<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing DTD v2.0 20040830//EN" "http://dtd.nlm.nih.gov/publishing/2.0/journalpublishing.dtd">
<article xmlns:xlink="http://www.w3.org/1999/xlink" article-type="review-article" dtd-version="2.0">
  <front>
    <journal-meta>
      <journal-id journal-id-type="publisher-id">JMU</journal-id>
      <journal-id journal-id-type="nlm-ta">JMIR Mhealth Uhealth</journal-id>
      <journal-title>JMIR mHealth and uHealth</journal-title>
      <issn pub-type="epub">2291-5222</issn>
      <publisher>
        <publisher-name>JMIR Publications</publisher-name>
        <publisher-loc>Toronto, Canada</publisher-loc>
      </publisher>
    </journal-meta>
    <article-meta>
      <article-id pub-id-type="publisher-id">v8i9e18694</article-id>
      <article-id pub-id-type="pmid">32897239</article-id>
      <article-id pub-id-type="doi">10.2196/18694</article-id>
      <article-categories>
        <subj-group subj-group-type="heading">
          <subject>Review</subject>
        </subj-group>
        <subj-group subj-group-type="article-type">
          <subject>Review</subject>
        </subj-group>
      </article-categories>
      <title-group>
        <article-title>Reliability and Validity of Commercially Available Wearable Devices for Measuring Steps, Energy Expenditure, and Heart Rate: Systematic Review</article-title>
      </title-group>
      <contrib-group>
        <contrib contrib-type="editor">
          <name>
            <surname>Eysenbach</surname>
            <given-names>Gunther</given-names>
          </name>
        </contrib>
      </contrib-group>
      <contrib-group>
        <contrib contrib-type="reviewer">
          <name>
            <surname>Lott</surname>
            <given-names>Donovan</given-names>
          </name>
        </contrib>
        <contrib contrib-type="reviewer">
          <name>
            <surname>Becker</surname>
            <given-names>Linda</given-names>
          </name>
        </contrib>
      </contrib-group>
      <contrib-group>
        <contrib id="contrib1" contrib-type="author" corresp="yes" equal-contrib="yes">
          <name name-style="western">
            <surname>Fuller</surname>
            <given-names>Daniel</given-names>
          </name>
          <degrees>PhD</degrees>
          <xref rid="aff1" ref-type="aff">1</xref>
          <address>
            <institution>School of Human Kinetics and Recreation</institution>
            <addr-line>Memorial University of Newfoundland</addr-line>
            <addr-line>St. John's, NL, A1C 5S7</addr-line>
            <country>Canada</country>
            <phone>1 7098647270</phone>
            <email>dfuller@mun.ca</email>
          </address>
          <xref rid="aff2" ref-type="aff">2</xref>
          <xref rid="aff3" ref-type="aff">3</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0002-2015-2955</ext-link>
        </contrib>
        <contrib id="contrib2" contrib-type="author" equal-contrib="yes">
          <name name-style="western">
            <surname>Colwell</surname>
            <given-names>Emily</given-names>
          </name>
          <degrees>MSc</degrees>
          <xref rid="aff1" ref-type="aff">1</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0002-1045-6234</ext-link>
        </contrib>
        <contrib id="contrib3" contrib-type="author" equal-contrib="yes">
          <name name-style="western">
            <surname>Low</surname>
            <given-names>Jonathan</given-names>
          </name>
          <degrees>MSc</degrees>
          <xref rid="aff1" ref-type="aff">1</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0002-1964-6237</ext-link>
        </contrib>
        <contrib id="contrib4" contrib-type="author">
          <name name-style="western">
            <surname>Orychock</surname>
            <given-names>Kassia</given-names>
          </name>
          <degrees>MSc</degrees>
          <xref rid="aff1" ref-type="aff">1</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0002-1355-8842</ext-link>
        </contrib>
        <contrib id="contrib5" contrib-type="author">
          <name name-style="western">
            <surname>Tobin</surname>
            <given-names>Melissa Ann</given-names>
          </name>
          <degrees>MSc</degrees>
          <xref rid="aff4" ref-type="aff">4</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0002-4128-4690</ext-link>
        </contrib>
        <contrib id="contrib6" contrib-type="author">
          <name name-style="western">
            <surname>Simango</surname>
            <given-names>Bo</given-names>
          </name>
          <degrees>MSc</degrees>
          <xref rid="aff1" ref-type="aff">1</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0003-3695-2815</ext-link>
        </contrib>
        <contrib id="contrib7" contrib-type="author">
          <name name-style="western">
            <surname>Buote</surname>
            <given-names>Richard</given-names>
          </name>
          <degrees>MSc</degrees>
          <xref rid="aff4" ref-type="aff">4</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0001-7981-7854</ext-link>
        </contrib>
        <contrib id="contrib8" contrib-type="author">
          <name name-style="western">
            <surname>Van Heerden</surname>
            <given-names>Desiree</given-names>
          </name>
          <xref rid="aff5" ref-type="aff">5</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0001-8994-8428</ext-link>
        </contrib>
        <contrib id="contrib9" contrib-type="author">
          <name name-style="western">
            <surname>Luan</surname>
            <given-names>Hui</given-names>
          </name>
          <degrees>PhD</degrees>
          <xref rid="aff6" ref-type="aff">6</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0001-9884-8509</ext-link>
        </contrib>
        <contrib id="contrib10" contrib-type="author">
          <name name-style="western">
            <surname>Cullen</surname>
            <given-names>Kimberley</given-names>
          </name>
          <degrees>PhD</degrees>
          <xref rid="aff1" ref-type="aff">1</xref>
          <xref rid="aff3" ref-type="aff">3</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0002-7904-9151</ext-link>
        </contrib>
        <contrib id="contrib11" contrib-type="author">
          <name name-style="western">
            <surname>Slade</surname>
            <given-names>Logan</given-names>
          </name>
          <degrees>BSc</degrees>
          <xref rid="aff4" ref-type="aff">4</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0003-2641-4124</ext-link>
        </contrib>
        <contrib id="contrib12" contrib-type="author">
          <name name-style="western">
            <surname>Taylor</surname>
            <given-names>Nathan G A</given-names>
          </name>
          <degrees>MPH</degrees>
          <xref rid="aff7" ref-type="aff">7</xref>
          <ext-link ext-link-type="orcid">https://orcid.org/0000-0001-6478-8165</ext-link>
        </contrib>
      </contrib-group>
      <aff id="aff1">
        <label>1</label>
        <institution>School of Human Kinetics and Recreation</institution>
        <institution>Memorial University</institution>
        <addr-line>St. John's, NL</addr-line>
        <country>Canada</country>
      </aff>
      <aff id="aff2">
        <label>2</label>
        <institution>Department of Computer Science</institution>
        <institution>Memorial University</institution>
        <addr-line>St. John's, NL</addr-line>
        <country>Canada</country>
      </aff>
      <aff id="aff3">
        <label>3</label>
        <institution>Division of Community Health and Humanities</institution>
        <institution>Faculty of Medicine</institution>
        <institution>Memorial University</institution>
        <addr-line>St. John's, NL</addr-line>
        <country>Canada</country>
      </aff>
      <aff id="aff4">
        <label>4</label>
        <institution>Faculty of Medicine</institution>
        <institution>Memorial University</institution>
        <addr-line>St. John's, NL</addr-line>
        <country>Canada</country>
      </aff>
      <aff id="aff5">
        <label>5</label>
        <institution>Faculty of Engineering</institution>
        <institution>Memorial University</institution>
        <addr-line>St. John's, NL</addr-line>
        <country>Canada</country>
      </aff>
      <aff id="aff6">
        <label>6</label>
        <institution>Department of Geography</institution>
        <institution>University of Oregon</institution>
        <addr-line>Eugene, OR</addr-line>
        <country>United States</country>
      </aff>
      <aff id="aff7">
        <label>7</label>
        <institution>School of Health Administration</institution>
        <institution>Dalhousie University</institution>
        <addr-line>Halifax, NS</addr-line>
        <country>Canada</country>
      </aff>
      <author-notes>
        <corresp>Corresponding Author: Daniel Fuller <email>dfuller@mun.ca</email></corresp>
      </author-notes>
      <pub-date pub-type="collection">
        <month>9</month>
        <year>2020</year>
      </pub-date>
      <pub-date pub-type="epub">
        <day>8</day>
        <month>9</month>
        <year>2020</year>
      </pub-date>
      <volume>8</volume>
      <issue>9</issue>
      <elocation-id>e18694</elocation-id>
      <history>
        <date date-type="received">
          <day>30</day>
          <month>3</month>
          <year>2020</year>
        </date>
        <date date-type="rev-request">
          <day>12</day>
          <month>6</month>
          <year>2020</year>
        </date>
        <date date-type="rev-recd">
          <day>22</day>
          <month>6</month>
          <year>2020</year>
        </date>
        <date date-type="accepted">
          <day>25</day>
          <month>6</month>
          <year>2020</year>
        </date>
      </history>
      <copyright-statement>©Daniel Fuller, Emily Colwell, Jonathan Low, Kassia Orychock, Melissa Ann Tobin, Bo Simango, Richard Buote, Desiree Van Heerden, Hui Luan, Kimberley Cullen, Logan Slade, Nathan G A Taylor. Originally published in JMIR mHealth and uHealth (http://mhealth.jmir.org), 08.09.2020.</copyright-statement>
      <copyright-year>2020</copyright-year>
      <license license-type="open-access" xlink:href="https://creativecommons.org/licenses/by/4.0/">
        <p>This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mHealth and uHealth, is properly cited. The complete bibliographic information, a link to the original publication on http://mhealth.jmir.org/, as well as this copyright and license information must be included.</p>
      </license>
      <self-uri xlink:href="http://mhealth.jmir.org/2020/9/e18694/" xlink:type="simple"/>
      <abstract>
        <sec sec-type="background">
          <title>Background</title>
          <p>Consumer-wearable activity trackers are small electronic devices that record fitness and health-related measures.</p>
        </sec>
        <sec sec-type="objective">
          <title>Objective</title>
          <p>The purpose of this systematic review was to examine the validity and reliability of commercial wearables in measuring step count, heart rate, and energy expenditure.</p>
        </sec>
        <sec sec-type="methods">
          <title>Methods</title>
          <p>We identified devices to be included in the review. Database searches were conducted in PubMed, Embase, and SPORTDiscus, and only articles published in the English language up to May 2019 were considered. Studies were excluded if they did not identify the device used and if they did not examine the validity or reliability of the device. Studies involving the general population and all special populations were included. We operationalized validity as criterion validity (as compared with other measures) and construct validity (degree to which the device is measuring what it claims). Reliability measures focused on intradevice and interdevice reliability.</p>
        </sec>
        <sec sec-type="results">
          <title>Results</title>
          <p>We included 158 publications examining nine different commercial wearable device brands. Fitbit was by far the most studied brand. In laboratory-based settings, Fitbit, Apple Watch, and Samsung appeared to measure steps accurately. Heart rate measurement was more variable, with Apple Watch and Garmin being the most accurate and Fitbit tending toward underestimation. For energy expenditure, no brand was accurate. We also examined validity between devices within a specific brand.</p>
        </sec>
        <sec sec-type="conclusions">
          <title>Conclusions</title>
          <p>Commercial wearable devices are accurate for measuring steps and heart rate in laboratory-based settings, but this varies by the manufacturer and device type. Devices are constantly being upgraded and redesigned to new models, suggesting the need for more current reviews and research.</p>
        </sec>
      </abstract>
      <kwd-group>
        <kwd>commercial wearable devices</kwd>
        <kwd>systematic review</kwd>
        <kwd>heart rate</kwd>
        <kwd>energy expenditure</kwd>
        <kwd>step count</kwd>
        <kwd>Fitbit</kwd>
        <kwd>Apple Watch</kwd>
        <kwd>Garmin</kwd>
        <kwd>Polar</kwd>
      </kwd-group>
    </article-meta>
  </front>
  <body>
    <sec sec-type="introduction">
      <title>Introduction</title>
      <p>Globally, physical inactivity is a pressing public health concern. A recent report suggested that about 23% of adults and 81% of school-going adolescents are not meeting physical activity guidelines [<xref ref-type="bibr" rid="ref1">1</xref>]. Government organizations have attempted to improve these numbers by implementing initiatives aimed at promoting physical activity. Though the successful promotion of physical activity is a complex multifacetted issue, behavior change is a well-established method to increase physical activity [<xref ref-type="bibr" rid="ref2">2</xref>]. Metrics defining physical activity guidelines from commercial wearable devices have been developed, including 10,000 steps per day [<xref ref-type="bibr" rid="ref3">3</xref>,<xref ref-type="bibr" rid="ref4">4</xref>] and 100 steps per minute for moderate to vigorous activity [<xref ref-type="bibr" rid="ref5">5</xref>]. However, research has shown variation in step count among devices, and the applicability of these metrics may vary by device brand and device type [<xref ref-type="bibr" rid="ref6">6</xref>].</p>
      <p>Research examining consumer wearable devices, such as watches, pendants, armbands, and other accessories, is associated with various labels including Quantified Self [<xref ref-type="bibr" rid="ref7">7</xref>] and mobile health (mHealth) [<xref ref-type="bibr" rid="ref8">8</xref>]. These consumer wearable devices are becoming increasingly popular for purchase and use. It has been estimated that in the year 2019, 225 million consumer wearables were sold [<xref ref-type="bibr" rid="ref9">9</xref>], and studies have suggested that more than a third of adults in Canada and Australia own and use a consumer wearable device [<xref ref-type="bibr" rid="ref10">10</xref>,<xref ref-type="bibr" rid="ref11">11</xref>]. Despite their popularity, research is equivocal about whether commercial wearable devices are valid and reliable methods for estimating metrics associated with physical activity including steps, heart rate, and energy expenditure.</p>
      <p>In a recent review of 10 articles, Bunn et al [<xref ref-type="bibr" rid="ref12">12</xref>] noted tendencies of wearables to underestimate energy expenditure, heart rate, and step count. Fitbit wearables were highly correlated with criterion measures of step count during laboratory-based assessment and had consistently high interdevice reliability for both step count and energy expenditure [<xref ref-type="bibr" rid="ref13">13</xref>]. However, this review found that these devices tended to underestimate energy expenditure, which is consistent with a separate review of Fitbit accuracy [<xref ref-type="bibr" rid="ref14">14</xref>] indicating that Fitbit wearables provide accurate measures only in limited circumstances.</p>
      <p>Commercial wearable devices have the potential to allow for population-level measurement of physical activity and large-scale behavior change. However, questions remain about their reliability and validity. This is especially true of smaller and newer manufacturers of wearable devices for which few or no reliability and validity studies have been conducted. The purpose of this systematic review was to outline and summarize information about the validity and reliability of wearables in measuring step count, heart rate, and energy expenditure in any population. The information summarized herein can be used to inform consumers and can aid researchers in study design when selecting physical activity monitoring devices.</p>
    </sec>
    <sec sec-type="methods">
      <title>Methods</title>
      <sec>
        <title>Design</title>
        <p>This systematic review was conducted and reported according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines [<xref ref-type="bibr" rid="ref14">14</xref>]. The review was not registered with PROSPERO. Full-length peer-reviewed original research articles, short reports, and letters to the editor published from January 1, 2000, through May 28, 2019, were included in the search. We limited the search to articles published after the year 2000 because commercial wearable devices were not truly available before that time.</p>
      </sec>
      <sec>
        <title>Search Strategy</title>
        <p>We conducted a literature search of the following databases: MEDLINE via PubMed (1946 to present); Embase (1947 to present); and SPORTDiscus with full text (1920 to present) via EBSCO. The reference lists of eligible papers were reviewed for additional pertinent references.</p>
        <p>A librarian (KR) developed the MEDLINE search strategy, which was peer reviewed by a second librarian according to the Peer Review of Electronic Search Strategies (PRESS) 2015 Guideline Statement [<xref ref-type="bibr" rid="ref15">15</xref>]. The MEDLINE strategy, which included Medical Subject Heading terms and text words, was translated for the other databases using database-specific controlled vocabulary. We searched the literature using multiple combinations and forms of the following key terms: <italic>accelerometer</italic>, <italic>fitness tracker</italic>, <italic>activity monitor</italic>, <italic>step count</italic>, <italic>wearable device</italic>, <italic>validity</italic>, <italic>reliability</italic>, <italic>accuracy</italic>, <italic>Fitbit</italic>, <italic>Garmin</italic>, <italic>Misfit</italic>, <italic>Jawbone</italic>, <italic>UnderArmour</italic>, <italic>Samsung</italic>, <italic>Apple watch</italic>, <italic>GENEactiv</italic>, <italic>Empatica</italic>, <italic>Mio</italic>, <italic>Amiigo</italic>, <italic>Xiaomi</italic>, <italic>Actigraph</italic>, <italic>Withings</italic>, and <italic>Sensewear</italic> (see <xref ref-type="supplementary-material" rid="app1">Multimedia Appendix 1</xref> for the full search strategies). An English language limit was applied. We included any abstracts and conference proceedings, as well as articles examining any population in the initial search. References were imported into EndNote X8 software (Clarivate Analytics) where duplicate references were removed. The remaining references were then imported into Covidence software (Veritas Health Innovation) for screening.</p>
      </sec>
      <sec>
        <title>Study Selection Strategy</title>
        <p>The web-based systematic review software Covidence was used for this review. The titles and abstracts of the studies included from the initial database search were independently assessed by at least two authors from the team. Conflicts arising during any step of the screening for inclusion/exclusion were resolved by a third author or by consensus. Following the title and abstract screening, full-text documents of the selected studies were searched and retrieved and were independently assessed for inclusion by at least two authors (EC, JL, and DF). Any conflicts were resolved by discussion and consensus. All reviewers strictly adhered to the defined inclusion criteria.</p>
      </sec>
      <sec>
        <title>Eligibility Criteria</title>
        <p>Studies that met the following criteria were included in the review: (1) use of any consumer-wearable model from the brand Apple Inc, Empatica, Fitbit, Garmin, Jawbone, Mio, Misfit, Polar, Samsung, UnderArmour, Withings, or Xiaomi; (2) specific examination of the reliability and validity measures of the aforementioned brands; and (3) examination of the device’s ability to measure a variable (step count, heart rate, or energy expenditure). Studies with fewer than 10 participants were excluded, as has been done in previous work [<xref ref-type="bibr" rid="ref13">13</xref>]. Validity of the wearable devices was defined as follows [<xref ref-type="bibr" rid="ref16">16</xref>]:</p>
        <list list-type="bullet">
          <list-item>
            <p>Criterion validity: comparing the devices to a criterion measure of steps, heart rate, or energy expenditure.</p>
          </list-item>
        </list>
