChemotherapy; chemo-radiotherapy5066 (8)Observational study
Roche et al (2014) [33]FranceGastrointestinalChemotherapy1651-89Pilot / feasibility study
Roscoe et al (2002) [34]United StatesBreastChemotherapy +/- radiotherapy10234-79Randomized controlled trial
Sarna et al (2001) [35]United StatesMixedRadiotherapy748-74Pilot / feasibility study
Savard et al (2009) [36]United StatesBreastChemotherapy9534-79Observational study
Solk et al (2019) [37]United StatesBreastChemotherapy6731-71Observational study
van der Meij et al (2012) [38]The NetherlandsLungChemo-radiotherapy4039-80Randomized controlled trial
Vassbakk-Brovold et al (2016) [39]NorwayMixedChemotherapy6659 (11)Pilot / feasibility study
Wright et al (2018) [40]United StatesGynaecologicalChemotherapy1060 (11)Pilot / feasibility study
Table 2. Description of wearables and adherence.
Planned wear time interval and study (year)Hardware / softwareDevice outcomePlanned wear time / valid wear timeAdherence description
7 days




Chevalier et al (2003) [19]Actigraph, Ambulatory Monitoring Inc / Action 3.8
  • Rest activity cycle (movements/period)
3 days / 3 days
  • 100% (10/10) of the patients wore the device for the full period

Dean et al (2013) [20]Motionlogger actigraph / Action 3
  • Sleep efficiency (%)
  • Sleep (hours)
  • Wake after sleep onset (minutes)
7 days / not reported
  • 86% (30/35) of the patients wore the device for the full period

Lowe et al (2014) [26]activPALa / not reported
  • Position time (hours/day)
  • Energy expenditure (metabolic equivalent of task [MET] h/day)
  • Step count (steps/day)
7 days / not reported
  • 77% (24/31) of the patients provided evaluable sensor data between 3 and 7 days

Roscoe et al (2002) [34]Mini-Motionlogger Actigraph / Action 3
  • Circadian consistency (I<Ob)
  • Daytime activity level (minutes)
  • Sleep (%)
72 hours at 2 timepoints / not reported
  • 89% (91/102) provided evaluable sensor data at second cycle of chemotherapy
  • 44% (45/102) provided evaluable sensor data at fourth cycle of chemotherapy

Vassbakk-Brovold et al 2016) [39]SenseWear Armband Pro3 or SenseWear Armband Minia / SenseWear version 6.1 for Pro3 and version 7.0 for Mini
  • Physical activity (minutes/week) recorded in 1-minute epochs
5 days / ≥19.2 hrs, for ≥1 day
  • 79 % (66/84) of the patients wore the device for the full period





8-30 days




Edbrooke et al (2019) [22]SenseWear accelerometera / not reported
  • Step count (steps/day)
  • Number of 10+ minutes step bouts/day
  • Duration of 10+ minutes bouts (minutes)
  • Cadence of 10+ minutes bouts (steps/min)
7 days at 3 timepoints / 8hrs/day, for ≥4 days
  • 87% (80/92) of the patients provided evaluable sensor data at baseline
  • 71%(65/92) of the patients provided evaluable sensor data at 9 weeks
  • 60% (55/92) of the patients provided evaluable data at 6 months

Innominato et al (2016) [24]Micro Motionlogger / Action 4
  • Circadian rest-activity (I<Ob)
30 days / not reported
  • Evaluable sensor data were available in 75 % of the total days (653/874)

Li et al (2019) [25]GENEActiv Original / not reported
  • Sleep efficiency (%)
  • Sleep duration (minutes)
  • Nighttime total wake time (minutes)
7 days at 3 timepoints / ≥5 days per timepoint
  • 97% (175/180) of the patients provided evaluable sensor data at T2
  • 76% (136/180) of the patients provided evaluable sensor data at T3

Low et al (2017) [11]Fitbit Charge HR / not reported
  • Step count (steps/day)
  • Floors climbed (n)
  • Sleep (minutes)
  • Awakenings (n)
  • Time in bed (minutes)
4 weeks / not reported
  • Evaluable sensor data were available in 75 % of the total days (295/392 days)