        <p>Reliability of the trackers included the following [<xref ref-type="bibr" rid="ref16">16</xref>]:</p>
        <list list-type="bullet">
          <list-item>
            <p>Intradevice reliability: consistent test-retest results conducted within the same device.</p>
          </list-item>
          <list-item>
            <p>Interdevice reliability: consistent results across the same model of wearable device measured at the same time and worn at the same location.</p>
          </list-item>
        </list>
        <p>The main exclusion criteria were non-English studies, opinion/magazine articles, and systematic reviews. The initial database search and title/abstract screening included articles examining the accuracy of research-grade wearable devices, but the number of returned results was unmanageable. In order to further elucidate the research question in regard to consumer-wearable devices, before full-text screening, the decision was made to exclude all studies examining the reliability and validity of research-grade devices (Actigraph, GENEactiv, Amiigo, Sensewear Armband, Yamax, Omron, Kenze Lifecorder, Digiwalker, Actical, and Actiheart). Studies in which heart rate and energy expenditure estimates were collected using a chest strap heart rate monitor and transmitted to a wearable device were also excluded. Following text screening, the decision was made to exclude abstracts and conference papers. Following data extraction, the decision was made to exclude all studies examining Jawbone commercial wearables, as the company’s application program interface (API) was taken offline in 2018, rendering associated devices defunct. Studies were included in the final review if they had extractable data for the following criterion validity measures: correlation coefficient, group mean or percentage difference, median or mean absolute percentage error (MAPE), or level-of-agreement analysis, or had correlation coefficients for reliability measures. Authors were not contacted if these data were not reported in published or supplementary material. The remaining articles were those that met the inclusion criteria (consumer-grade wearables).</p>
      </sec>
      <sec>
        <title>Risk of Bias</title>
        <p>In our risk of bias assessment, comparisons that did not report group percentage differences or correlation coefficients (n=192) were excluded from the quantitative analysis. However, rather than exclude these comparisons and studies from the review completely, we included them in a narrative summary of how the measures reported were or were not consistent with exploration of percentage measurement error and correlation.</p>
      </sec>
      <sec>
        <title>Data Extraction</title>
        <p>We first conducted and documented an in-depth web search of the available consumer-wearable models and their specifications (placement, size, weight, cost, and connectivity). The data extraction process then consisted of the following: (1) categorizing the selected full-text articles into reliability or validity studies (EC, JL, and DF); (2) using a modification of the modified Consensus-Based Standards for the Selection of Health Status Measurement Instruments (COSMIN) validation subscale used by Feehan et al [<xref ref-type="bibr" rid="ref13">13</xref>] and an a priori modified COSMIN reliability subscale (<xref ref-type="supplementary-material" rid="app2">Multimedia Appendix 2</xref>) to assess the quality and risk of bias of each study (EC and DF); (3) extracting the key characteristics from each selected publication and compiling them into tables. Details from each reviewer were compared, and inconsistencies were resolved through consensus before compiling the results (EC and DF).</p>
        <p>Data extracted included characteristics of studies, participants, and devices, including study setting and activity type, outcomes measured, and type of criterion measure used. Correlation coefficients were extracted for all reliability comparisons reported in each study. Correlation coefficients, percentage difference and group mean values, MAPE values, and level-of-agreement data were extracted for all validity comparisons where available. Where group percentage differences were not reported, we calculated group percentage error ([wearable<sub>mean</sub> – criterion<sub>mean</sub>]/criterion<sub>mean</sub> × 100) to allow for comparison across studies. We split a small number of studies (n=10) into “substudies” (n=21), where separate populations were examined in the same publication (see <xref ref-type="supplementary-material" rid="app3">Multimedia Appendix 3</xref> for a more detailed breakdown).</p>
      </sec>
      <sec>
        <title>Syntheses</title>
        <p>Given the wide range of testing conditions and reported outcomes, we were unable to conduct meta-analyses of the extracted data. We instead conducted a narrative synthesis of the available quantitative data within each examined measure (step count, heart rate, and energy expenditure) using correlation comparisons and group percentage difference as the common metrics for criterion validity and correlation coefficient as the common metric for reliability.</p>
        <p>Our interpretation of measurement accuracy was focused on acceptable limits of percentage difference of ±3% in controlled settings and percentage difference of ±10% in free-living settings, as outlined in previous work [<xref ref-type="bibr" rid="ref13">13</xref>]. We interpreted correlation coefficients as follows: 0 to &#60;0.2, very weak; ≥0.2 to &#60;0.4, weak; ≥0.4 to &#60;0.6, moderate; ≥0.6 to &#60;0.8, strong; and ≥0.8 to 1.0, very strong [<xref ref-type="bibr" rid="ref17">17</xref>]. We completed all quantitative analyses and plots using RStudio version 1.2.1335 (RStudio Inc) and R version 3.6.0 (The R Foundation).</p>
        <p>Secondary analyses explored device brand. Brands were only included in these analyses when the group had 10 or more comparisons available for the measure. Studies that did not report data allowing for the examination of group percentage measurement error were still included in the review if they reported level of agreement or MAPE data. Such studies were included in the risk of bias assessment, the synthesis of study characteristics, and the narrative synthesis of study results.</p>
      </sec>
      <sec>
        <title>Availability of Data and Materials</title>
        <p>Data are publicly available on the BeapLab Dataverse [<xref ref-type="bibr" rid="ref18">18</xref>], and the analysis code is available on Github [<xref ref-type="bibr" rid="ref19">19</xref>].</p>
      </sec>
    </sec>
    <sec sec-type="results">
      <title>Results</title>
      <p>The initial literature search from the three databases yielded 34,890 unique citations (13,679 [39.21%] from PubMed, 17,560 [50.33%] from Embase, and 3651 [10.46%] from SPORTDiscus). Fourteen additional records were identified through other sources (eg, article reference lists and social media). After duplicate references were removed, 21,083 citations remained. Based on the subsequent title and abstract screening, 20,541 were rejected because they did not meet the inclusion criteria or met the exclusion criteria. Of the 542 that remained for full-text screening, 385 (71.0%) were further excluded for the following reasons: research-grade devices (n=311, 57.4%), wrong variable examined (n=24, 4.4%), fewer than 10 participants (n=14, 2.6%), abstracts (n=13, 2.4%), wrong consumer-grade brand examined (n=10, 1.9%; devices were Yamax, Omron, Kenz Lifecorder, Digiwalker, and uniaxial Actical/Actiheart), no extractable data (n=10, 1.9%), not peer reviewed (n=2, 0.4%), and conference paper (n=1, 0.2%). As a result, a total of 158 publications were included in this systematic review (<xref rid="figure1" ref-type="fig">Figure 1</xref>) [<xref ref-type="bibr" rid="ref14">14</xref>]. <xref ref-type="table" rid="table1">Table 1</xref> shows the details of the device brand, model, year, and status (current model or discontinued) in the included studies.</p>
      <fig id="figure1" position="float">
        <label>Figure 1</label>
        <caption>
          <p>PRISMA flow chart for systematic review of the reliability and validity of commercial wearable devices.</p>
        </caption>
        <graphic xlink:href="mhealth_v8i9e18694_fig1.png" alt-version="no" mimetype="image" position="float" xlink:type="simple"/>
      </fig>
      <table-wrap position="float" id="table1">
        <label>Table 1</label>
        <caption>
          <p>Device brand, model, year, current status, wear location, and studies used for the current systematic review.</p>
        </caption>
        <table width="1000" cellpadding="5" cellspacing="0" border="1" rules="groups" frame="hsides">
          <col width="90"/>
          <col width="130"/>
          <col width="100"/>
          <col width="190"/>
          <col width="130"/>
          <col width="360"/>
          <thead>
            <tr valign="top">
              <td>Brand</td>
              <td>Model</td>
              <td>Year</td>
              <td>Status</td>
              <td>Wear location</td>
              <td>Studies</td>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td>Apple</td>
              <td>Watch</td>
              <td>2015</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref20">20</xref>-<xref ref-type="bibr" rid="ref40">40</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Apple</td>
              <td>Watch Series 2</td>
              <td>2016</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref41">41</xref>-<xref ref-type="bibr" rid="ref44">44</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Alta</td>
              <td>2016</td>
              <td>Current model</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref45">45</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Blaze</td>
              <td>2016</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref22">22</xref>,<xref ref-type="bibr" rid="ref40">40</xref>,<xref ref-type="bibr" rid="ref43">43</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Charge</td>
              <td>2014</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref45">45</xref>-<xref ref-type="bibr" rid="ref56">56</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Charge 2</td>
              <td>2016</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref23">23</xref>,<xref ref-type="bibr" rid="ref30">30</xref>,<xref ref-type="bibr" rid="ref43">43</xref>,<xref ref-type="bibr" rid="ref44">44</xref>,<xref ref-type="bibr" rid="ref57">57</xref>-<xref ref-type="bibr" rid="ref63">63</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Charge HR</td>
              <td>2015</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref20">20</xref>,<xref ref-type="bibr" rid="ref21">21</xref>,<xref ref-type="bibr" rid="ref29">29</xref>,<xref ref-type="bibr" rid="ref32">32</xref>,<xref ref-type="bibr" rid="ref34">34</xref>,<xref ref-type="bibr" rid="ref36">36</xref>,<xref ref-type="bibr" rid="ref38">38</xref>,<xref ref-type="bibr" rid="ref45">45</xref>,<xref ref-type="bibr" rid="ref53">53</xref>,<xref ref-type="bibr" rid="ref64">64</xref>-<xref ref-type="bibr" rid="ref82">82</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Classic</td>
              <td>2009</td>
              <td>Discontinued</td>
              <td>Ankle/foot or waist/hip</td>
              <td>[<xref ref-type="bibr" rid="ref83">83</xref>-<xref ref-type="bibr" rid="ref87">87</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Flex</td>
              <td>2013</td>
              <td>Discontinued</td>
              <td>Thigh or wrist</td>
              <td>[<xref ref-type="bibr" rid="ref45">45</xref>,<xref ref-type="bibr" rid="ref50">50</xref>,<xref ref-type="bibr" rid="ref72">72</xref>,<xref ref-type="bibr" rid="ref79">79</xref>,<xref ref-type="bibr" rid="ref80">80</xref>,<xref ref-type="bibr" rid="ref88">88</xref>-<xref ref-type="bibr" rid="ref117">117</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Flex 2</td>
              <td>2017</td>
              <td>Current model</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref113">113</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Force</td>
              <td>2013</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref118">118</xref>,<xref ref-type="bibr" rid="ref119">119</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>One</td>
              <td>2012</td>
              <td>Discontinued</td>
              <td>Ankle/foot, pant pocket, waist/hip, or wrist</td>
              <td>[<xref ref-type="bibr" rid="ref34">34</xref>,<xref ref-type="bibr" rid="ref49">49</xref>,<xref ref-type="bibr" rid="ref52">52</xref>,<xref ref-type="bibr" rid="ref73">73</xref>,<xref ref-type="bibr" rid="ref80">80</xref>,<xref ref-type="bibr" rid="ref88">88</xref>,<xref ref-type="bibr" rid="ref90">90</xref>,<xref ref-type="bibr" rid="ref92">92</xref>,<xref ref-type="bibr" rid="ref93">93</xref>,<xref ref-type="bibr" rid="ref98">98</xref>,<xref ref-type="bibr" rid="ref100">100</xref>,<xref ref-type="bibr" rid="ref102">102</xref>,<xref ref-type="bibr" rid="ref103">103</xref>,<xref ref-type="bibr" rid="ref110">110</xref>, <xref ref-type="bibr" rid="ref116">116</xref>-<xref ref-type="bibr" rid="ref118">118</xref>,<xref ref-type="bibr" rid="ref120">120</xref>-<xref ref-type="bibr" rid="ref138">138</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Surge</td>
              <td>2015</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref27">27</xref>,<xref ref-type="bibr" rid="ref35">35</xref>,<xref ref-type="bibr" rid="ref42">42</xref>,<xref ref-type="bibr" rid="ref45">45</xref>,<xref ref-type="bibr" rid="ref54">54</xref>,<xref ref-type="bibr" rid="ref82">82</xref>,<xref ref-type="bibr" rid="ref139">139</xref>-<xref ref-type="bibr" rid="ref143">143</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Ultra</td>
              <td>2011</td>
              <td>Discontinued</td>
              <td>Chest, pant pocket, upper arm, waist/hip, or wrist</td>
              <td>[<xref ref-type="bibr" rid="ref85">85</xref>,<xref ref-type="bibr" rid="ref144">144</xref>-<xref ref-type="bibr" rid="ref148">148</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Fitbit</td>
              <td>Zip</td>
              <td>2012</td>
              <td>Current model</td>
              <td>Ankle/foot, pant pocket, shin, or waist/hip</td>
              <td>[<xref ref-type="bibr" rid="ref34">34</xref>,<xref ref-type="bibr" rid="ref45">45</xref>,<xref ref-type="bibr" rid="ref46">46</xref>,<xref ref-type="bibr" rid="ref51">51</xref>,<xref ref-type="bibr" rid="ref88">88</xref>,<xref ref-type="bibr" rid="ref89">89</xref>,<xref ref-type="bibr" rid="ref92">92</xref>,<xref ref-type="bibr" rid="ref93">93</xref>,<xref ref-type="bibr" rid="ref96">96</xref>,<xref ref-type="bibr" rid="ref103">103</xref>,<xref ref-type="bibr" rid="ref112">112</xref>,<xref ref-type="bibr" rid="ref119">119</xref>,<xref ref-type="bibr" rid="ref127">127</xref>,<xref ref-type="bibr" rid="ref129">129</xref>, <xref ref-type="bibr" rid="ref131">131</xref>,<xref ref-type="bibr" rid="ref141">141</xref>,<xref ref-type="bibr" rid="ref149">149</xref>-<xref ref-type="bibr" rid="ref161">161</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Fenix 3 HR</td>
              <td>2016</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref41">41</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Forerunner 225</td>
              <td>2015</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref21">21</xref>,<xref ref-type="bibr" rid="ref162">162</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Forerunner 235</td>
              <td>2015</td>
              <td>Current model</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref40">40</xref>,<xref ref-type="bibr" rid="ref139">139</xref>,<xref ref-type="bibr" rid="ref163">163</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Forerunner 405CX</td>
              <td>2009</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref164">164</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Forerunner 735XT</td>
              <td>2016</td>
              <td>Current model</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref35">35</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Forerunner 920XT</td>
              <td>2014</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref53">53</xref>] </td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Vivoactive</td>
              <td>2015</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref53">53</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Vivofit</td>
              <td>2014</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref35">35</xref>,<xref ref-type="bibr" rid="ref46">46</xref>,<xref ref-type="bibr" rid="ref50">50</xref>,<xref ref-type="bibr" rid="ref52">52</xref>,<xref ref-type="bibr" rid="ref53">53</xref>,<xref ref-type="bibr" rid="ref89">89</xref>,<xref ref-type="bibr" rid="ref92">92</xref>,<xref ref-type="bibr" rid="ref104">104</xref>,<xref ref-type="bibr" rid="ref114">114</xref>,<xref ref-type="bibr" rid="ref122">122</xref>,<xref ref-type="bibr" rid="ref130">130</xref>,<xref ref-type="bibr" rid="ref143">143</xref>,<xref ref-type="bibr" rid="ref150">150</xref>,<xref ref-type="bibr" rid="ref159">159</xref>, <xref ref-type="bibr" rid="ref165">165</xref>-<xref ref-type="bibr" rid="ref169">169</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Vivofit 2</td>
              <td>2015</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref34">34</xref>,<xref ref-type="bibr" rid="ref127">127</xref>,<xref ref-type="bibr" rid="ref170">170</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Vivofit 3</td>
              <td>2016</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref168">168</xref>,<xref ref-type="bibr" rid="ref171">171</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Vivosmart</td>
              <td>2014</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref32">32</xref>,<xref ref-type="bibr" rid="ref53">53</xref>,<xref ref-type="bibr" rid="ref75">75</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Vivosmart HR</td>
              <td>2015</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref36">36</xref>,<xref ref-type="bibr" rid="ref43">43</xref>,<xref ref-type="bibr" rid="ref65">65</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Garmin</td>
              <td>Vivosmart HR+</td>
              <td>2016</td>
              <td>Current model</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref58">58</xref>,<xref ref-type="bibr" rid="ref63">63</xref>,<xref ref-type="bibr" rid="ref140">140</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Mio</td>
              <td>Alpha</td>
              <td>2013</td>
              <td>Discontinued </td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref25">25</xref>,<xref ref-type="bibr" rid="ref38">38</xref>,<xref ref-type="bibr" rid="ref71">71</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Mio</td>
              <td>Fuse</td>
              <td>2015</td>
              <td>Discontinued </td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref54">54</xref>,<xref ref-type="bibr" rid="ref64">64</xref>] </td>
            </tr>
            <tr valign="top">
              <td>Misfit</td>
              <td>Flash</td>
              <td>2015</td>
              <td>Discontinued</td>
              <td>Waist/hip</td>
              <td>[<xref ref-type="bibr" rid="ref32">32</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Misfit</td>
              <td>Shine</td>
              <td>2012</td>
              <td>Discontinued</td>
              <td>Ankle/foot, chest, pant pocket, waist/hip, or wrist</td>
              <td>[<xref ref-type="bibr" rid="ref74">74</xref>,<xref ref-type="bibr" rid="ref79">79</xref>,<xref ref-type="bibr" rid="ref89">89</xref>,<xref ref-type="bibr" rid="ref96">96</xref>,<xref ref-type="bibr" rid="ref99">99</xref>,<xref ref-type="bibr" rid="ref104">104</xref>,<xref ref-type="bibr" rid="ref105">105</xref>,<xref ref-type="bibr" rid="ref131">131</xref>,<xref ref-type="bibr" rid="ref159">159</xref>,<xref ref-type="bibr" rid="ref169">169</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Polar</td>
              <td>A300</td>
              <td>2015</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref172">172</xref>] </td>
            </tr>
            <tr valign="top">
              <td>Polar</td>
              <td>A360</td>
              <td>2015</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref25">25</xref>,<xref ref-type="bibr" rid="ref43">43</xref>,<xref ref-type="bibr" rid="ref140">140</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Polar</td>
              <td>Active</td>
              <td>2011</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref173">173</xref>] </td>
            </tr>
            <tr valign="top">
              <td>Polar</td>
              <td>Loop</td>
              <td>2013</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref32">32</xref>,<xref ref-type="bibr" rid="ref50">50</xref>,<xref ref-type="bibr" rid="ref53">53</xref>,<xref ref-type="bibr" rid="ref79">79</xref>,<xref ref-type="bibr" rid="ref89">89</xref>,<xref ref-type="bibr" rid="ref167">167</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Polar</td>
              <td>M600</td>
              <td>2016</td>
              <td>Current</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref56">56</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Polar</td>
              <td>V800</td>
              <td>2016</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref174">174</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Samsung</td>
              <td>Gear 2</td>
              <td>2014</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref140">140</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Samsung</td>
              <td>Gear S</td>
              <td>2014</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref32">32</xref>,<xref ref-type="bibr" rid="ref38">38</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Samsung</td>
              <td>Gear S2</td>
              <td>2015</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref35">35</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Samsung</td>
              <td>Gear S3</td>
              <td>2016</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref42">42</xref>,<xref ref-type="bibr" rid="ref44">44</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Withings</td>
              <td>Pulse</td>
              <td>2013</td>
              <td>Discontinued</td>
              <td>Collar, pant pocket, waist/hip, or wrist</td>
              <td> [<xref ref-type="bibr" rid="ref89">89</xref>,<xref ref-type="bibr" rid="ref96">96</xref>,<xref ref-type="bibr" rid="ref131">131</xref>,<xref ref-type="bibr" rid="ref166">166</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Withings</td>
              <td>Pulse O2</td>
              <td>2013</td>
              <td>Discontinued</td>
              <td>Collar, waist/hip, or wrist</td>
              <td>[<xref ref-type="bibr" rid="ref104">104</xref>,<xref ref-type="bibr" rid="ref122">122</xref>,<xref ref-type="bibr" rid="ref123">123</xref>,<xref ref-type="bibr" rid="ref169">169</xref>,<xref ref-type="bibr" rid="ref175">175</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Withings</td>
              <td>Pulse Ox</td>
              <td>2014</td>
              <td>Current model</td>
              <td>Waist/hip or wrist</td>
              <td>[<xref ref-type="bibr" rid="ref53">53</xref>,<xref ref-type="bibr" rid="ref58">58</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Xiaomi</td>
              <td>Mi Band</td>
              <td>2014</td>
              <td>Discontinued </td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref42">42</xref>]</td>
            </tr>
            <tr valign="top">
              <td>Xiaomi</td>
              <td>Mi Band 2</td>
              <td>2016</td>
              <td>Discontinued</td>
              <td>Wrist</td>
              <td>[<xref ref-type="bibr" rid="ref81">81</xref>]</td>
            </tr>
          </tbody>
        </table>
      </table-wrap>
      <sec>
        <title>Study and Participant Characteristics</title>
        <p>Of the 158 publications included, 143 were full-text research articles, 10 were brief reports, and five were letters to the editor. Publication year ranged from 2013 to 2019, with the amount of publications increasing from 2013 to 2017 (2013, n=2; 2014, n=8; 2015, n=11; 2016, n=30; 2017, n=43). We also included an additional 40 and 24 studies published in 2018 and 2019, respectively.</p>
        <p>Within those 158 publications, 169 studies/substudies were identified. Among these, 168 (99.4%) examined validity and 19 (11.2%) examined reliability. Moreover, 126 studies examined step count (125 validity and 16 reliability), 32 examined heart rate (32 validity and 3 reliability), and 43 examined energy expenditure (42 validity and 5 reliability) (<xref rid="figure2" ref-type="fig">Figure 2</xref>). Furthermore, 130 examined populations in a controlled environment and 48 examined populations in a free-living environment. A total of 1838 comparisons were identified, of which 166 examined reliability (mean 8, SD 11 per reliability study; range 1-40) and 1672 examined validity (mean 10, SD 15 per validity study; range 1-98).</p>
        <fig id="figure2" position="float">
          <label>Figure 2</label>
          <caption>
            <p>Number of studies published per year by measurement type. EE: energy expenditure; HR: heart rate; SC: step count.</p>
          </caption>
          <graphic xlink:href="mhealth_v8i9e18694_fig2.png" alt-version="no" mimetype="image" position="float" xlink:type="simple"/>
        </fig>
        <p>The 169 studies/substudies comprised a total of 5934 participants, with a mean of 35 (SD 27) participants per study (range 10-185). One hundred and sixty-one studies reported sex, and 51.08% (2861/5601) of participants were female. One hundred and fifty-eight studies reported age, with a mean participant age of 36.8 years (SD 18.3; range 3.7-87 years). One hundred and fifty-nine studies examined adult populations (age ≥18 years) and 10 studies examined children. One hundred and thirty-three studies included only healthy participants, while the other 36 studies included participants with mobility limitations and/or chronic diseases (<xref ref-type="supplementary-material" rid="app4">Multimedia Appendix 4</xref>).</p>
        <p>Fitbit consumer-grade wearables were examined most frequently (144 studies examining 12 models), followed by Garmin (42 studies, 13 models), Apple (28 studies, 2 models), Polar (15 studies, 6 models), Misfit (13 studies, 2 models), Withings (12 studies, 2 models), Samsung (8 studies, 4 models), Mio (6 studies, 2 models), and Xiaomi (2 studies, 2 models) (a complete list of examined models is provided in <xref ref-type="supplementary-material" rid="app5">Multimedia Appendix 5</xref>) (<xref rid="figure3" ref-type="fig">Figure 3</xref>). Wearables were typically examined while worn on the wrist (n=131, examining at least one wrist-worn device) or at the waist/hip (n=71, locations included the waist, hip, belt, and pants pocket). Substantially fewer studies examined wearables worn on the torso (n=14, locations included the chest, bra, lanyard, and shirt collar) and lower limb (n=13, locations included the thigh, shin, ankle, and foot).</p>
        <fig id="figure3" position="float">
          <label>Figure 3</label>
          <caption>
            <p>Line graph of studies published per year by device brand.</p>
          </caption>
          <graphic xlink:href="mhealth_v8i9e18694_fig3.png" alt-version="no" mimetype="image" position="float" xlink:type="simple"/>
        </fig>
      </sec>
      <sec>
        <title>Risk of Bias</title>
        <p>Of 169 studies, 140 (82.8%; 1640 of 1838 [89.23%] comparisons) were rated fair or poor for sample size (&#60;50 participants), but were not excluded from the analysis owing to the paucity of studies with excellent (≥100 participants, n=7) and good (50-99 participants, n=22) sample sizes. We additionally explored the potential for bias related to sample size in step count, heart rate, and energy expenditure by examining the percentage error dispersion by sample size using scatter plots (<xref rid="figure4" ref-type="fig">Figure 4</xref>).</p>
        <fig id="figure4" position="float">
          <label>Figure 4</label>
          <caption>
            <p>Mean percentage error (MPE) plots by study sample size for step count, heart rate, and energy expenditure. The solid black line represents zero. The solid grey line represents average MPE for all data points. The dashed grey lines represent the 95% CIs.</p>
          </caption>
          <graphic xlink:href="mhealth_v8i9e18694_fig4.png" alt-version="no" mimetype="image" position="float" xlink:type="simple"/>
        </fig>
        <p>In these examinations, we saw no apparent systematic bias for measurement error beyond a small number of comparisons showing extreme overestimation (four comparisons in step count and five comparisons in energy expenditure). The four extreme outliers for step count involved measurement during sedentary and light physical activity in a single study with fewer than 40 participants [<xref ref-type="bibr" rid="ref20">20</xref>] and were likely inflated by the limited number of steps accumulated during those bouts. As a result, we excluded these four comparisons from the quantitative syntheses. Upon closer examination of the five extreme outliers for energy expenditure (four occurred in a study with greater than 60 participants [<xref ref-type="bibr" rid="ref21">21</xref>] and one occurred in a study with fewer than 40 participants [<xref ref-type="bibr" rid="ref41">41</xref>]), we determined that these were likely true reflections of tendencies to overestimate energy expenditure during sedentary and low-intensity activities, and therefore, we included these five comparisons in the quantitative syntheses.</p>
      </sec>
      <sec>
        <title>Validity: Controlled Settings</title>
        <p>We examined criterion validity for step count, heart rate, and energy expenditure separately for controlled and free-living settings. For controlled settings, we also had sufficient data to examine validity by brand and devices within brands.</p>
        <sec>
          <title>Validity for Step Count in Controlled Settings</title>
          <p>A total of 90 studies (979 comparisons) examined wearable device step count measurements compared with reference standard criterion measures of manual counting [<xref ref-type="bibr" rid="ref32">32</xref>,<xref ref-type="bibr" rid="ref34">34</xref>-<xref ref-type="bibr" rid="ref38">38</xref>,<xref ref-type="bibr" rid="ref42">42</xref>,<xref ref-type="bibr" rid="ref46">46</xref>,<xref ref-type="bibr" rid="ref47">47</xref>,<xref ref-type="bibr" rid="ref50">50</xref>-<xref ref-type="bibr" rid="ref53">53</xref>,<xref ref-type="bibr" rid="ref57">57</xref>,<xref ref-type="bibr" rid="ref58">58</xref>,<xref ref-type="bibr" rid="ref72">72</xref>,<xref ref-type="bibr" rid="ref80">80</xref>-<xref ref-type="bibr" rid="ref84">84</xref>,<xref ref-type="bibr" rid="ref88">88</xref>-<xref ref-type="bibr" rid="ref102">102</xref>,<xref ref-type="bibr" rid="ref109">109</xref>, <xref ref-type="bibr" rid="ref114">114</xref>-<xref ref-type="bibr" rid="ref125">125</xref>,<xref ref-type="bibr" rid="ref138">138</xref>-<xref ref-type="bibr" rid="ref141">141</xref>,<xref ref-type="bibr" rid="ref144">144</xref>-<xref ref-type="bibr" rid="ref147">147</xref>,<xref ref-type="bibr" rid="ref149">149</xref>-<xref ref-type="bibr" rid="ref153">153</xref>,<xref ref-type="bibr" rid="ref158">158</xref>-<xref ref-type="bibr" rid="ref161">161</xref>,<xref ref-type="bibr" rid="ref165">165</xref>,<xref ref-type="bibr" rid="ref169">169</xref>-<xref ref-type="bibr" rid="ref171">171</xref>,<xref ref-type="bibr" rid="ref173">173</xref>] 
          and accelerometry [<xref ref-type="bibr" rid="ref20">20</xref>,<xref ref-type="bibr" rid="ref60">60</xref>,<xref ref-type="bibr" rid="ref64">64</xref>-<xref ref-type="bibr" rid="ref66">66</xref>,<xref ref-type="bibr" rid="ref85">85</xref>,<xref ref-type="bibr" rid="ref103">103</xref>,<xref ref-type="bibr" rid="ref109">109</xref>,<xref ref-type="bibr" rid="ref126">126</xref>-<xref ref-type="bibr" rid="ref128">128</xref>,<xref ref-type="bibr" rid="ref148">148</xref>, <xref ref-type="bibr" rid="ref154">154</xref>,<xref ref-type="bibr" rid="ref164">164</xref>] (<xref ref-type="supplementary-material" rid="app6">Multimedia Appendix 6</xref>). Of these, 67 studies recruited healthy adults (mean age 35.4 years, SD 17.4 years), 20 studies recruited adults living with limited mobility/chronic diseases (mean age 60.1 years, SD 10.5 years), two studies recruited children living with limited mobility/chronic diseases (mean age 12.5 years, SD 2.9 years), and one study recruited healthy children (mean age 3.7 years, SD 0.6 years). Wearable devices were worn on the lower limb (foot, ankle, shin, and thigh), torso, waist/hip, and wrist.</p>
          <p>Group measurement error was reported or calculable for 805 of the 979 comparisons, regardless of the criterion measure. Of these, 45.2% (n=364) were within ±3% measurement error, 42.7% (n=344) were below −3% measurement error, and 12.1% (n=97) were above 3% measurement error. The overall tendency was to underestimate step count (mean: −9%, median: −2%).</p>
        </sec>
        <sec>
          <title>Validity for Heart Rate in Controlled Settings</title>
          <p>A total of 29 studies (266 comparisons) examined wearable device heart rate measurements compared with reference standard criterion measures, including electrocardiography [<xref ref-type="bibr" rid="ref22">22</xref>,<xref ref-type="bibr" rid="ref23">23</xref>,<xref ref-type="bibr" rid="ref38">38</xref>-<xref ref-type="bibr" rid="ref40">40</xref>,<xref ref-type="bibr" rid="ref43">43</xref>,<xref ref-type="bibr" rid="ref44">44</xref>,<xref ref-type="bibr" rid="ref54">54</xref>,<xref ref-type="bibr" rid="ref61">61</xref>,<xref ref-type="bibr" rid="ref62">62</xref>,<xref ref-type="bibr" rid="ref67">67</xref>-<xref ref-type="bibr" rid="ref70">70</xref>,<xref ref-type="bibr" rid="ref142">142</xref>,<xref ref-type="bibr" rid="ref162">162</xref>,<xref ref-type="bibr" rid="ref176">176</xref>], Polar brand chest straps [<xref ref-type="bibr" rid="ref20">20</xref>,<xref ref-type="bibr" rid="ref21">21</xref>,<xref ref-type="bibr" rid="ref24">24</xref>-<xref ref-type="bibr" rid="ref28">28</xref>,<xref ref-type="bibr" rid="ref58">58</xref>,<xref ref-type="bibr" rid="ref63">63</xref>,<xref ref-type="bibr" rid="ref71">71</xref>,<xref ref-type="bibr" rid="ref163">163</xref>], and pulse oximetry [<xref ref-type="bibr" rid="ref66">66</xref>], in controlled settings (a detailed list of the criterion measures used is presented in <xref ref-type="supplementary-material" rid="app6">Multimedia Appendix 6</xref>). Of these, 24 studies recruited healthy adults (mean age 29.8 years, SD 10.5 years), four studies recruited adults living with limited mobility/chronic diseases (mean age 59.6 years, SD 9.0 years), and one study recruited children undergoing surgery (mean age 8.2 years, SD 3.1 years). All wearable devices were worn on the wrist.</p>
          <p>Group measurement error was reported or calculable for 177 of 266 comparisons, regardless of the criterion measure. Of these, 56.5% (n=100) were within ±3% measurement error, 24.9% (n=44) were below −3% measurement error, and 18.6% (n=33) were above 3% measurement error. There was a slight overall tendency toward underestimation of heart rate (estimated median error: −1%).</p>
        </sec>
        <sec>
          <title>Validity for Energy Expenditure in Controlled Settings</title>
          <p>A total of 36 studies (312 comparisons) examined wearable device energy expenditure measurements compared with reference standard criterion measures, including direct calorimetry [<xref ref-type="bibr" rid="ref86">86</xref>,<xref ref-type="bibr" rid="ref104">104</xref>] and indirect calorimetry [<xref ref-type="bibr" rid="ref20">20</xref>,<xref ref-type="bibr" rid="ref21">21</xref>,<xref ref-type="bibr" rid="ref29">29</xref>-<xref ref-type="bibr" rid="ref31">31</xref>,<xref ref-type="bibr" rid="ref38">38</xref>,<xref ref-type="bibr" rid="ref39">39</xref>,<xref ref-type="bibr" rid="ref41">41</xref>-<xref ref-type="bibr" rid="ref43">43</xref>,<xref ref-type="bibr" rid="ref53">53</xref>,<xref ref-type="bibr" rid="ref55">55</xref>,<xref ref-type="bibr" rid="ref63">63</xref>,<xref ref-type="bibr" rid="ref66">66</xref>,<xref ref-type="bibr" rid="ref73">73</xref>,<xref ref-type="bibr" rid="ref85">85</xref>,<xref ref-type="bibr" rid="ref87">87</xref>,<xref ref-type="bibr" rid="ref93">93</xref>,<xref ref-type="bibr" rid="ref95">95</xref>,<xref ref-type="bibr" rid="ref97">97</xref>,<xref ref-type="bibr" rid="ref103">103</xref>, <xref ref-type="bibr" rid="ref105">105</xref>,<xref ref-type="bibr" rid="ref116">116</xref>,<xref ref-type="bibr" rid="ref117">117</xref>,<xref ref-type="bibr" rid="ref129">129</xref>,<xref ref-type="bibr" rid="ref130">130</xref>,<xref ref-type="bibr" rid="ref142">142</xref>,<xref ref-type="bibr" rid="ref143">143</xref>,<xref ref-type="bibr" rid="ref146">146</xref>,<xref ref-type="bibr" rid="ref148">148</xref>,<xref ref-type="bibr" rid="ref159">159</xref>,<xref ref-type="bibr" rid="ref165">165</xref>,<xref ref-type="bibr" rid="ref166">166</xref>,<xref ref-type="bibr" rid="ref177">177</xref>], in controlled settings. Of these, 35 studies recruited healthy adults (mean age 27.2 years, SD 7.1 years), and one study recruited adults living with cardiovascular disease (mean age 64.2 years, SD 2.3 years). Wearable devices were worn on the wrist, waist/hip, and torso.</p>
          <p>Group measurement error was reported or calculable for 305 of the 312 comparisons, regardless of the criterion measure. Of these, 9.2% (n=28) were within ±3% measurement error, 54.1% (n=165) were below −3% measurement error, and 36.7% (n=112) were above 3% measurement error. Studies showed a tendency to underestimate energy expenditure and to provide inaccurate measures of energy expenditure compared with the criterion.</p>
        </sec>
      </sec>
      <sec>
        <title>Validity in Controlled Settings by Brand</title>
        <p><xref rid="figure5" ref-type="fig">Figure 5</xref> shows the mean percentage error (MPE) for step count, heart rate, and energy expenditure by device brand for devices with 10 or more comparisons. <xref rid="figure6" ref-type="fig">Figure 6</xref> shows the MPE for step count, heart rate, and energy expenditure by device brand and model for devices with 10 or more comparisons.</p>
        <fig id="figure5" position="float">
          <label>Figure 5</label>
          <caption>
            <p>Box plots representing mean percentage error (MPE) for steps, heart rate, and energy expenditure by device brand for devices with 10 or more comparisons.</p>
          </caption>
          <graphic xlink:href="mhealth_v8i9e18694_fig5.png" alt-version="no" mimetype="image" position="float" xlink:type="simple"/>
        </fig>
        <fig id="figure6" position="float">
          <label>Figure 6</label>
          <caption>
            <p>Box plots representing mean percentage error (MPE) for steps, heart rate, and energy expenditure by device brand and model for devices with 10 or more comparisons.</p>
          </caption>
          <graphic xlink:href="mhealth_v8i9e18694_fig6.png" alt-version="no" mimetype="image" position="float" xlink:type="simple"/>
        </fig>
        <sec>
          <title>Validity for Step Count by Brand</title>
          <p>We observed that the error level varied by device brand (<xref rid="figure5" ref-type="fig">Figure 5</xref>). Withings and Misfit wearables consistently underestimated step count, and Apple and Samsung had less measurement variability than other brands. There are possible interactions between the number and size of studies and device wear location that may influence the brand comparisons. For example, Apple Watch and Samsung have the tightest ranges for step count estimates but have relatively fewer studies compared with other brands.</p>
        </sec>
        <sec>
          <title>Validity for Heart Rate by Brand</title>
          <p>For heart rate, measurement error also varied by device brand (<xref rid="figure5" ref-type="fig">Figure 5</xref>). Apple Watch was within ±3% 71% (35/49) of the time, while Fitbit wearables were within ±3% 51% (36/71) of the time and Garmin wearables were within ±3% 49% (23/47) of the time. Despite similar ±3% measurement error rates, Fitbit appeared to underestimate heart rate more than Apple Watch and Garmin.</p>
        </sec>
        <sec>
          <title>Validity for Energy Expenditure by Brand</title>
          <p>For energy expenditure estimates, no brand of wearable was within ±3% measurement error more than 13% of the time (<xref rid="figure5" ref-type="fig">Figure 5</xref>). Underestimation of energy expenditure (less than −3%) was observed in Garmin wearables 69% (37/51) of the time and in Withings wearables 74% (34/46) of the time. Conversely, Apple wearables overestimated energy expenditure 58% (18/31) of the time and Polar wearables overestimated energy expenditure 69% (9/13) of the time. Fitbit devices tended to provide inaccurate measures compared with the criterion, underestimating 48.4% (76/157) of the time and overestimating 39.5% (62/157) of the time, despite the boxplot in <xref rid="figure5" ref-type="fig">Figure 5</xref> showing a reasonable median value for accuracy.</p>
        </sec>
      </sec>
      <sec>
        <title>Validity: Free-Living Settings</title>
        <p>There were relatively few studies on wearable device validity in free-living conditions. Fitbit was the only brand with more than 10 studies published for step count validity in free-living conditions, and no brands had more than 10 studies for heart rate or energy expenditure. As a result, we have not shown plots of MPE for free-living conditions.</p>
        <sec>
          <title>Validity for Step Count in Free-Living Settings</title>