Mouri et al (2018) [27]Kenz Lifecorder‐GSa / Lifelyzer‐05 coach
  • Step count (steps/day)
  • Physical activity (minutes/day) (physical activity was rated ≥1.8 METs)
7 days at 3 timepoints / ≥5 hrs/day
  • 93% (28/30) of the patients wore the device for the full period

Ohri et al (2019) [29]Garmin Vivofit a / not reported
  • Step count (steps/day)
Up to 3 weeks / not reported
  • Evaluable sensor data were available in 94 % of the total days (741/791)

Ortiz-Tudela et al (2014) [31]Mini-Motionlogger Actigraph / Action 4
  • Rest-activity (I<Ob)
  • Wrist accelerations (acc/minute)
10-14 days split into 4 periods of 3-4 days / not reported
  • 86% (42/49) of the patients provided evaluable sensor data the full period

Roche et al (2014) [33]Mini-Motionlogger and VitalSense / Action 4, version 1.10
  • Rest-activity (I<Ob)
  • Wrist accelerations (acc/minute)
  • Skin surface temperature (°C/minute)
12 days split into 3 periods of 4 days/ not reported
  • 100% (16/16) of patients provided evaluable sensor data at baseline
  • 63% (10/16) of patients provided evaluable sensor data during therapy and after therapy administration

Sarna et al (2001) [35]Actiwatch 2 / not reported
  • Wrist movement (n/second)
  • Physical activity (15-minute intervals)
5 days at 2 timepoints/ ≥3 days per timepoint
  • 100% (7/7) of the patients wore the device the full period

Savard et al (2009) [36]Actillume / Action 3
  • Circadian rhythm variables (calculated from orientation and movement)
72 hrs at 7 timepoints/ not reported
  • 91% (86/95) of patients provided evaluable sensor data at baseline (first cycle of chemotherapy week 1: 80%, week 2: 73% and week 3: 76%; fourth cycle of chemotherapy week 1: 74%, week 2 63% and week 3: 68%)

Solk et al (2019) [37]ActiGraph, model wGT3X-BT / ActiLife, version 6.13.3
  • Activity data (1-minutes intervals)
10 days at 3 timepoints / ≥10 hrs/day
  • 84% (63/75) of the patients provided evaluable sensor data for the full period

van der Meij et al (2012) [38]PAM accelerometer, model AM101a / not reported
  • Physical activity (index score, 3 points reflects about 10 min of walking)
7 days at 3 timepoints / ≥3 full days
  • 65% (26/40) of the patients wore the device for the full period

Wright et al (2018) [40]Fitbit Zip and Fitbit Charge 2 / Fitabase
  • Step count (steps/day)
  • Heart rate
30 days / ≥4 days/week
  • 90% (9/10) of the patients wore the devices for the full period

31-90 days




Broderick et al (2019) [17]Microsoft Band 2 / not reported
  • Step count (steps/day)
  • Heart rate
  • Calories (calories/hour)
60 days / ≥6 hrs/day
  • Evaluable sensor data were available 86 % of the days (only day 1-14 included)

Champ et al (2018) [18]Misfit Shinea / not reported
  • Step count (steps/day)
  • Calories (calories/day)
  • Walking distance (miles)
  • Sleep (hours)
10 weeks / not reported
  • 90% (9/10) of the patients wore the device for the full period

Gupta A et al (2018) [23]Fitbit Flex / not reported
  • Step count (steps/day)
  • Physical activity (sedentary minutes/day)
  • Sleep (minutes)
12 weeks / ≥1 steps/day recorded
  • 96% (23/24) wore the device for >50% of the period

Nyrop et al (2018) [28]Fitbit Zipa / not reported
  • Step count (steps/day)
6-12 weeks / ≥3 weeks
  • 79% (100/127) of the patients provided evaluable sensor data

Ohri et al (2017) [30]Garmin / not reported
  • Step count (steps/day)
Up to 80 days / 80% of the days
  • Evaluable sensor data were available 94 % of the days
>91 days




Dreher et al (2019) [21]Fitbit Charge HR or Fitbit Charge 2 / Fitabase
  • Step count (steps/day)
  • Heart rate
  • Sleep data
Up to 270 days / ≥10 hrs/day
  • Evaluable sensor data were available in 45% of the days across 9 months

Parker et al (2019) [32]ActiGraph GT3X+a / ActiLife Software, Version 6
  • Physical activity (minutes/week) (1-min epochs)
14 days at each therapy phase / ≥10 hrs/day, for ≥7 days per timepoint
  • 88 % (44/50) of the patients provided evaluable sensor data

aPlacement other than wrist (anterior mid-thigh, hip, triceps muscle waist, not reported).

bbI<O is computed as the percentage of activity epochs when in-bed (I), whose values are lower than the median level of activity when out-of-bed(O).

Adherence data, presenting how many patients were able to use or collect data from the wearable device, or how many evaluable days the wearable was worn, were collected. Adherence varied across studies, from 60%-100% and 45%-94%, respectively, but was differently measured and reported. Valid wear time was defined in 16 of the 25 studies. Different hardware and software were used. The most frequent placement of the wearable was the wrist.

In Table 3, study outcomes were grouped as wearable outcome, PRO, and clinical outcome, and their respective subtopics can be seen for each included study, showing that 1 study could have more than 1 topic assigned. Of the 25 studies included, 19 had the purpose of monitoring PA (wearable outcome) as an outcome. The second most frequent topic among wearable outcomes was circadian rhythm, monitored in 8 studies. With respect to PRO, 9 studies examined quality of life, 7 studied mental health, 7 studied physical activity, and 8 looked at symptoms. Clinical outcomes were few, 4 comprising performance status, 2 looking at adverse events, and 2 studying hospitalization (Table 3).

Table 3. Study outcomes
Cancer type and study (year)Wearable outcomesPatient-reported outcomesClinical outcomes


Circ. rhythmaPhys. activitybSkin temp.cSleepMental healthPhys. activityQoLdSymptomsOtherAdverse eventsPerf. statuseHospitalizationOther
Breast



Roscoe et al (2002) [34]






Savard J et al (2009) [36]










Champ et al (2017) [18]










Li et al (2019) [25]









Nyrop et al (2018) [28]



Dreher et al (2019) [21]










Solk et al (2019) [37]







Gastrointestinal

Chevalier et al (2003) [19]











Roche et al (2014) [33]










Low et al (2017) [11]









Parker et al (2019) [32]









Gynecological

Wright et al (2018) [40]





















Lung

van der Meij et al (2012) [38]










Dean et al (2013) [20]








Edbrooke et al (2019) [22]







Ohri et al (2019) [18]









Mixed

Sarna et al (2001) [35]









Ortiz-Tudela et al (2014) [31]











Lowe et al (2014) [26]









Innominato et al (2016) [24]











Vassbakk-Brovold et al (2016) [39]










Ohri et al (2017) [30]









Gupta et al (2018) [23]







Mouri et al (2018) [27]











Broderick et al (2019) [17]








aCirc. rhythm: Circadian rhythm.

bPhys. activity: Physical activity.

cSkin temp.: Skin temperature.

dQoL: Quality of life.

ePerf. status: Performance status.

Of the 17 PRO studies, only 9 studies reported correlations between the wearable outcome and the PRO (Table 4). It was primarily physical activity, which was compared with the PROs (7/9).

Synonyms for “wearable” were also collected for each study (data not shown) while reading through the full text, which reflected both the terms used to address the technology in general and the terms describing the actual device used in the study. The most commonly used term was “accelerometer,” which was used in 12 studies; next was “actigraph,” which was mentioned in 8 studies. The term “tracker” was used in 6 studies as a part of several terms, including “activity tracker,” “wearable activity tracker,” and “fitness tracker.” The latter term was used similarly to “monitor,” which was mentioned in 9 studies.

Table 4. Studies that reported relationships between wearable outcomes and patient-reported outcomes (PRO; n=9).
PROWearable outcome

Circadian rhythmPhysical activitySkin temperatureSleep
Mental health[34][28],[22], [23]a[23]
Physical activity[28], [22], [26], [39], [17]
Quality of life[28], [22], [26], [23][23]
Symptoms[28], [11], [22], [26][11]
Others[34] (fatigue)[28] (fatigue), [23] (fatigue), [17] (fatigue, sleep)[40] (sleep), [23] (fatigue)

aNo relationship reported.