          <p>A total of 42 studies (84 comparisons) examined wearable device step count measurements compared with the reference standard criterion measure of accelerometry [<xref ref-type="bibr" rid="ref33">33</xref>,<xref ref-type="bibr" rid="ref45">45</xref>,<xref ref-type="bibr" rid="ref48">48</xref>,<xref ref-type="bibr" rid="ref49">49</xref>, <xref ref-type="bibr" rid="ref56">56</xref>,<xref ref-type="bibr" rid="ref59">59</xref>,<xref ref-type="bibr" rid="ref60">60</xref>,<xref ref-type="bibr" rid="ref64">64</xref>,<xref ref-type="bibr" rid="ref74">74</xref>-<xref ref-type="bibr" rid="ref76">76</xref>,<xref ref-type="bibr" rid="ref89">89</xref>,<xref ref-type="bibr" rid="ref96">96</xref>,<xref ref-type="bibr" rid="ref101">101</xref>,<xref ref-type="bibr" rid="ref106">106</xref>-<xref ref-type="bibr" rid="ref112">112</xref>,<xref ref-type="bibr" rid="ref120">120</xref>,<xref ref-type="bibr" rid="ref131">131</xref>-<xref ref-type="bibr" rid="ref136">136</xref>, <xref ref-type="bibr" rid="ref149">149</xref>,<xref ref-type="bibr" rid="ref154">154</xref>-<xref ref-type="bibr" rid="ref156">156</xref>,<xref ref-type="bibr" rid="ref159">159</xref>,<xref ref-type="bibr" rid="ref167">167</xref>,<xref ref-type="bibr" rid="ref168">168</xref>,<xref ref-type="bibr" rid="ref172">172</xref>-<xref ref-type="bibr" rid="ref175">175</xref>] in free-living settings (<xref ref-type="supplementary-material" rid="app6">Multimedia Appendix 6</xref>). Of these, 28 studies recruited healthy adults (mean age 33.7 years, SD 13.9 years), nine studies recruited adults living with limited mobility/chronic diseases (mean age 60.1 years, SD 11.2 years), four studies recruited healthy children (mean age 12.5 years, SD 2.6 years), and one study recruited children living with cardiac diseases (mean age 13 years, SD 2.2 years). Wearable devices were worn on the lower limb (foot, ankle, and shin), torso, waist/hip, and wrist.</p>
          <p>Group measurement error was reported or calculable for 69 of the 84 comparisons, regardless of the criterion measure. Of these, 42% (n=29) were within ±10% measurement error, 17% (n=12) were below −10% measurement error, and 41% (n=28) were above 10% measurement error. The overall tendency was slight overestimation of step count (mean: 5%, median: 6%). Among the remaining comparisons, 11 of 15 reported MAPE, of which 40% (n=6) were below 10% measurement error and 60% (n=9) were above 10% measurement error.</p>
        </sec>
        <sec>
          <title>Validity for Heart Rate in Free-Living Settings</title>
          <p>Three studies (five comparisons) examined wearable device heart rate compared with the reference standard criterion measure of a Polar brand chest strap in free-living settings [<xref ref-type="bibr" rid="ref75">75</xref>,<xref ref-type="bibr" rid="ref77">77</xref>,<xref ref-type="bibr" rid="ref78">78</xref>]. Of these, one study recruited healthy adults (mean age 25.4 years, SD 3.7 years), one study recruited healthy children (mean age 8 years, SD 1.8 years), and one study recruited adults recovering from stroke (mean age 64.4 years, SD 15 years). All wearable devices were worn on the wrist. Group measurement error was reported or calculable for one of the five comparisons, with the Fitbit Charge HR falling within ±10% measurement error in the study examining healthy children. Three of the four remaining comparisons examined the Fitbit Charge HR in adults and noted underestimation of heart rate that varied depending on activity intensity, but all reported that MAPE values fell within 10% measurement error. Correlation coefficients were strong to very strong in four of the five comparisons and moderate in one comparison examining estimation during high-intensity activity.</p>
        </sec>
        <sec>
          <title>Validity for Energy Expenditure in Free-Living Settings</title>
          <p>Nine studies (22 comparisons) examined energy expenditure in free-living settings compared with the criterion measures of doubly labeled water [<xref ref-type="bibr" rid="ref104">104</xref>] and accelerometry [<xref ref-type="bibr" rid="ref29">29</xref>,<xref ref-type="bibr" rid="ref49">49</xref>,<xref ref-type="bibr" rid="ref79">79</xref>,<xref ref-type="bibr" rid="ref101">101</xref>,<xref ref-type="bibr" rid="ref131">131</xref>,<xref ref-type="bibr" rid="ref172">172</xref>,<xref ref-type="bibr" rid="ref174">174</xref>,<xref ref-type="bibr" rid="ref175">175</xref>]. Eight studies recruited healthy adults (mean age 27.7 years, SD 3.8 years) and one study recruited adults with chronic obstructive pulmonary disease (mean age 66.4 years, SD 7.4 years). Wearable devices were worn on the wrist or waist/hip.</p>
          <p>Group measurement error was reported or calculable for 17 of the 22 comparisons, regardless of the criterion measure. Of these, 18% (n=3) were within ±10% measurement error, 53% (n=9) were below −10% measurement error, and 29% (n=5) were above 10% measurement error. There was an overall tendency to underestimate energy expenditure (mean: −3%, median: −11%). Xiaomi data were not analyzed in a single indirect calorimetry study owing to the lack of data [<xref ref-type="bibr" rid="ref53">53</xref>].</p>
        </sec>
      </sec>
      <sec>
        <title>Reliability</title>
        <p>Nineteen studies (166 comparisons) with sample sizes ranging from 11 [<xref ref-type="bibr" rid="ref94">94</xref>] to 56 [<xref ref-type="bibr" rid="ref151">151</xref>] reported inter- or intradevice reliability for Apple (seven comparisons), Fitbit (92 comparisons), Garmin (22 comparisons), Polar (one comparison), and Withings (44 comparisons). The majority of comparisons (153/166) reported interdevice reliability for step count, heart rate, or energy expenditure. No studies reported intradevice reliability for heart rate or energy expenditure. We have not reported between-brand comparisons for inter- or intradevice reliability owing to the small number of comparisons for each brand.</p>
        <sec>
          <title>Interdevice Reliability for Step Count</title>
          <p>Twelve studies (51 comparisons) with sample sizes ranging from 13 [<xref ref-type="bibr" rid="ref117">117</xref>,<xref ref-type="bibr" rid="ref138">138</xref>] to 56 [<xref ref-type="bibr" rid="ref151">151</xref>] reported on interdevice reliability for step count [<xref ref-type="bibr" rid="ref50">50</xref>,<xref ref-type="bibr" rid="ref58">58</xref>,<xref ref-type="bibr" rid="ref72">72</xref>,<xref ref-type="bibr" rid="ref85">85</xref>,<xref ref-type="bibr" rid="ref94">94</xref>,<xref ref-type="bibr" rid="ref110">110</xref>,<xref ref-type="bibr" rid="ref113">113</xref>,<xref ref-type="bibr" rid="ref116">116</xref>,<xref ref-type="bibr" rid="ref117">117</xref>,<xref ref-type="bibr" rid="ref121">121</xref>,<xref ref-type="bibr" rid="ref125">125</xref>, <xref ref-type="bibr" rid="ref138">138</xref>,<xref ref-type="bibr" rid="ref151">151</xref>,<xref ref-type="bibr" rid="ref161">161</xref>,<xref ref-type="bibr" rid="ref171">171</xref>]. The majority of correlation coefficients for step count interdevice reliability were very strong (n=35), with only a small number (n=3) being reported as strong.</p>
        </sec>
        <sec>
          <title>Intradevice Reliability for Step Count</title>
          <p>Two studies (13 comparisons) reported on intradevice reliability for step count, with sample sizes of 20 [<xref ref-type="bibr" rid="ref82">82</xref>] and 24 [<xref ref-type="bibr" rid="ref150">150</xref>]. Intradevice reliability correlations were very weak (n=1), weak (n=2), moderate (n=5), strong (n=2), and very strong (n=3). The mean correlation coefficient was 0.58.</p>
        </sec>
        <sec>
          <title>Interdevice Reliability for Heart Rate</title>
          <p>Three studies (23 comparisons) examined interdevice reliability for heart rate [<xref ref-type="bibr" rid="ref24">24</xref>,<xref ref-type="bibr" rid="ref26">26</xref>,<xref ref-type="bibr" rid="ref58">58</xref>], with analyzed sample sizes ranging from 13 [<xref ref-type="bibr" rid="ref24">24</xref>] to 21 [<xref ref-type="bibr" rid="ref26">26</xref>]. Apple Watch showed very good interdevice reliability at 5-s epochs during treadmill bouts at 4, 7, and 10 km/h, with reliability increasing and standard typical error decreasing with increasing pace [<xref ref-type="bibr" rid="ref26">26</xref>]. Similar standard typical error levels were seen in maximum heart rate measured during a single incremental maximal oxygen uptake test performed on a treadmill and heart rate taken from the highest 30-s mean heart rate, with somewhat lower correlation coefficients [<xref ref-type="bibr" rid="ref24">24</xref>]. In the examination of interdevice reliability in healthy older adults, Fitbit Charge 2 showed good reliability during treadmill and overground bouts and poor reliability during hand movement tasks such as dusting [<xref ref-type="bibr" rid="ref58">58</xref>]. During the same tasks, Garmin Vivosmart HR+ showed good reliability during all tasks and had narrower limits of agreement than Fitbit.</p>
        </sec>
        <sec>
          <title>Interdevice Reliability for Energy Expenditure</title>
          <p>Five studies (50 comparisons) reported on interdevice reliability [<xref ref-type="bibr" rid="ref85">85</xref>,<xref ref-type="bibr" rid="ref113">113</xref>,<xref ref-type="bibr" rid="ref116">116</xref>,<xref ref-type="bibr" rid="ref117">117</xref>,<xref ref-type="bibr" rid="ref166">166</xref>], with analyzed sample sizes ranging from 13 [<xref ref-type="bibr" rid="ref117">117</xref>] to 29 [<xref ref-type="bibr" rid="ref113">113</xref>]. All five studies recruited healthy adults (mean age 26.3 years, SD 3.9 years). Correlation coefficients were reported for 16 of 50 comparisons. Of these, 13% (n=2) were rated very weak, 6% (n=1) were rated moderate, 6% (n=1) were rated strong, and 75% (n=12) were rated very strong.</p>
        </sec>
      </sec>
    </sec>
    <sec sec-type="discussion">
      <title>Discussion</title>
      <sec>
        <title>Overview</title>
        <p>The purpose of this study was to examine the validity and inter- and intradevice reliability of commercial wearable devices in measuring steps, heart rate, and energy expenditure. Our review focused on both a breadth of devices and reproducibility. Our review included nine brands and 45 devices with the number of comparisons ranging from 201 for the Fitbit Zip to one for the Garmin Forerunner 405CX and the Polar M600. For comparison, two recent reviews from 2017 included two brands and 16 devices [<xref ref-type="bibr" rid="ref13">13</xref>] and seven brands and eight devices [<xref ref-type="bibr" rid="ref79">79</xref>]. A review from 2016 included eight devices [<xref ref-type="bibr" rid="ref32">32</xref>]. Along with this review, we have published our dataset and code to reproduce our findings.</p>
        <p>Our bias assessment showed no apparent bias toward studies of different sample sizes. However, there is a strong overrepresentation of studies with 20 participants. There were some outliers in our findings; however, considering the number of included comparisons, this is to be expected.</p>
      </sec>
      <sec>
        <title>Reliability and Validity</title>
        <p>Criterion validity of commercial wearables varied by study type (controlled or free-living), brand, and device. For step count, our review showed that in controlled laboratory settings, a higher proportion of devices showed accuracy, and this was within a tighter limit of acceptable accuracy compared with free-living conditions. In both controlled and free-living studies, when not correctly estimating steps, devices tended to underestimate values. Validity compared with criteria was the best for Apple Watch and Garmin, while the MPE values for Fitbit, Samsung, and Withings fell within ±3% on average. Within brands, devices appeared to vary, with Fitbit Classic tending to overestimate steps, while Fitbit Charge tending to underestimate steps; however, the variability observed could be attributed to differences in the number of comparisons for each device and in wear locations of the devices. Our findings are consistent with previous reviews [<xref ref-type="bibr" rid="ref178">178</xref>].</p>
        <p>In controlled settings across all devices, heart rate was accurately measured with only a very small tendency for underestimation. Heart rate validity was only sufficiently tested in Apple Watch, Fitbit, and Garmin devices. Heart rate measured by photoplethysmography is only available in relatively new commercial wearable devices. All of the brands measured heart rate to within ±3% on average in controlled settings. There were few studies examining the validity of heart rate measures in free-living conditions, but it appears that Fitbit devices may underestimate heart rate depending on activity intensity. All devices were within acceptable measurement error for heart rate. To our knowledge, this is the first systematic review to examine heart rate validity, and it appears that devices are able to measure heart rate within acceptable limits.</p>
        <p>Energy expenditure estimates varied widely with less than 10% of estimates falling within acceptable limits in controlled settings. In many of the studies, there did appear to be a tendency for systematic over or underestimation. On average, only Fitbit measured energy expenditure to within acceptable limits, but there was wide variation around the estimate. Energy expenditure estimates also varied by model, with the Fitbit Classic underestimating the value considerably and Fitbit Charge HR overestimating the value. We hypothesize that Fitbit may provide the best, though still not acceptable, measure of energy expenditure because the algorithm employs a published equation for estimating resting metabolic rate [<xref ref-type="bibr" rid="ref179">179</xref>]. To our knowledge, the other brands do not publish information about the energy expenditure estimates. There does not appear to be a relationship among more accurate estimates of energy expenditure in devices that include heart rate (<xref ref-type="supplementary-material" rid="app7">Multimedia Appendix 7</xref>).</p>
        <p>Interdevice reliabilities for steps, heart rate, and energy expenditure were all very strong. However, compared with validity studies, there were fewer reliability studies, and we were not able to conduct comparisons between brands or devices owing to small sample sizes. Sufficient data for intradevice reliability was only available for step count. The results showed considerable variability within the same device for step count for Fitbit Charge HR, Fitbit Surge, Fitbit Zip, and Garmin Vivofit, with five, five, one, and two comparisons, respectively.</p>
      </sec>
      <sec>
        <title>Future Research</title>
        <p>Future research in this area should focus on the following three main topics: relevance and age of the devices tested, data acquisition from the devices, and algorithms used by companies. First, relevance of the devices is important. Owing to rapidly developing technology, the majority of the tested devices included in this review are now out of date or discontinued. The nature of the consumer technology market is such that updated product iterations are commissioned even before the original iteration of a device is released. For example, the newest Apple Watch included in the review is the Series 2 watch. The Series 5 watch was released in the fall of 2019. The results are similar for all devices and brands; the Fitbit Charge HR is a popular model for validity and reliability studies, likely because of its moderate price point (approximately US $150) compared with more expensive models (eg, Garmin Fenix 5, approximately US $500). Given the current device specialization, device relevance (eg, swimming or sleep-specific watches), and price difference between devices, continuing to conduct the types of reliability and validity studies reported here will be a challenge. The increasing pace of device release combined with device specialization makes this type of research challenging.</p>
        <p>Second, few studies reported on how data were acquired from the devices. We believe this has implications for the scale of and usability of the data collected. For example, in order to collect data, we infer that some studies counted the steps recorded on the device in short time intervals instead of connecting the device to a platform after recording. Other studies exported and downloaded data from user accounts on the brand website, while others collected data from the brand API. Collecting data from the device API is the best and most scalable method for physical activity researchers when using wearable device data. In order to do so, we must develop interdisciplinary collaborations and open source tools to allow these data to be collected (eg, Open mHealth) [<xref ref-type="bibr" rid="ref180">180</xref>].</p>
        <p>Third, the algorithms used in consumer wearables are constantly changing based on sensor development and technological advances. Companies can update their devices’ firmware and algorithm at any time. When the device is synced, the firmware is updated. Feehan et al discussed the importance of firmware updates in their review [<xref ref-type="bibr" rid="ref13">13</xref>]. While we believe this is important, it is clear that companies must be more open about the algorithms they are using to estimate steps, heart rate, and energy expenditure. Given the continuing release of new devices, firmware and algorithm updates to existing devices, and lack of availability of raw data, we believe researchers may need to shift focus from traditional reliability and validity research to studies that can provide open estimates for physical activity intensities or sleep standardized across devices. These studies will need to use device APIs and machine learning methods in collaboration with interdisciplinary teams in order to move the field forward.</p>
      </sec>
      <sec>
        <title>Limitations</title>
        <p>Over the course of time that it took to complete this review, much has changed with market share, technology, and even research methodologies. Though the market share of companies was a large determining factor of what devices were included in this review, the consumer wearable market is volatile. On November 1, 2019, Google purchased Fitbit for US $2.3 billion, a massive shift for the consumer wearable device market [<xref ref-type="bibr" rid="ref181">181</xref>,<xref ref-type="bibr" rid="ref182">182</xref>]. Further to this limitation is the ever-changing nature of consumer technology. As <xref ref-type="table" rid="table1">Table 1</xref> shows, many of the devices utilized in the studies included in this review are so out of date that they are no longer available on the market. There is some potential for bias when including only English language studies in systematic reviews. However, studies have shown that the effect may be small in general but may be difficult to measure for an individual systematic review [<xref ref-type="bibr" rid="ref183">183</xref>,<xref ref-type="bibr" rid="ref184">184</xref>].</p>
      </sec>
      <sec>
        <title>Conclusion</title>
        <p>This systematic review of 158 publications included assessments of consumer wearable devices from nine brands (Apple Inc, Fitbit, Garmin, Mio, Misfit, Polar, Samsung, Withings, and Xiaomi), with a focus on the reliability and validity of the devices in measuring heart rate, energy expenditure, and step count. This review examined the validity of consumer wearable devices in free-living and laboratory settings and further highlighted results of the inter- and intradevice reliability of the nine consumer wearable brands. Among the studies included, Fitbit was studied the most and Xiaomi and Mio were studied the least. Apple and Samsung had the highest validity for step count, and Apple, Fitbit, and Garmin were accurate nearly 50% of the time. No brand fell within the acceptable accuracy limits for energy expenditure. Interdevice reliabilities for steps, heart rate, and calories were all very strong. Sufficient data for intradevice reliability were only available for step count, and the results showed considerable variability. There was no specific device or brand that involved a complete assessment across all measures, and no specific brand stood out as the “gold standard” in fitness wearables. This review highlights the validity and reliability of readily available wearable devices from brands and serves to guide researchers in making decisions about including them in their research. As new devices and models enter the market, up-to-date documentation can help direct their use in the research setting.</p>
      </sec>
    </sec>
  </body>
  <back>
    <app-group>
      <supplementary-material id="app1">
        <label>Multimedia Appendix 1</label>
        <p>Search strategies.</p>
        <media xlink:href="mhealth_v8i9e18694_app1.pdf" xlink:title="PDF File  (Adobe PDF File), 35 KB"/>
      </supplementary-material>
      <supplementary-material id="app2">
        <label>Multimedia Appendix 2</label>
        <p>Modified Consensus-Based Standards for the Selection of Health Status Measurement Instruments (COSMIN) risk of bias.</p>
        <media xlink:href="mhealth_v8i9e18694_app2.pdf" xlink:title="PDF File  (Adobe PDF File), 131 KB"/>
      </supplementary-material>
      <supplementary-material id="app3">
        <label>Multimedia Appendix 3</label>
        <p>Substudies.</p>
        <media xlink:href="mhealth_v8i9e18694_app3.pdf" xlink:title="PDF File  (Adobe PDF File), 11 KB"/>
      </supplementary-material>
      <supplementary-material id="app4">
        <label>Multimedia Appendix 4</label>
        <p>Study demographics.</p>
        <media xlink:href="mhealth_v8i9e18694_app4.pdf" xlink:title="PDF File  (Adobe PDF File), 254 KB"/>