Summary of Findings

The use of wearable sensor devices has become a popular self-awareness gadget for many people today, especially when it comes to measuring physical activity [41]. In this review, we demonstrate the heterogeneous use of wearables during cancer treatment reported in research studies. In a search of the literature, 1269 studies were identified, of which 25 were included in our review. These studies represented different cancer types, with most focusing on mixed cancer types or breast cancer solely. Treatment given in the studies was primarily chemotherapy. Study types were pilot/feasibility, observational, and randomized controlled studies with sample sizes varying from 7 to 180 patients. All studies included in the review used a wearable with an accelerometer, but monitoring duration varied (3-270 days); and even though most studies (19/25) had physical activity as the wearable outcome, the device outcome for physical activity varied. With respect to adherence to using the wearable, a considerable variation was seen. Of the 17 PRO studies, 9 studies made comparisons between wearable outcome and PRO. In general, we noticed a broad variation in study designs, definitions, and outcomes within this field. This was also reflected in the number of synonyms for “wearable” used as terms to address the technology in general and as terms to describe the actual device used in a study.

Our Focus

Other studies have reviewed the use and effects of eHealth tools, such as those for patient self-reporting of medication management and use, and have concluded that more high-quality research is needed before standard implementation of such tools can occur [42]. In oncology, there is a growing urge to include patients in the management of their own illness, making wearable devices a valuable tool in cancer therapy. However, technical and clinical feasibility are essential aspects to explore [15], as the device measurements depend on patients using them. In this review, we have additionally focused on a somewhat understudied issue—adherence. We report adherence in relation to wear time and report how many patients were able to use or collect data with the wearable device, or how many evaluable days the wearable was worn. Adherence appears to have a wide range, and no data are available on missing wear time, which leaves us questioning if patients only wear wearables when they feel fit or when it is convenient. Thus, before designing large intervention studies using wearables, one must consider defining minimum wear time or conditions with mandatory wear time, since changes in these parameters might influence results [13,14,43].

Strengths and Limitations

To our knowledge, this is the first review of the use of wearables in clinical trials during cancer treatment. This review was limited to studies that included adherence as an issue, but was not restricted to specific types of wearables. However, we and others believe that the choice of wearable outcome is highly important. Determination of which variables to measure is crucial to ensure the purpose of the wearable when incorporated into patients’ daily routines [1]. Outcomes must be of specific value regarding the individual patient’s course of disease. We agree with other researchers that it is unnecessary to monitor many different kinds of variables if the outcome is not going to have a significant impact on the patient’s treatment [41]. At present, most studies tend to choose PA as the wearable outcome, but this may be because knowledge from the field of fitness and fitness training has grown [44]. Other wearable outcomes might be more relevant in the cancer setting, but this remains to be studied. Further, the objective measure of PA across studies varies. Of 9 studies comparing the wearable outcome and PRO, only 2 were randomized controlled trials. Further investigations in this area is needed before conclusions can be drawn. We suggest that future research include measurements of relevant, well-defined outcomes and be based on guidelines within the field, where such exist [45,46].

Many different terms are being used to describe electronic devices for use in health care [47], which can make it difficult to get an overview of the field and to keep up with how far research has come. In general, the terms used to describe the technology differ, from being very specific (eg, “wearable activity tracker” or “physical activity monitor”) to being broad, and covering several types of devices (eg, “telehealth,” “mobile health,” “wearable devices,” and “remote patient monitoring”) [7,48]. Consensus regarding the terms used could be helpful for both indexing studies and categorizing results by the specific types of devices used; such uniformity would improve research in the field and probably lead to more and improved knowledge.

The patient populations represented in this review mostly reflect breast and mixed cancer populations. Additionally, most studies are from the United States. It is questionable if results from such studies can be transferred to other diagnoses, countries, and cultural settings.

Implications for Future Research

As in this review, overall adherence to wear time or to wearable device interventions in general are difficult to compare. This is because almost every study has a different way of defining how many minutes or hours of wear time should count as a valid active day. By establishing standards for definitions of wear time, this could allow results across different patient populations to be compared more easily. This could also be solved by using a parameter other than step count as a measure of physical activity [41].