      </supplementary-material>
      <supplementary-material id="app5">
        <label>Multimedia Appendix 5</label>
        <p>Device models.</p>
        <media xlink:href="mhealth_v8i9e18694_app5.pdf" xlink:title="PDF File  (Adobe PDF File), 197 KB"/>
      </supplementary-material>
      <supplementary-material id="app6">
        <label>Multimedia Appendix 6</label>
        <p>Criterion measures.</p>
        <media xlink:href="mhealth_v8i9e18694_app6.pdf" xlink:title="PDF File  (Adobe PDF File), 193 KB"/>
      </supplementary-material>
      <supplementary-material id="app7">
        <label>Multimedia Appendix 7</label>
        <p>Energy expenditure mean absolute percentage error stratified by heart rate measurement on the device.</p>
        <media xlink:href="mhealth_v8i9e18694_app7.pdf" xlink:title="PDF File  (Adobe PDF File), 5 KB"/>
      </supplementary-material>
    </app-group>
    <glossary>
      <title>Abbreviations</title>
      <def-list>
        <def-item>
          <term id="abb1">API</term>
          <def>
            <p>application program interface</p>
          </def>
        </def-item>
        <def-item>
          <term id="abb2">COSMIN</term>
          <def>
            <p>Consensus-Based Standards for the Selection of Health Status Measurement Instruments</p>
          </def>
        </def-item>
        <def-item>
          <term id="abb3">MAPE</term>
          <def>
            <p>mean absolute percentage error</p>
          </def>
        </def-item>
        <def-item>
          <term id="abb4">MPE</term>
          <def>
            <p>mean percentage error</p>
          </def>
        </def-item>
      </def-list>
    </glossary>
    <ack>
      <p>The authors would like to thank Kristen Romme, a librarian at Memorial University, for writing and executing the searches. They would also like to thank Dr Sherri Tran who assisted with article screening as part of her medical training. Funding for this research was provided by Dr Fuller’s Canada Research Chair (#950-230773).</p>
    </ack>
    <fn-group>
      <fn fn-type="con">
        <p>DF, EC, and JF conceptualized the paper and lead the review. EC, JF, KO, MT, BS, RB, DvH, LS, HL, and NT conducted screening. DF, EC, JF, and KC conducted methodological review and data analysis. All authors contributed to writing of the manuscript and approved the submitted version of the manuscript.</p>
      </fn>
      <fn fn-type="conflict">
        <p>Author JL was employed by Garmin Inc. during the publication process but after completion of the paper.	All other authors have no conflicts to declare.</p>
      </fn>
    </fn-group>
    <ref-list>
      <ref id="ref1">
        <label>1</label>
        <nlm-citation citation-type="web">
          <person-group person-group-type="author">
            <collab>World Health Organization</collab>
          </person-group>
          <article-title>Global recommendations on physical activity for health</article-title>
          <source>World Health Organization</source>
          <access-date>2020-08-17</access-date>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://www.who.int/dietphysicalactivity/factsheet_recommendations/en/">https://www.who.int/dietphysicalactivity/factsheet_recommendations/en/</ext-link>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref2">
        <label>2</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Conn</surname>
              <given-names>VS</given-names>
            </name>
            <name name-style="western">
              <surname>Hafdahl</surname>
              <given-names>AR</given-names>
            </name>
            <name name-style="western">
              <surname>Mehr</surname>
              <given-names>DR</given-names>
            </name>
          </person-group>
          <article-title>Interventions to increase physical activity among healthy adults: meta-analysis of outcomes</article-title>
          <source>Am J Public Health</source>
          <year>2011</year>
          <month>04</month>
          <volume>101</volume>
          <issue>4</issue>
          <fpage>751</fpage>
          <lpage>8</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/21330590"/>
          </comment>
          <pub-id pub-id-type="doi">10.2105/AJPH.2010.194381</pub-id>
          <pub-id pub-id-type="medline">21330590</pub-id>
          <pub-id pub-id-type="pii">AJPH.2010.194381</pub-id>
          <pub-id pub-id-type="pmcid">PMC3052337</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref3">
        <label>3</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Tudor-Locke</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Craig</surname>
              <given-names>CL</given-names>
            </name>
            <name name-style="western">
              <surname>Brown</surname>
              <given-names>WJ</given-names>
            </name>
            <name name-style="western">
              <surname>Clemes</surname>
              <given-names>SA</given-names>
            </name>
            <name name-style="western">
              <surname>De</surname>
              <given-names>CK</given-names>
            </name>
            <name name-style="western">
              <surname>Giles-Corti</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Hatano</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Inoue</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Matsudo</surname>
              <given-names>SM</given-names>
            </name>
            <name name-style="western">
              <surname>Mutrie</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Oppert</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Rowe</surname>
              <given-names>DA</given-names>
            </name>
            <name name-style="western">
              <surname>Schmidt</surname>
              <given-names>MD</given-names>
            </name>
            <name name-style="western">
              <surname>Schofield</surname>
              <given-names>GM</given-names>
            </name>
            <name name-style="western">
              <surname>Spence</surname>
              <given-names>JC</given-names>
            </name>
            <name name-style="western">
              <surname>Teixeira</surname>
              <given-names>PJ</given-names>
            </name>
            <name name-style="western">
              <surname>Tully</surname>
              <given-names>MA</given-names>
            </name>
            <name name-style="western">
              <surname>Blair</surname>
              <given-names>SN</given-names>
            </name>
          </person-group>
          <article-title>How many steps/day are enough? For adults</article-title>
          <source>Int J Behav Nutr Phys Act</source>
          <year>2011</year>
          <month>07</month>
          <day>28</day>
          <volume>8</volume>
          <fpage>79</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://ijbnpa.biomedcentral.com/articles/10.1186/1479-5868-8-79"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/1479-5868-8-79</pub-id>
          <pub-id pub-id-type="medline">21798015</pub-id>
          <pub-id pub-id-type="pii">1479-5868-8-79</pub-id>
          <pub-id pub-id-type="pmcid">PMC3197470</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref4">
        <label>4</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Schneider</surname>
              <given-names>PL</given-names>
            </name>
            <name name-style="western">
              <surname>Bassett</surname>
              <given-names>DR</given-names>
            </name>
            <name name-style="western">
              <surname>Thompson</surname>
              <given-names>DL</given-names>
            </name>
            <name name-style="western">
              <surname>Pronk</surname>
              <given-names>NP</given-names>
            </name>
            <name name-style="western">
              <surname>Bielak</surname>
              <given-names>KM</given-names>
            </name>
          </person-group>
          <article-title>Effects of a 10,000 steps per day goal in overweight adults</article-title>
          <source>Am J Health Promot</source>
          <year>2006</year>
          <volume>21</volume>
          <issue>2</issue>
          <fpage>85</fpage>
          <lpage>9</lpage>
          <pub-id pub-id-type="doi">10.4278/0890-1171-21.2.85</pub-id>
          <pub-id pub-id-type="medline">17152246</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref5">
        <label>5</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Marshall</surname>
              <given-names>SJ</given-names>
            </name>
            <name name-style="western">
              <surname>Levy</surname>
              <given-names>SS</given-names>
            </name>
            <name name-style="western">
              <surname>Tudor-Locke</surname>
              <given-names>CE</given-names>
            </name>
            <name name-style="western">
              <surname>Kolkhorst</surname>
              <given-names>FW</given-names>
            </name>
            <name name-style="western">
              <surname>Wooten</surname>
              <given-names>KM</given-names>
            </name>
            <name name-style="western">
              <surname>Ji</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Macera</surname>
              <given-names>CA</given-names>
            </name>
            <name name-style="western">
              <surname>Ainsworth</surname>
              <given-names>BE</given-names>
            </name>
          </person-group>
          <article-title>Translating physical activity recommendations into a pedometer-based step goal: 3000 steps in 30 minutes</article-title>
          <source>Am J Prev Med</source>
          <year>2009</year>
          <month>05</month>
          <volume>36</volume>
          <issue>5</issue>
          <fpage>410</fpage>
          <lpage>5</lpage>
          <pub-id pub-id-type="doi">10.1016/j.amepre.2009.01.021</pub-id>
          <pub-id pub-id-type="medline">19362695</pub-id>
          <pub-id pub-id-type="pii">S0749-3797(09)00087-7</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref6">
        <label>6</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Case</surname>
              <given-names>MA</given-names>
            </name>
            <name name-style="western">
              <surname>Burwick</surname>
              <given-names>HA</given-names>
            </name>
            <name name-style="western">
              <surname>Volpp</surname>
              <given-names>KG</given-names>
            </name>
            <name name-style="western">
              <surname>Patel</surname>
              <given-names>MS</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of smartphone applications and wearable devices for tracking physical activity data</article-title>
          <source>JAMA</source>
          <year>2015</year>
          <month>02</month>
          <day>10</day>
          <volume>313</volume>
          <issue>6</issue>
          <fpage>625</fpage>
          <lpage>6</lpage>
          <pub-id pub-id-type="doi">10.1001/jama.2014.17841</pub-id>
          <pub-id pub-id-type="medline">25668268</pub-id>
          <pub-id pub-id-type="pii">2108876</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref7">
        <label>7</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Almalki</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Gray</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Sanchez</surname>
              <given-names>FM</given-names>
            </name>
          </person-group>
          <article-title>The use of self-quantification systems for personal health information: big data management activities and prospects</article-title>
          <source>Health Inf Sci Syst</source>
          <year>2015</year>
          <volume>3</volume>
          <issue>Suppl 1 HISA Big Data in Biomedicine and Healthcare 2013 Con</issue>
          <fpage>S1</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/26019809"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/2047-2501-3-S1-S1</pub-id>
          <pub-id pub-id-type="medline">26019809</pub-id>
          <pub-id pub-id-type="pii">2047-2501-3-S1-S1</pub-id>
          <pub-id pub-id-type="pmcid">PMC4437547</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref8">
        <label>8</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Lupton</surname>
              <given-names>D</given-names>
            </name>
          </person-group>
          <article-title>Quantifying the body: monitoring and measuring health in the age of mHealth technologies</article-title>
          <source>Critical Public Health</source>
          <year>2013</year>
          <month>12</month>
          <volume>23</volume>
          <issue>4</issue>
          <fpage>393</fpage>
          <lpage>403</lpage>
          <pub-id pub-id-type="doi">10.1080/09581596.2013.794931</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref9">
        <label>9</label>
        <nlm-citation citation-type="web">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Costello</surname>
              <given-names>K</given-names>
            </name>
          </person-group>
          <article-title>Gartner Says Worldwide Wearable Device Sales to Grow 26 Percent in 2019</article-title>
          <source>Gartner, Inc</source>
          <year>2018</year>
          <access-date>2020-08-17</access-date>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://www.gartner.com/en/newsroom/press-releases/2018-11-29-gartner-says-worldwide-wearable-device-sales-to-grow-">https://www.gartner.com/en/newsroom/press-releases/2018-11-29-gartner-says-worldwide-wearable-device-sales-to-grow-</ext-link>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref10">
        <label>10</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Alley</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Schoeppe</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Guertler</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Jennings</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Duncan</surname>
              <given-names>MJ</given-names>
            </name>
            <name name-style="western">
              <surname>Vandelanotte</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
          <article-title>Interest and preferences for using advanced physical activity tracking devices: results of a national cross-sectional survey</article-title>
          <source>BMJ Open</source>
          <year>2016</year>
          <month>07</month>
          <day>07</day>
          <volume>6</volume>
          <issue>7</issue>
          <fpage>e011243</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://bmjopen.bmj.com/cgi/pmidlookup?view=long&#38;pmid=27388359"/>
          </comment>
          <pub-id pub-id-type="doi">10.1136/bmjopen-2016-011243</pub-id>
          <pub-id pub-id-type="medline">27388359</pub-id>
          <pub-id pub-id-type="pii">bmjopen-2016-011243</pub-id>
          <pub-id pub-id-type="pmcid">PMC4947799</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref11">
        <label>11</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Macridis</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Johnston</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Johnson</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Vallance</surname>
              <given-names>JK</given-names>
            </name>
          </person-group>
          <article-title>Consumer physical activity tracking device ownership and use among a population-based sample of adults</article-title>
          <source>PLoS One</source>
          <year>2018</year>
          <volume>13</volume>
          <issue>1</issue>
          <fpage>e0189298</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0189298"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0189298</pub-id>
          <pub-id pub-id-type="medline">29293532</pub-id>
          <pub-id pub-id-type="pii">PONE-D-17-16001</pub-id>
          <pub-id pub-id-type="pmcid">PMC5749689</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref12">
        <label>12</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Bunn</surname>
              <given-names>JA</given-names>
            </name>
            <name name-style="western">
              <surname>Navalta</surname>
              <given-names>JW</given-names>
            </name>
            <name name-style="western">
              <surname>Fountaine</surname>
              <given-names>CJ</given-names>
            </name>
            <name name-style="western">
              <surname>Reece</surname>
              <given-names>JD</given-names>
            </name>
          </person-group>
          <article-title>Current State of Commercial Wearable Technology in Physical Activity Monitoring 2015-2017</article-title>
          <source>Int J Exerc Sci</source>
          <year>2018</year>
          <volume>11</volume>
          <issue>7</issue>
          <fpage>503</fpage>
          <lpage>515</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/29541338"/>
          </comment>
          <pub-id pub-id-type="medline">29541338</pub-id>
          <pub-id pub-id-type="pmcid">PMC5841672</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref13">
        <label>13</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Feehan</surname>
              <given-names>LM</given-names>
            </name>
            <name name-style="western">
              <surname>Geldman</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Sayre</surname>
              <given-names>EC</given-names>
            </name>
            <name name-style="western">
              <surname>Park</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Ezzat</surname>
              <given-names>AM</given-names>
            </name>
            <name name-style="western">
              <surname>Yoo</surname>
              <given-names>JY</given-names>
            </name>
            <name name-style="western">
              <surname>Hamilton</surname>
              <given-names>CB</given-names>
            </name>
            <name name-style="western">
              <surname>Li</surname>
              <given-names>LC</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of Fitbit Devices: Systematic Review and Narrative Syntheses of Quantitative Data</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2018</year>
          <month>08</month>
          <day>09</day>
          <volume>6</volume>
          <issue>8</issue>
          <fpage>e10527</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://mhealth.jmir.org/2018/8/e10527/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/10527</pub-id>
          <pub-id pub-id-type="medline">30093371</pub-id>
          <pub-id pub-id-type="pii">v6i8e10527</pub-id>
          <pub-id pub-id-type="pmcid">PMC6107736</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref14">
        <label>14</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Moher</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Shamseer</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Clarke</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Ghersi</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Liberati</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Petticrew</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Shekelle</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Stewart</surname>
              <given-names>LA</given-names>
            </name>
            <collab>PRISMA-P Group</collab>
          </person-group>
          <article-title>Preferred reporting items for systematic review and meta-analysis protocols (PRISMA-P) 2015 statement</article-title>
          <source>Syst Rev</source>
          <year>2015</year>
          <month>01</month>
          <day>01</day>
          <volume>4</volume>
          <fpage>1</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://systematicreviewsjournal.biomedcentral.com/articles/10.1186/2046-4053-4-1"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/2046-4053-4-1</pub-id>
          <pub-id pub-id-type="medline">25554246</pub-id>
          <pub-id pub-id-type="pii">2046-4053-4-1</pub-id>
          <pub-id pub-id-type="pmcid">PMC4320440</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref15">
        <label>15</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>McGowan</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Sampson</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Salzwedel</surname>
              <given-names>DM</given-names>
            </name>
            <name name-style="western">
              <surname>Cogo</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Foerster</surname>
              <given-names>V</given-names>
            </name>
            <name name-style="western">
              <surname>Lefebvre</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
          <article-title>PRESS Peer Review of Electronic Search Strategies: 2015 Guideline Statement</article-title>
          <source>J Clin Epidemiol</source>
          <year>2016</year>
          <month>07</month>
          <volume>75</volume>
          <fpage>40</fpage>
          <lpage>6</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://linkinghub.elsevier.com/retrieve/pii/S0895-4356(16)00058-5"/>
          </comment>
          <pub-id pub-id-type="doi">10.1016/j.jclinepi.2016.01.021</pub-id>
          <pub-id pub-id-type="medline">27005575</pub-id>
          <pub-id pub-id-type="pii">S0895-4356(16)00058-5</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref16">
        <label>16</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Higgins</surname>
              <given-names>PA</given-names>
            </name>
            <name name-style="western">
              <surname>Straub</surname>
              <given-names>AJ</given-names>
            </name>
          </person-group>
          <article-title>Understanding the error of our ways: mapping the concepts of validity and reliability</article-title>
          <source>Nurs Outlook</source>
          <year>2006</year>
          <volume>54</volume>
          <issue>1</issue>
          <fpage>23</fpage>
          <lpage>9</lpage>
          <pub-id pub-id-type="doi">10.1016/j.outlook.2004.12.004</pub-id>
          <pub-id pub-id-type="medline">16487776</pub-id>
          <pub-id pub-id-type="pii">S0029-6554(05)00018-7</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref17">
        <label>17</label>
        <nlm-citation citation-type="book">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Swinscow</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Campbell</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <source>Statistics at Square One</source>
          <year>2009</year>
          <publisher-loc>Oxford</publisher-loc>
          <publisher-name>BMJ Books</publisher-name>
        </nlm-citation>
      </ref>
      <ref id="ref18">
        <label>18</label>
        <nlm-citation citation-type="web">
          <article-title>Replication Data for: Systematic Review of the Reliability and Validity of Commercially Available Wearable Devices for Measuring Steps, Energy Expenditure, and Heart Rate</article-title>
          <source>Harvard Dataverse</source>
          <access-date>2020-08-06</access-date>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dataverse.harvard.edu/dataset.xhtml?persistentId=doi:10.7910/DVN/O7GQIM">https://dataverse.harvard.edu/dataset.xhtml?persistentId=doi:10.7910/DVN/O7GQIM</ext-link>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref19">
        <label>19</label>
        <nlm-citation citation-type="web">
          <article-title>wearable_systematic_review</article-title>
          <source>GitHub</source>
          <access-date>2020-08-06</access-date>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://github.com/walkabillylab/wearable_systematic_review">https://github.com/walkabillylab/wearable_systematic_review</ext-link>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref20">
        <label>20</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Bai</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Hibbing</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Mantis</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Welk</surname>
              <given-names>G</given-names>
            </name>
          </person-group>
          <article-title>Comparative evaluation of heart rate-based monitors: Apple Watch vs Fitbit Charge HR</article-title>
          <source>J Sports Sci</source>
          <year>2018</year>
          <month>08</month>
          <volume>36</volume>
          <issue>15</issue>
          <fpage>1734</fpage>
          <lpage>1741</lpage>
          <pub-id pub-id-type="doi">10.1080/02640414.2017.1412235</pub-id>
          <pub-id pub-id-type="medline">29210326</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref21">
        <label>21</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Dooley</surname>
              <given-names>EE</given-names>
            </name>
            <name name-style="western">
              <surname>Golaszewski</surname>
              <given-names>NM</given-names>
            </name>
            <name name-style="western">
              <surname>Bartholomew</surname>
              <given-names>JB</given-names>
            </name>
          </person-group>
          <article-title>Estimating Accuracy at Exercise Intensities: A Comparative Study of Self-Monitoring Heart Rate and Physical Activity Wearable Devices</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2017</year>
          <month>03</month>
          <day>16</day>
          <volume>5</volume>
          <issue>3</issue>
          <fpage>e34</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://mhealth.jmir.org/2017/3/e34/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/mhealth.7043</pub-id>
          <pub-id pub-id-type="medline">28302596</pub-id>
          <pub-id pub-id-type="pii">v5i3e34</pub-id>
          <pub-id pub-id-type="pmcid">PMC5374271</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref22">
        <label>22</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Koshy</surname>
              <given-names>AN</given-names>
            </name>
            <name name-style="western">
              <surname>Sajeev</surname>
              <given-names>JK</given-names>
            </name>
            <name name-style="western">
              <surname>Nerlekar</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Brown</surname>
              <given-names>AJ</given-names>
            </name>
            <name name-style="western">
              <surname>Rajakariar</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Zureik</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Wong</surname>
              <given-names>MC</given-names>
            </name>
            <name name-style="western">
              <surname>Roberts</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Street</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Cooke</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Teh</surname>
              <given-names>AW</given-names>
            </name>
          </person-group>
          <article-title>Smart watches for heart rate assessment in atrial arrhythmias</article-title>
          <source>Int J Cardiol</source>
          <year>2018</year>
          <month>09</month>
          <day>01</day>
          <volume>266</volume>
          <fpage>124</fpage>
          <lpage>127</lpage>
          <pub-id pub-id-type="doi">10.1016/j.ijcard.2018.02.073</pub-id>
          <pub-id pub-id-type="medline">29887428</pub-id>
          <pub-id pub-id-type="pii">S0167-5273(18)30401-7</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref23">
        <label>23</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Thomson</surname>
              <given-names>EA</given-names>
            </name>
            <name name-style="western">
              <surname>Nuss</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Comstock</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Reinwald</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Blake</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Pimentel</surname>
              <given-names>RE</given-names>
            </name>
            <name name-style="western">
              <surname>Tracy</surname>
              <given-names>BL</given-names>
            </name>
            <name name-style="western">
              <surname>Li</surname>
              <given-names>K</given-names>
            </name>
          </person-group>
          <article-title>Heart rate measures from the Apple Watch, Fitbit Charge HR 2, and electrocardiogram across different exercise intensities</article-title>
          <source>J Sports Sci</source>
          <year>2019</year>
          <month>06</month>
          <volume>37</volume>
          <issue>12</issue>
          <fpage>1411</fpage>
          <lpage>1419</lpage>
          <pub-id pub-id-type="doi">10.1080/02640414.2018.1560644</pub-id>
          <pub-id pub-id-type="medline">30657025</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref24">
        <label>24</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Abt</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Bray</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Benson</surname>
              <given-names>AC</given-names>
            </name>
          </person-group>
          <article-title>The validity and inter-device variability of the Apple Watch™ for measuring maximal heart rate</article-title>
          <source>J Sports Sci</source>
          <year>2018</year>
          <month>07</month>
          <volume>36</volume>
          <issue>13</issue>
          <fpage>1447</fpage>
          <lpage>1452</lpage>
          <pub-id pub-id-type="doi">10.1080/02640414.2017.1397282</pub-id>
          <pub-id pub-id-type="medline">29090987</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref25">
        <label>25</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Bunn</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Wells</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Manor</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Webster</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <article-title>Evaluation of Earbud and Wristwatch Heart Rate Monitors during Aerobic and Resistance Training</article-title>
          <source>Int J Exerc Sci</source>
          <year>2019</year>
          <volume>12</volume>
          <issue>4</issue>
          <fpage>374</fpage>
          <lpage>384</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/30899350"/>
          </comment>
          <pub-id pub-id-type="medline">30899350</pub-id>
          <pub-id pub-id-type="pmcid">PMC6413847</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref26">
        <label>26</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Khushhal</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Nichols</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Evans</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Gleadall-Siddall</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Page</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>O'Doherty</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Carroll</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Ingle</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Abt</surname>
              <given-names>G</given-names>
            </name>
          </person-group>
          <article-title>Validity and Reliability of the Apple Watch for Measuring Heart Rate During Exercise</article-title>
          <source>Sports Med Int Open</source>
          <year>2017</year>
          <month>10</month>
          <volume>1</volume>
          <issue>6</issue>
          <fpage>E206</fpage>
          <lpage>E211</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/30539109"/>
          </comment>
          <pub-id pub-id-type="doi">10.1055/s-0043-120195</pub-id>
          <pub-id pub-id-type="medline">30539109</pub-id>
          <pub-id pub-id-type="pii">smio03-2017-0034</pub-id>
          <pub-id pub-id-type="pmcid">PMC6226089</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref27">
        <label>27</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Pope</surname>
              <given-names>ZC</given-names>
            </name>
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>JE</given-names>
            </name>
            <name name-style="western">
              <surname>Zeng</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Gao</surname>
              <given-names>Z</given-names>
            </name>
          </person-group>
          <article-title>Validation of Four Smartwatches in Energy Expenditure and Heart Rate Assessment During Exergaming</article-title>
          <source>Games Health J</source>
          <year>2019</year>
          <month>06</month>
          <volume>8</volume>
          <issue>3</issue>
          <fpage>205</fpage>
          <lpage>212</lpage>
          <pub-id pub-id-type="doi">10.1089/g4h.2018.0087</pub-id>
          <pub-id pub-id-type="medline">31045446</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref28">
        <label>28</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Sañudo</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>De Hoyo</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Muñoz-López</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Perry</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Abt</surname>
              <given-names>G</given-names>
            </name>
          </person-group>
          <article-title>Pilot Study Assessing the Influence of Skin Type on the Heart Rate Measurements Obtained by Photoplethysmography with the Apple Watch</article-title>
          <source>J Med Syst</source>
          <year>2019</year>
          <month>05</month>
          <day>22</day>
          <volume>43</volume>
          <issue>7</issue>
          <fpage>195</fpage>
          <pub-id pub-id-type="doi">10.1007/s10916-019-1325-2</pub-id>
          <pub-id pub-id-type="medline">31119387</pub-id>
          <pub-id pub-id-type="pii">10.1007/s10916-019-1325-2</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref29">
        <label>29</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Chowdhury</surname>
              <given-names>EA</given-names>
            </name>
            <name name-style="western">
              <surname>Western</surname>
              <given-names>MJ</given-names>
            </name>
            <name name-style="western">
              <surname>Nightingale</surname>
              <given-names>TE</given-names>
            </name>
            <name name-style="western">
              <surname>Peacock</surname>
              <given-names>OJ</given-names>
            </name>
            <name name-style="western">
              <surname>Thompson</surname>
              <given-names>D</given-names>
            </name>
          </person-group>
          <article-title>Assessment of laboratory and daily energy expenditure estimates from consumer multi-sensor physical activity monitors</article-title>
          <source>PLoS One</source>
          <year>2017</year>
          <volume>12</volume>
          <issue>2</issue>
          <fpage>e0171720</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0171720"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0171720</pub-id>
          <pub-id pub-id-type="medline">28234979</pub-id>
          <pub-id pub-id-type="pii">PONE-D-16-36931</pub-id>
          <pub-id pub-id-type="pmcid">PMC5325221</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref30">
        <label>30</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Nuss</surname>
              <given-names>KJ</given-names>
            </name>
            <name name-style="western">
              <surname>Thomson</surname>
              <given-names>EA</given-names>
            </name>
            <name name-style="western">
              <surname>Courtney</surname>
              <given-names>JB</given-names>
            </name>
            <name name-style="western">
              <surname>Comstock</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Reinwald</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Blake</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>E.</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Tracy</surname>
              <given-names>BL</given-names>
            </name>
            <name name-style="western">
              <surname>Li</surname>
              <given-names>K</given-names>
            </name>
          </person-group>
          <article-title>Assessment of Accuracy of Overall Energy Expenditure Measurements for the Fitbit Charge HR 2 and Apple Watch</article-title>
          <source>Am J Health Behav</source>
          <year>2019</year>
          <month>05</month>
          <day>01</day>
          <volume>43</volume>
          <issue>3</issue>
          <fpage>498</fpage>
          <lpage>505</lpage>
          <pub-id pub-id-type="doi">10.5993/ajhb.43.3.5</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref31">
        <label>31</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Zhang</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Godin</surname>
              <given-names>SD</given-names>
            </name>
            <name name-style="western">
              <surname>Owens</surname>
              <given-names>MV</given-names>
            </name>
          </person-group>
          <article-title>Measuring the validity and reliability of the Apple Watch as a physical activity monitor</article-title>
          <source>J Sports Med Phys Fitness</source>
          <year>2019</year>
          <month>05</month>
          <volume>59</volume>
          <issue>5</issue>
          <pub-id pub-id-type="doi">10.23736/s0022-4707.18.08339-1</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref32">
        <label>32</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Fokkema</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Kooiman</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Krijnen</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Van</surname>
              <given-names>DS</given-names>
            </name>
            <name name-style="western">
              <surname>De</surname>
              <given-names>GM</given-names>
            </name>
          </person-group>
          <article-title>Reliability and validity of ten consumer activity trackers depend on walking speed</article-title>
          <source>Medicine and Science in Sport and Exercise</source>
          <year>2017</year>
          <volume>49</volume>
          <issue>4</issue>
          <fpage>793</fpage>
          <lpage>800</lpage>
          <pub-id pub-id-type="doi">10.1249/mss.0000000000001146</pub-id>
          <pub-id pub-id-type="pmcid">28319983</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref33">
        <label>33</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Breteler</surname>
              <given-names>MJ</given-names>
            </name>
            <name name-style="western">
              <surname>Janssen</surname>
              <given-names>JH</given-names>
            </name>
            <name name-style="western">
              <surname>Spiering</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Kalkman</surname>
              <given-names>CJ</given-names>
            </name>
            <name name-style="western">
              <surname>van Solinge</surname>
              <given-names>WW</given-names>
            </name>
            <name name-style="western">
              <surname>Dohmen</surname>
              <given-names>DA</given-names>
            </name>
          </person-group>
          <article-title>Measuring Free-Living Physical Activity With Three Commercially Available Activity Monitors for Telemonitoring Purposes: Validation Study</article-title>
          <source>JMIR Form Res</source>
          <year>2019</year>
          <month>04</month>
          <day>24</day>
          <volume>3</volume>
          <issue>2</issue>
          <fpage>e11489</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://formative.jmir.org/2019/2/e11489/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/11489</pub-id>
          <pub-id pub-id-type="medline">31017587</pub-id>
          <pub-id pub-id-type="pii">v3i2e11489</pub-id>
          <pub-id pub-id-type="pmcid">PMC6505372</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref34">
        <label>34</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Gaz</surname>
              <given-names>DV</given-names>
            </name>
            <name name-style="western">
              <surname>Rieck</surname>
              <given-names>TM</given-names>
            </name>
            <name name-style="western">
              <surname>Peterson</surname>
              <given-names>NW</given-names>
            </name>
            <name name-style="western">
              <surname>Ferguson</surname>
              <given-names>JA</given-names>
            </name>
            <name name-style="western">
              <surname>Schroeder</surname>
              <given-names>DR</given-names>
            </name>
            <name name-style="western">
              <surname>Dunfee</surname>
              <given-names>HA</given-names>
            </name>
            <name name-style="western">
              <surname>Henderzahs-Mason</surname>
              <given-names>JM</given-names>
            </name>
            <name name-style="western">
              <surname>Hagen</surname>
              <given-names>PT</given-names>
            </name>
          </person-group>
          <article-title>Determining the Validity and Accuracy of Multiple Activity-Tracking Devices in Controlled and Free-Walking Conditions</article-title>
          <source>Am J Health Promot</source>
          <year>2018</year>
          <month>11</month>
          <volume>32</volume>
          <issue>8</issue>
          <fpage>1671</fpage>
          <lpage>1678</lpage>
          <pub-id pub-id-type="doi">10.1177/0890117118763273</pub-id>
          <pub-id pub-id-type="medline">29558811</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref35">
        <label>35</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Modave</surname>
              <given-names>F</given-names>
            </name>
            <name name-style="western">