This review provides an overview of the frequently used wearables in oncology during therapy. Many of the wearables used have similar competencies, which might suggest the need to expand research into using more advanced wearables like smartwatches; according to Lu et al [2], very few clinical trials using smartwatches could be identified in 2016. Besides providing extended opportunities to measure variables, smartwatches also benefit from having a user-friendly display, which can make it feasible for patients to track their own activity status. Only 2 of the 25 reviewed studies used real-time feedback [24,40], and this feature might play an important role in motivating patients and possibly detecting worsening or severe symptoms earlier [11]. Technology allows plenty of different wearables to be applied in the oncology setting. However, it remains unknown if wearables can improve essential outcomes like overall survival or lead to other improvements in cancer treatment. Only future, well-designed research studies based on guidelines for this field with clinically relevant outcomes can help us decide when and where to apply these important tools.

This review provides an inventory for the status of wearables in clinical trials and can be used in addition to the CTTI studies database when designing new clinical trials with wearables [15].

Conclusion

This review provides an overview of the use of wearable devices in oncology care for patients with solid tumors receiving antineoplastic treatment. We extracted data from studies monitoring patients with cancer and presented these results specifically regarding adherence, the device outcomes, and the types of wearables used.

We found that definitions of outcome measures and adherence varied across studies, and limited consensus among studies existed on which variables to monitor during treatment.

Less heterogeneity and better consensus in terms of use and establishing standards for definitions of wearable outcomes and adherence would improve the comparisons of outcomes among studies using wearables. Adherence and consistent definitions are crucial for drawing conclusions from data from wearable studies in oncology. Additionally, research using advanced wearable devices and active use of the data are encouraged to further explore the potential of wearables in oncology during treatment. Especially, randomized clinical studies are warranted to create consensus on when and how to implement in oncological practice.

Acknowledgments

CHM is supported by the Faculty of Health and Medical Sciences, University of Copenhagen, Denmark.

Conflicts of Interest

None declared.

Multimedia Appendix 1

Search strategy.

PDF File (Adobe PDF File), 62 KB

  1. Chung AE, Jensen RE, Basch EM. Leveraging Emerging Technologies and the. J Oncol Pract 2016 Oct;12(10):863-866 [FREE Full text] [CrossRef] [Medline]
  2. Lu T, Fu C, Ma MH, Fang C, Turner AM. Healthcare Applications of Smart Watches. A Systematic Review. Appl Clin Inform 2016 Sep 14;7(3):850-869. [CrossRef] [Medline]
  3. Pennell NA, Dicker AP, Tran C, Jim HSL, Schwartz DL, Stepanski EJ. mHealth: Mobile Technologies to Virtually Bring the Patient Into an Oncology Practice. Am Soc Clin Oncol Educ Book 2017;37:144-154. [CrossRef] [Medline]
  4. Piwek L, Ellis DA, Andrews S, Joinson A. The Rise of Consumer Health Wearables: Promises and Barriers. PLoS Med 2016 Feb;13(2):e1001953 [FREE Full text] [CrossRef] [Medline]
  5. Jim HSL, Hoogland AI, Brownstein NC, Barata A, Dicker AP, Knoop H, et al. Innovations in research and clinical care using patient-generated health data. CA Cancer J Clin 2020 May;70(3):182-199 [FREE Full text] [CrossRef] [Medline]
  6. Ravizza A, De Maria C, Di Pietro L, Sternini F, Audenino AL, Bignardi C. Comprehensive Review on Current and Future Regulatory Requirements on Wearable Sensors in Preclinical and Clinical Testing. Front Bioeng Biotechnol 2019;7:313 [FREE Full text] [CrossRef] [Medline]
  7. Dias D, Paulo Silva Cunha J. Wearable Health Devices—Vital Sign Monitoring, Systems and Technologies. Sensors 2018 Jul 25;18(8):2414. [CrossRef]
  8. Cancer today 2018. World Health Organization: WHO. 2019 Dec 01.   URL: https://gco.iarc.fr/today/data/factsheets/populations/900-world-fact-sheets.pdf [accessed 2019-12-01]
  9. Beg MS, Gupta A, Stewart T, Rethorst CD. Promise of Wearable Physical Activity Monitors in Oncology Practice. JOP 2017 Feb;13(2):82-89. [CrossRef]
  10. Reilly CM, Bruner DW, Mitchell SA, Minasian LM, Basch E, Dueck AC, et al. A literature synthesis of symptom prevalence and severity in persons receiving active cancer treatment. Support Care Cancer 2013 Jun;21(6):1525-1550 [FREE Full text] [CrossRef] [Medline]
  11. Low CA, Dey AK, Ferreira D, Kamarck T, Sun W, Bae S, et al. Estimation of Symptom Severity During Chemotherapy From Passively Sensed Data: Exploratory Study. J Med Internet Res 2017 Dec 19;19(12):e420 [FREE Full text] [CrossRef] [Medline]
  12. Coughlin SS, Caplan LS, Stone R. Use of consumer wearable devices to promote physical activity among breast, prostate, and colorectal cancer survivors: a review of health intervention studies. J Cancer Surviv 2020 Jun;14(3):386-392. [CrossRef] [Medline]
  13. Developing Novel Endpoints Generated by Mobile Technologies for Use in Clinical Trials 2017. Clinical Trials Transformation Initiative. 2020 Jun 08.   URL: https://www.ctti-clinicaltrials.org/files/novelendpoints-recs.pdf [accessed 2020-06-08]
  14. Byrom B, Watson C, Doll H, Coons SJ, Eremenco S, Ballinger R, ePRO Consortium. Selection of and Evidentiary Considerations for Wearable Devices and Their Measurements for Use in Regulatory Decision Making: Recommendations from the ePRO Consortium. Value Health 2018 Jun;21(6):631-639 [FREE Full text] [CrossRef] [Medline]
  15. Bakker JP, Goldsack JC, Clarke M, Coravos A, Geoghegan C, Godfrey A, et al. A systematic review of feasibility studies promoting the use of mobile technologies in clinical research. NPJ Digit Med 2019;2:47 [FREE Full text] [CrossRef] [Medline]
  16. 2020. International prospective register of systematic reviews, PROSPERO. 2020 May 20.   URL: https://www.crd.york.ac.uk/prospero/display_record.php?RecordID=154386 [accessed 2020-05-20]
  17. Broderick JE, May M, Schwartz JE, Li M, Mejia A, Nocera L, et al. Patient reported outcomes can improve performance status assessment: a pilot study. J Patient Rep Outcomes 2019 Jul 16;3(1):41 [FREE Full text] [CrossRef] [Medline]
  18. Champ CE, Ohri N, Klement RJ, Cantor M, Beriwal S, Glaser SM, et al. Assessing Changes in the Activity Levels of Breast Cancer Patients During Radiation Therapy. Clinical Breast Cancer 2018 Feb;18(1):e1-e6. [CrossRef]