              <surname>Guo</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Bian</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Gurka</surname>
              <given-names>MJ</given-names>
            </name>
            <name name-style="western">
              <surname>Parish</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Smith</surname>
              <given-names>MD</given-names>
            </name>
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>AM</given-names>
            </name>
            <name name-style="western">
              <surname>Buford</surname>
              <given-names>TW</given-names>
            </name>
          </person-group>
          <article-title>Mobile Device Accuracy for Step Counting Across Age Groups</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2017</year>
          <month>06</month>
          <day>28</day>
          <volume>5</volume>
          <issue>6</issue>
          <fpage>e88</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://mhealth.jmir.org/2017/6/e88/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/mhealth.7870</pub-id>
          <pub-id pub-id-type="medline">28659255</pub-id>
          <pub-id pub-id-type="pii">v5i6e88</pub-id>
          <pub-id pub-id-type="pmcid">PMC5508112</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref36">
        <label>36</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Sears</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Alvalos</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Lawson</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>McAlister</surname>
              <given-names>I</given-names>
            </name>
            <name name-style="western">
              <surname>Bunn</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>Wrist-Worn Physical Activity Trackers Tend To Underestimate Steps During Walking</article-title>
          <source>International Journal of Exercise Science</source>
          <year>2017</year>
          <volume>10</volume>
          <issue>5</issue>
          <fpage>764</fpage>
          <lpage>773</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://digitalcommons.wku.edu/cgi/viewcontent.cgi?article=2033&#38;context=ijes#:~:text=In%20conclusion%2C%20the%20wrist%2Dworn,best%20of%20the%20devices%20tested"/>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref37">
        <label>37</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Veerabhadrappa</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Moran</surname>
              <given-names>MD</given-names>
            </name>
            <name name-style="western">
              <surname>Renninger</surname>
              <given-names>MD</given-names>
            </name>
            <name name-style="western">
              <surname>Rhudy</surname>
              <given-names>MB</given-names>
            </name>
            <name name-style="western">
              <surname>Dreisbach</surname>
              <given-names>SB</given-names>
            </name>
            <name name-style="western">
              <surname>Gift</surname>
              <given-names>KM</given-names>
            </name>
          </person-group>
          <article-title>Tracking Steps on Apple Watch at Different Walking Speeds</article-title>
          <source>J Gen Intern Med</source>
          <year>2018</year>
          <month>06</month>
          <volume>33</volume>
          <issue>6</issue>
          <fpage>795</fpage>
          <lpage>796</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/29633143"/>
          </comment>
          <pub-id pub-id-type="doi">10.1007/s11606-018-4332-y</pub-id>
          <pub-id pub-id-type="medline">29633143</pub-id>
          <pub-id pub-id-type="pii">10.1007/s11606-018-4332-y</pub-id>
          <pub-id pub-id-type="pmcid">PMC5975150</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref38">
        <label>38</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Wallen</surname>
              <given-names>MP</given-names>
            </name>
            <name name-style="western">
              <surname>Gomersall</surname>
              <given-names>SR</given-names>
            </name>
            <name name-style="western">
              <surname>Keating</surname>
              <given-names>SE</given-names>
            </name>
            <name name-style="western">
              <surname>Wisløff</surname>
              <given-names>U</given-names>
            </name>
            <name name-style="western">
              <surname>Coombes</surname>
              <given-names>JS</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of Heart Rate Watches: Implications for Weight Management</article-title>
          <source>PLoS One</source>
          <year>2016</year>
          <volume>11</volume>
          <issue>5</issue>
          <fpage>e0154420</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0154420"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0154420</pub-id>
          <pub-id pub-id-type="medline">27232714</pub-id>
          <pub-id pub-id-type="pii">PONE-D-15-55600</pub-id>
          <pub-id pub-id-type="pmcid">PMC4883747</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref39">
        <label>39</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Falter</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Budts</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Goetschalckx</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Cornelissen</surname>
              <given-names>V</given-names>
            </name>
            <name name-style="western">
              <surname>Buys</surname>
              <given-names>R</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of Apple Watch Measurements for Heart Rate and Energy Expenditure in Patients With Cardiovascular Disease: Cross-Sectional Study</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2019</year>
          <month>03</month>
          <day>19</day>
          <volume>7</volume>
          <issue>3</issue>
          <fpage>e11889</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://mhealth.jmir.org/2019/3/e11889/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/11889</pub-id>
          <pub-id pub-id-type="medline">30888332</pub-id>
          <pub-id pub-id-type="pii">v7i3e11889</pub-id>
          <pub-id pub-id-type="pmcid">PMC6444219</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref40">
        <label>40</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Gillinov</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Etiwy</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Wang</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Blackburn</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Phelan</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Gillinov</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Houghtaling</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Javadikasgari</surname>
              <given-names>H</given-names>
            </name>
            <name name-style="western">
              <surname>Desai</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <article-title>Variable Accuracy of Wearable Heart Rate Monitors during Aerobic Exercise</article-title>
          <source>Med Sci Sports Exerc</source>
          <year>2017</year>
          <month>08</month>
          <volume>49</volume>
          <issue>8</issue>
          <fpage>1697</fpage>
          <lpage>1703</lpage>
          <pub-id pub-id-type="doi">10.1249/MSS.0000000000001284</pub-id>
          <pub-id pub-id-type="medline">28709155</pub-id>
          <pub-id pub-id-type="pii">00005768-201708000-00022</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref41">
        <label>41</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>H</given-names>
            </name>
            <name name-style="western">
              <surname>Park</surname>
              <given-names>S</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of swimming wearable watches for estimating energy expenditure</article-title>
          <source>International Journal of Applied Sport Science</source>
          <year>2018</year>
          <volume>30</volume>
          <issue>1</issue>
          <fpage>80</fpage>
          <lpage>90</lpage>
          <pub-id pub-id-type="doi">10.24985/ijass.2018.30.1.80</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref42">
        <label>42</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Xie</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Wen</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Liang</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Jia</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Gao</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Lei</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>Evaluating the Validity of Current Mainstream Wearable Devices in Fitness Tracking Under Various Physical Activities: Comparative Study</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2018</year>
          <month>04</month>
          <day>12</day>
          <volume>6</volume>
          <issue>4</issue>
          <fpage>e94</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://mhealth.jmir.org/2018/4/e94/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/mhealth.9754</pub-id>
          <pub-id pub-id-type="medline">29650506</pub-id>
          <pub-id pub-id-type="pii">v6i4e94</pub-id>
          <pub-id pub-id-type="pmcid">PMC5920198</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref43">
        <label>43</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Boudreaux</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Hebert</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Hollander</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Williams</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Cormier</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Naquin</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Gillan</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Gusew</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Kraemer</surname>
              <given-names>R</given-names>
            </name>
          </person-group>
          <article-title>Validity of Wearable Activity Monitors during Cycling and Resistance Exercise</article-title>
          <source>Med Sci Sports Exerc</source>
          <year>2018</year>
          <month>03</month>
          <volume>50</volume>
          <issue>3</issue>
          <fpage>624</fpage>
          <lpage>633</lpage>
          <pub-id pub-id-type="doi">10.1249/MSS.0000000000001471</pub-id>
          <pub-id pub-id-type="medline">29189666</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref44">
        <label>44</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Hwang</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Kim</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Choi</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Cho</surname>
              <given-names>MS</given-names>
            </name>
            <name name-style="western">
              <surname>Nam</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Kim</surname>
              <given-names>Y</given-names>
            </name>
          </person-group>
          <article-title>Assessing Accuracy of Wrist-Worn Wearable Devices in Measurement of Paroxysmal Supraventricular Tachycardia Heart Rate</article-title>
          <source>Korean Circ J</source>
          <year>2019</year>
          <month>05</month>
          <volume>49</volume>
          <issue>5</issue>
          <fpage>437</fpage>
          <lpage>445</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://e-kcj.org/DOIx.php?id=10.4070/kcj.2018.0323"/>
          </comment>
          <pub-id pub-id-type="doi">10.4070/kcj.2018.0323</pub-id>
          <pub-id pub-id-type="medline">30808083</pub-id>
          <pub-id pub-id-type="pii">49.e34</pub-id>
          <pub-id pub-id-type="pmcid">PMC6511533</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref45">
        <label>45</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Brewer</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Swanson</surname>
              <given-names>BT</given-names>
            </name>
            <name name-style="western">
              <surname>Ortiz</surname>
              <given-names>A</given-names>
            </name>
          </person-group>
          <article-title>Validity of Fitbit's active minutes as compared with a research-grade accelerometer and self-reported measures</article-title>
          <source>BMJ Open Sport Exerc Med</source>
          <year>2017</year>
          <volume>3</volume>
          <issue>1</issue>
          <fpage>e000254</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://doi.org/10.1136/bmjsem-2017-000254"/>
          </comment>
          <pub-id pub-id-type="doi">10.1136/bmjsem-2017-000254</pub-id>
          <pub-id pub-id-type="medline">29018543</pub-id>
          <pub-id pub-id-type="pii">bmjsem-2017-000254</pub-id>
          <pub-id pub-id-type="pmcid">PMC5604706</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref46">
        <label>46</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Hergenroeder</surname>
              <given-names>AL</given-names>
            </name>
            <name name-style="western">
              <surname>Barone Gibbs</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Kotlarczyk</surname>
              <given-names>MP</given-names>
            </name>
            <name name-style="western">
              <surname>Perera</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Kowalsky</surname>
              <given-names>RJ</given-names>
            </name>
            <name name-style="western">
              <surname>Brach</surname>
              <given-names>JS</given-names>
            </name>
          </person-group>
          <article-title>Accuracy and Acceptability of Commercial-Grade Physical Activity Monitors in Older Adults</article-title>
          <source>J Aging Phys Act</source>
          <year>2019</year>
          <month>04</month>
          <day>01</day>
          <volume>27</volume>
          <issue>2</issue>
          <fpage>222</fpage>
          <lpage>229</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/30117355"/>
          </comment>
          <pub-id pub-id-type="doi">10.1123/japa.2018-0036</pub-id>
          <pub-id pub-id-type="medline">30117355</pub-id>
          <pub-id pub-id-type="pmcid">PMC7023914</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref47">
        <label>47</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Husted</surname>
              <given-names>HM</given-names>
            </name>
            <name name-style="western">
              <surname>Llewellyn</surname>
              <given-names>TL</given-names>
            </name>
          </person-group>
          <article-title>The Accuracy of Pedometers in Measuring Walking Steps on a Treadmill in College Students</article-title>
          <source>Int J Exerc Sci</source>
          <year>2017</year>
          <volume>10</volume>
          <issue>1</issue>
          <fpage>146</fpage>
          <lpage>153</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/28479955"/>
          </comment>
          <pub-id pub-id-type="medline">28479955</pub-id>
          <pub-id pub-id-type="pmcid">PMC5214549</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref48">
        <label>48</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>DeShaw</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Ellingson</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Bai</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Lansing</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Perez</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Welk</surname>
              <given-names>G</given-names>
            </name>
          </person-group>
          <article-title>Methods for Activity Monitor Validation Studies: An Example With the Fitbit Charge</article-title>
          <source>Journal for the Measurement of Physical Behaviour</source>
          <year>2018</year>
          <volume>1</volume>
          <issue>3</issue>
          <fpage>130</fpage>
          <lpage>135</lpage>
          <pub-id pub-id-type="doi">10.1123/jmpb.2018-0017</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref49">
        <label>49</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Hargens</surname>
              <given-names>TA</given-names>
            </name>
            <name name-style="western">
              <surname>Deyarmin</surname>
              <given-names>KN</given-names>
            </name>
            <name name-style="western">
              <surname>Snyder</surname>
              <given-names>KM</given-names>
            </name>
            <name name-style="western">
              <surname>Mihalik</surname>
              <given-names>AG</given-names>
            </name>
            <name name-style="western">
              <surname>Sharpe</surname>
              <given-names>LE</given-names>
            </name>
          </person-group>
          <article-title>Comparison of wrist-worn and hip-worn activity monitors under free living conditions</article-title>
          <source>J Med Eng Technol</source>
          <year>2017</year>
          <month>04</month>
          <volume>41</volume>
          <issue>3</issue>
          <fpage>200</fpage>
          <lpage>207</lpage>
          <pub-id pub-id-type="doi">10.1080/03091902.2016.1271046</pub-id>
          <pub-id pub-id-type="medline">28078908</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref50">
        <label>50</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Smith</surname>
              <given-names>JD</given-names>
            </name>
            <name name-style="western">
              <surname>Guerra</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Burkholder</surname>
              <given-names>BG</given-names>
            </name>
          </person-group>
          <article-title>The validity and accuracy of wrist-worn activity monitors in lower-limb prosthesis users</article-title>
          <source>Disabil Rehabil</source>
          <year>2019</year>
          <month>04</month>
          <day>12</day>
          <fpage>1</fpage>
          <lpage>7</lpage>
          <pub-id pub-id-type="doi">10.1080/09638288.2019.1587792</pub-id>
          <pub-id pub-id-type="medline">30978125</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref51">
        <label>51</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Toth</surname>
              <given-names>LP</given-names>
            </name>
            <name name-style="western">
              <surname>Park</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Springer</surname>
              <given-names>CM</given-names>
            </name>
            <name name-style="western">
              <surname>Feyerabend</surname>
              <given-names>MD</given-names>
            </name>
            <name name-style="western">
              <surname>Steeves</surname>
              <given-names>JA</given-names>
            </name>
            <name name-style="western">
              <surname>Bassett</surname>
              <given-names>DR</given-names>
            </name>
          </person-group>
          <article-title>Video-Recorded Validation of Wearable Step Counters under Free-living Conditions</article-title>
          <source>Medicine &#38; Science in Sports &#38; Exercise</source>
          <year>2018</year>
          <volume>50</volume>
          <issue>6</issue>
          <fpage>1315</fpage>
          <lpage>1322</lpage>
          <pub-id pub-id-type="doi">10.1249/mss.0000000000001569</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref52">
        <label>52</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Treacy</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Hassett</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Schurr</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Chagpar</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Paul</surname>
              <given-names>SS</given-names>
            </name>
            <name name-style="western">
              <surname>Sherrington</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
          <article-title>Validity of Different Activity Monitors to Count Steps in an Inpatient Rehabilitation Setting</article-title>
          <source>Phys Ther</source>
          <year>2017</year>
          <month>05</month>
          <day>01</day>
          <volume>97</volume>
          <issue>5</issue>
          <fpage>581</fpage>
          <lpage>588</lpage>
          <pub-id pub-id-type="doi">10.1093/ptj/pzx010</pub-id>
          <pub-id pub-id-type="medline">28339904</pub-id>
          <pub-id pub-id-type="pii">3059635</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref53">
        <label>53</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Wahl</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Düking</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Droszez</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Wahl</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Mester</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>Criterion-Validity of Commercially Available Physical Activity Tracker to Estimate Step Count, Covered Distance and Energy Expenditure during Sports Conditions</article-title>