  19. Chevalier V, Mormont M, Curé H, Chollet P. Assessment of circadian rhythms by actimetry in healthy subjects and patients with advanced colorectal cancer. Oncol Rep 2003;10(3):733-737. [Medline]
  20. Dean GE, Redeker NS, Wang Y, Rogers AE, Dickerson SS, Steinbrenner LM, et al. Sleep, mood, and quality of life in patients receiving treatment for lung cancer. Oncol Nurs Forum 2013 Sep;40(5):441-451 [FREE Full text] [CrossRef] [Medline]
  21. Dreher N, Hadeler EK, Hartman SJ, Wong EC, Acerbi I, Rugo HS, et al. Fitbit Usage in Patients With Breast Cancer Undergoing Chemotherapy. Clin Breast Cancer 2019 Dec;19(6):443-449.e1. [CrossRef] [Medline]
  22. Edbrooke L, Granger CL, Clark RA, Denehy L. Physical Activity Levels Are Low in Inoperable Lung Cancer: Exploratory Analyses from a Randomised Controlled Trial. J Clin Med 2019 Aug 23;8(9) [FREE Full text] [CrossRef] [Medline]
  23. Gupta A, Stewart T, Bhulani N, Dong Y, Rahimi Z, Crane K, et al. Feasibility of Wearable Physical Activity Monitors in Patients With Cancer. JCO Clin Cancer Inform 2018 Dec;2:1-10. [CrossRef] [Medline]
  24. Innominato PF, Komarzynski S, Mohammad-Djafari A, Arbaud A, Ulusakarya A, Bouchahda M, et al. Clinical Relevance of the First Domomedicine Platform Securing Multidrug Chronotherapy Delivery in Metastatic Cancer Patients at Home: The inCASA European Project. J Med Internet Res 2016 Nov 25;18(11):e305 [FREE Full text] [CrossRef] [Medline]
  25. Li W, Kwok CC, Chan DC, Ho AW, Ho C, Zhang J, et al. Disruption of sleep, sleep-wake activity rhythm, and nocturnal melatonin production in breast cancer patients undergoing adjuvant chemotherapy: prospective cohort study. Sleep Med 2019 Mar;55:14-21. [CrossRef] [Medline]
  26. Lowe SS, Danielson B, Beaumont C, Watanabe SM, Baracos VE, Courneya KS. Associations Between Objectively Measured Physical Activity and Quality of Life in Cancer Patients With Brain Metastases. Journal of Pain and Symptom Management 2014 Sep;48(3):322-332. [CrossRef]
  27. Mouri T, Naito T, Morikawa A, Tatematsu N, Miura S, Okayama T, et al. Promotion of Behavioral Change and the Impact on Quality of Life in Elderly Patients with Advanced Cancer: A Physical Activity Intervention of the Multimodal Nutrition and Exercise Treatment for Advanced Cancer Program. Asia Pac J Oncol Nurs 2018 Dec;5(4):383-390 [FREE Full text] [CrossRef] [Medline]
  28. Nyrop KA, Deal AM, Choi SK, Wagoner CW, Lee JT, Wood A, et al. Measuring and understanding adherence in a home-based exercise intervention during chemotherapy for early breast cancer. Breast Cancer Res Treat 2017 Nov 9;168(1):43-55. [CrossRef]
  29. Ohri N, Halmos B, Bodner WR, Cheng H, Guha C, Kalnicki S, et al. Daily Step Counts: A New Prognostic Factor in Locally Advanced Non-Small Cell Lung Cancer? Int J Radiat Oncol Biol Phys 2019 Nov 15;105(4):745-751. [CrossRef] [Medline]
  30. Ohri N, Kabarriti R, Bodner WR, Mehta KJ, Shankar V, Halmos B, et al. Continuous Activity Monitoring During Concurrent Chemoradiotherapy. International Journal of Radiation Oncology*Biology*Physics 2017 Apr;97(5):1061-1065. [CrossRef]
  31. Ortiz-Tudela E, Iurisci I, Beau J, Karaboue A, Moreau T, Rol MA, et al. The circadian rest-activity rhythm, a potential safety pharmacology endpoint of cancer chemotherapy. Int J Cancer 2014 Jun 1;134(11):2717-2725 [FREE Full text] [CrossRef] [Medline]
  32. Parker NH, Ngo-Huang A, Lee RE, O'Connor DP, Basen-Engquist KM, Petzel MQB, et al. Physical activity and exercise during preoperative pancreatic cancer treatment. Support Care Cancer 2019 Jun;27(6):2275-2284. [CrossRef] [Medline]
  33. Roche VP, Mohamad-Djafari A, Innominato PF, Karaboué A, Gorbach A, Lévi FA. Thoracic surface temperature rhythms as circadian biomarkers for cancer chronotherapy. Chronobiol Int 2014 Apr;31(3):409-420 [FREE Full text] [CrossRef] [Medline]
  34. Roscoe JA, Morrow GR, Hickok JT, Bushunow P, Matteson S, Rakita D, et al. Temporal interrelationships among fatigue, circadian rhythm and depression in breast cancer patients undergoing chemotherapy treatment. Support Care Cancer 2002 May;10(4):329-336. [CrossRef] [Medline]
  35. Sarna L, Conde F. Physical activity and fatigue during radiation therapy: a pilot study using actigraph monitors. Oncol Nurs Forum 2001 Jul;28(6):1043-1046. [Medline]
  36. Savard J, Liu L, Natarajan L, Rissling MB, Neikrug AB, He F, et al. Breast cancer patients have progressively impaired sleep-wake activity rhythms during chemotherapy. Sleep 2009 Sep;32(9):1155-1160 [FREE Full text] [CrossRef] [Medline]
  37. Solk P, Gavin K, Fanning J, Welch W, Lloyd G, Cottrell A, et al. Feasibility and acceptability of intensive longitudinal data collection of activity and patient-reported outcomes during chemotherapy for breast cancer. Qual Life Res 2019 Dec;28(12):3333-3346. [CrossRef] [Medline]
  38. van der Meij BS, Langius JAE, Spreeuwenberg MD, Slootmaker SM, Paul MA, Smit EF, et al. Oral nutritional supplements containing n-3 polyunsaturated fatty acids affect quality of life and functional status in lung cancer patients during multimodality treatment: an RCT. Eur J Clin Nutr 2012 Mar;66(3):399-404 [FREE Full text] [CrossRef] [Medline]
  39. Vassbakk-Brovold K, Kersten C, Fegran L, Mjåland O, Mjåland S, Seiler S, et al. Cancer patients participating in a lifestyle intervention during chemotherapy greatly over-report their physical activity level: a validation study. BMC Sports Sci Med Rehabil 2016;8:10 [FREE Full text] [CrossRef] [Medline]
  40. Wright AA, Raman N, Staples P, Schonholz S, Cronin A, Carlson K, et al. The HOPE Pilot Study: Harnessing Patient-Reported Outcomes and Biometric Data to Enhance Cancer Care. JCO Clin Cancer Inform 2018 Dec;2:1-12. [CrossRef] [Medline]
  41. Wood WA, Basch E. From Intuition to Execution: Realizing the Potential of Wearables in Oncology. J Oncol Pract 2017 Feb;13(2):90-92. [CrossRef] [Medline]
  42. Lancaster K, Abuzour A, Khaira M. The Use and Effects of Electronic Health Tools for Patient Self-Monitoring and Reporting of Outcomes Following Medication Useystematic Review. Journal of medical Internet research Dec 18;. PubMed 2018;20(12):e294. [CrossRef] [Medline]
  43. Cox SM, Lane A, Volchenboum SL. Use of Wearable, Mobile, and Sensor Technology in Cancer Clinical Trials. JCO Clin Cancer Inform 2018 Dec;2:1-11. [CrossRef] [Medline]
  44. Brickwood K, Watson G, O'Brien J, Williams AD. Consumer-Based Wearable Activity Trackers Increase Physical Activity Participation: Systematic Review and Meta-Analysis. JMIR Mhealth Uhealth 2019 Apr 12;7(4):e11819 [FREE Full text] [CrossRef] [Medline]
  45. Goldsack JC, Coravos A, Bakker JP, Bent B, Dowling AV, Fitzer-Attas C, et al. Verification, analytical validation, and clinical validation (V3): the foundation of determining fit-for-purpose for Biometric Monitoring Technologies (BioMeTs). NPJ Digit Med 2020;3:55 [FREE Full text] [CrossRef] [Medline]
  46. Holländer-Mieritz C, Johansen C, Pappot H. eHealth-mind the gap. Acta Oncol 2020 Aug;59(8):877-878. [CrossRef] [Medline]
  47. Vegesna A, Tran M, Angelaccio M, Arcona S. Remote Patient Monitoring via Non-Invasive Digital Technologies: A Systematic Review. Telemed J E Health 2017 Jan;23(1):3-17 [FREE Full text] [CrossRef] [Medline]
  48. Weinstein RS, Lopez AM, Joseph BA, Erps KA, Holcomb M, Barker GP, et al. Telemedicine, telehealth, and mobile health applications that work: opportunities and barriers. Am J Med 2014 Mar;127(3):183-187. [CrossRef] [Medline]


CTTI: Clinical Transformation Initiative
GI: gastrointestinal
GYN: gynecological
PA: physical activity
PRO: patient-reported outcome
RCT: randomized controlled trial


Edited by G Eysenbach; submitted 02.07.20; peer-reviewed by W Wood; comments to author 10.07.20; revised version received 04.09.20; accepted 07.09.20; published 11.11.20

Copyright

©Ulrikke Lyng Beauchamp, Helle Pappot, Cecilie Holländer-Mieritz. Originally published in JMIR mHealth and uHealth (http://mhealth.jmir.org), 11.11.2020.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mHealth and uHealth, is properly cited. The complete bibliographic information, a link to the original publication on http://mhealth.jmir.org/, as well as this copyright and license information must be included.