          <source>Front Physiol</source>
          <year>2017</year>
          <volume>8</volume>
          <fpage>725</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://doi.org/10.3389/fphys.2017.00725"/>
          </comment>
          <pub-id pub-id-type="doi">10.3389/fphys.2017.00725</pub-id>
          <pub-id pub-id-type="medline">29018355</pub-id>
          <pub-id pub-id-type="pmcid">PMC5615304</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref54">
        <label>54</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Cadmus-Bertram</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Gangnon</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Wirkus</surname>
              <given-names>EJ</given-names>
            </name>
            <name name-style="western">
              <surname>Thraen-Borowski</surname>
              <given-names>KM</given-names>
            </name>
            <name name-style="western">
              <surname>Gorzelitz-Liebhauser</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>The Accuracy of Heart Rate Monitoring by Some Wrist-Worn Activity Trackers</article-title>
          <source>Ann Intern Med</source>
          <year>2017</year>
          <month>04</month>
          <day>18</day>
          <volume>166</volume>
          <issue>8</issue>
          <fpage>610</fpage>
          <lpage>612</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/28395305"/>
          </comment>
          <pub-id pub-id-type="doi">10.7326/L16-0353</pub-id>
          <pub-id pub-id-type="medline">28395305</pub-id>
          <pub-id pub-id-type="pii">2618339</pub-id>
          <pub-id pub-id-type="pmcid">PMC5564399</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref55">
        <label>55</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Dondzila</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Garner</surname>
              <given-names>D</given-names>
            </name>
          </person-group>
          <article-title>Comparative accuracy of fitness tracking modalities in quantifying energy expenditure</article-title>
          <source>J Med Eng Technol</source>
          <year>2016</year>
          <month>08</month>
          <volume>40</volume>
          <issue>6</issue>
          <fpage>325</fpage>
          <lpage>9</lpage>
          <pub-id pub-id-type="doi">10.1080/03091902.2016.1197978</pub-id>
          <pub-id pub-id-type="medline">27280592</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref56">
        <label>56</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Degroote</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>De Bourdeaudhuij</surname>
              <given-names>I</given-names>
            </name>
            <name name-style="western">
              <surname>Verloigne</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Poppe</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Crombez</surname>
              <given-names>G</given-names>
            </name>
          </person-group>
          <article-title>The Accuracy of Smart Devices for Measuring Physical Activity in Daily Life: Validation Study</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2018</year>
          <month>12</month>
          <day>13</day>
          <volume>6</volume>
          <issue>12</issue>
          <fpage>e10972</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://mhealth.jmir.org/2018/12/e10972/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/10972</pub-id>
          <pub-id pub-id-type="medline">30545810</pub-id>
          <pub-id pub-id-type="pii">v6i12e10972</pub-id>
          <pub-id pub-id-type="pmcid">PMC6315254</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref57">
        <label>57</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Poojary</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Arora</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Britto</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Polen</surname>
              <given-names>Z</given-names>
            </name>
            <name name-style="western">
              <surname>Arena</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Babu</surname>
              <given-names>AS</given-names>
            </name>
          </person-group>
          <article-title>Validity of Mobile-Based Technology vs Direct Observation in Measuring Number of Steps and Distance Walked in 6 Minutes</article-title>
          <source>Mayo Clin Proc</source>
          <year>2018</year>
          <month>12</month>
          <volume>93</volume>
          <issue>12</issue>
          <fpage>1873</fpage>
          <lpage>1874</lpage>
          <pub-id pub-id-type="doi">10.1016/j.mayocp.2018.09.003</pub-id>
          <pub-id pub-id-type="medline">30522597</pub-id>
          <pub-id pub-id-type="pii">S0025-6196(18)30706-7</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref58">
        <label>58</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Tedesco</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Sica</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Ancillao</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Timmons</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Barton</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>O'Flynn</surname>
              <given-names>B</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of consumer-level and research-grade activity trackers in ambulatory settings in older adults</article-title>
          <source>PLoS One</source>
          <year>2019</year>
          <volume>14</volume>
          <issue>5</issue>
          <fpage>e0216891</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0216891"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0216891</pub-id>
          <pub-id pub-id-type="medline">31112585</pub-id>
          <pub-id pub-id-type="pii">PONE-D-19-00617</pub-id>
          <pub-id pub-id-type="pmcid">PMC6529154</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref59">
        <label>59</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Collins</surname>
              <given-names>JE</given-names>
            </name>
            <name name-style="western">
              <surname>Yang</surname>
              <given-names>HY</given-names>
            </name>
            <name name-style="western">
              <surname>Trentadue</surname>
              <given-names>TP</given-names>
            </name>
            <name name-style="western">
              <surname>Gong</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Losina</surname>
              <given-names>E</given-names>
            </name>
          </person-group>
          <article-title>Validation of the Fitbit Charge 2 compared to the ActiGraph GT3X+ in older adults with knee osteoarthritis in free-living conditions</article-title>
          <source>PLoS One</source>
          <year>2019</year>
          <volume>14</volume>
          <issue>1</issue>
          <fpage>e0211231</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0211231"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0211231</pub-id>
          <pub-id pub-id-type="medline">30699159</pub-id>
          <pub-id pub-id-type="pii">PONE-D-18-27517</pub-id>
          <pub-id pub-id-type="pmcid">PMC6353569</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref60">
        <label>60</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Keating</surname>
              <given-names>XD</given-names>
            </name>
            <name name-style="western">
              <surname>Liu</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Liu</surname>
              <given-names>X</given-names>
            </name>
            <name name-style="western">
              <surname>Shangguan</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Guan</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Chen</surname>
              <given-names>L</given-names>
            </name>
          </person-group>
          <article-title>Validity of Fitbit Charge 2 in Controlled College Physical Education Settings</article-title>
          <source>ICHPER-SD Journal of Research</source>
          <year>2012</year>
          <volume>9</volume>
          <issue>2</issue>
          <fpage>28</fpage>
          <lpage>35</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://eric.ed.gov/?id=EJ1197296"/>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref61">
        <label>61</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Benedetto</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Caldato</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Bazzan</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Greenwood</surname>
              <given-names>DC</given-names>
            </name>
            <name name-style="western">
              <surname>Pensabene</surname>
              <given-names>V</given-names>
            </name>
            <name name-style="western">
              <surname>Actis</surname>
              <given-names>P</given-names>
            </name>
          </person-group>
          <article-title>Assessment of the Fitbit Charge 2 for monitoring heart rate</article-title>
          <source>PLoS One</source>
          <year>2018</year>
          <volume>13</volume>
          <issue>2</issue>
          <fpage>e0192691</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0192691"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0192691</pub-id>
          <pub-id pub-id-type="medline">29489850</pub-id>
          <pub-id pub-id-type="pii">PONE-D-17-29021</pub-id>
          <pub-id pub-id-type="pmcid">PMC5831032</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref62">
        <label>62</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Haghayegh</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Khoshnevis</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Smolensky</surname>
              <given-names>MH</given-names>
            </name>
            <name name-style="western">
              <surname>Diller</surname>
              <given-names>KR</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of PurePulse photoplethysmography technology of Fitbit Charge 2 for assessment of heart rate during sleep</article-title>
          <source>Chronobiol Int</source>
          <year>2019</year>
          <month>07</month>
          <volume>36</volume>
          <issue>7</issue>
          <fpage>927</fpage>
          <lpage>933</lpage>
          <pub-id pub-id-type="doi">10.1080/07420528.2019.1596947</pub-id>
          <pub-id pub-id-type="medline">30990098</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref63">
        <label>63</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Reddy</surname>
              <given-names>RK</given-names>
            </name>
            <name name-style="western">
              <surname>Pooni</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Zaharieva</surname>
              <given-names>DP</given-names>
            </name>
            <name name-style="western">
              <surname>Senf</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>El Youssef</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Dassau</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Doyle Iii</surname>
              <given-names>FJ</given-names>
            </name>
            <name name-style="western">
              <surname>Clements</surname>
              <given-names>MA</given-names>
            </name>
            <name name-style="western">
              <surname>Rickels</surname>
              <given-names>MR</given-names>
            </name>
            <name name-style="western">
              <surname>Patton</surname>
              <given-names>SR</given-names>
            </name>
            <name name-style="western">
              <surname>Castle</surname>
              <given-names>JR</given-names>
            </name>
            <name name-style="western">
              <surname>Riddell</surname>
              <given-names>MC</given-names>
            </name>
            <name name-style="western">
              <surname>Jacobs</surname>
              <given-names>PG</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2018</year>
          <month>12</month>
          <day>10</day>
          <volume>6</volume>
          <issue>12</issue>
          <fpage>e10338</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://mhealth.jmir.org/2018/12/e10338/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/10338</pub-id>
          <pub-id pub-id-type="medline">30530451</pub-id>
          <pub-id pub-id-type="pii">v6i12e10338</pub-id>
          <pub-id pub-id-type="pmcid">PMC6305876</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref64">
        <label>64</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Dondzila</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Lewis</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Lopez</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Parker</surname>
              <given-names>T</given-names>
            </name>
          </person-group>
          <article-title>Congruent Accuracy of Wrist-worn Activity Trackers during Controlled and Free-living Conditions</article-title>
          <source>International Journal of Exercise Science</source>
          <year>2018</year>
          <volume>11</volume>
          <issue>7</issue>
          <fpage>575</fpage>
          <lpage>584</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://pdfs.semanticscholar.org/5ace/6153feeec68af823bc618bab960de413fcd5.pdf"/>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref65">
        <label>65</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Lamont</surname>
              <given-names>RM</given-names>
            </name>
            <name name-style="western">
              <surname>Daniel</surname>
              <given-names>HL</given-names>
            </name>
            <name name-style="western">
              <surname>Payne</surname>
              <given-names>CL</given-names>
            </name>
            <name name-style="western">
              <surname>Brauer</surname>
              <given-names>SG</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of wearable physical activity trackers in people with Parkinson's disease</article-title>
          <source>Gait Posture</source>
          <year>2018</year>
          <month>06</month>
          <volume>63</volume>
          <fpage>104</fpage>
          <lpage>108</lpage>
          <pub-id pub-id-type="doi">10.1016/j.gaitpost.2018.04.034</pub-id>
          <pub-id pub-id-type="medline">29729611</pub-id>
          <pub-id pub-id-type="pii">S0966-6362(18)30436-3</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref66">
        <label>66</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Montoye</surname>
              <given-names>AH</given-names>
            </name>
            <name name-style="western">
              <surname>Mitrzyk</surname>
              <given-names>JR</given-names>
            </name>
            <name name-style="western">
              <surname>Molesky</surname>
              <given-names>MJ</given-names>
            </name>
          </person-group>
          <article-title>Comparative Accuracy of a Wrist-Worn Activity Tracker and a Smart Shirt for Physical Activity Assessment</article-title>
          <source>Measurement in Physical Education and Exercise Science</source>
          <year>2017</year>
          <month>06</month>
          <day>08</day>
          <volume>21</volume>
          <issue>4</issue>
          <fpage>201</fpage>
          <lpage>211</lpage>
          <pub-id pub-id-type="doi">10.1080/1091367X.2017.1331166</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref67">
        <label>67</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Jo</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Lewis</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Directo</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Kim</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Dolezal</surname>
              <given-names>B</given-names>
            </name>
          </person-group>
          <article-title>Validation of Biofeedback Wearables for Photoplethysmographic Heart Rate Tracking</article-title>
          <source>J Sports Sci Med</source>
          <year>2016</year>
          <month>09</month>
          <volume>15</volume>
          <issue>3</issue>
          <fpage>540</fpage>
          <lpage>547</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/27803634"/>
          </comment>
          <pub-id pub-id-type="medline">27803634</pub-id>
          <pub-id pub-id-type="pmcid">PMC4974868</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref68">
        <label>68</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Kroll</surname>
              <given-names>RR</given-names>
            </name>
            <name name-style="western">
              <surname>Boyd</surname>
              <given-names>JG</given-names>
            </name>
            <name name-style="western">
              <surname>Maslove</surname>
              <given-names>DM</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of a Wrist-Worn Wearable Device for Monitoring Heart Rates in Hospital Inpatients: A Prospective Observational Study</article-title>
          <source>J Med Internet Res</source>
          <year>2016</year>
          <month>09</month>
          <day>20</day>
          <volume>18</volume>
          <issue>9</issue>
          <fpage>e253</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://www.jmir.org/2016/9/e253/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/jmir.6025</pub-id>
          <pub-id pub-id-type="medline">27651304</pub-id>
          <pub-id pub-id-type="pii">v18i9e253</pub-id>
          <pub-id pub-id-type="pmcid">PMC5050383</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref69">
        <label>69</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Pelizzo</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Guddo</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Puglisi</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>De Silvestri</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Comparato</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Valenza</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Bordonaro</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Calcaterra</surname>
              <given-names>V</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of a Wrist-Worn Heart Rate Sensing Device during Elective Pediatric Surgical Procedures</article-title>
          <source>Children (Basel)</source>
          <year>2018</year>
          <month>03</month>
          <day>08</day>
          <volume>5</volume>
          <issue>3</issue>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://www.mdpi.com/resolver?pii=children5030038"/>
          </comment>
          <pub-id pub-id-type="doi">10.3390/children5030038</pub-id>
          <pub-id pub-id-type="medline">29518020</pub-id>
          <pub-id pub-id-type="pii">children5030038</pub-id>
          <pub-id pub-id-type="pmcid">PMC5867497</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref70">
        <label>70</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Powierza</surname>
              <given-names>CS</given-names>
            </name>
            <name name-style="western">
              <surname>Clark</surname>
              <given-names>MD</given-names>
            </name>
            <name name-style="western">
              <surname>Hughes</surname>
              <given-names>JM</given-names>
            </name>
            <name name-style="western">
              <surname>Carneiro</surname>
              <given-names>KA</given-names>
            </name>
            <name name-style="western">
              <surname>Mihalik</surname>
              <given-names>JP</given-names>
            </name>
          </person-group>
          <article-title>Validation Of A Self-monitoring Tool For Use In Post-concussion Syndrome Therapy</article-title>
          <source>Medicine &#38; Science in Sports &#38; Exercise</source>
          <year>2016</year>
          <volume>48</volume>
          <fpage>798</fpage>
          <pub-id pub-id-type="doi">10.1249/01.mss.0000487390.22076.d1</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref71">
        <label>71</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Stahl</surname>
              <given-names>SE</given-names>
            </name>
            <name name-style="western">
              <surname>An</surname>
              <given-names>H</given-names>
            </name>
            <name name-style="western">
              <surname>Dinkel</surname>
              <given-names>DM</given-names>
            </name>
            <name name-style="western">
              <surname>Noble</surname>
              <given-names>JM</given-names>
            </name>
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough?</article-title>
          <source>BMJ Open Sport Exerc Med</source>
          <year>2016</year>
          <volume>2</volume>
          <issue>1</issue>
          <fpage>e000106</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://doi.org/10.1136/bmjsem-2015-000106"/>
          </comment>
          <pub-id pub-id-type="doi">10.1136/bmjsem-2015-000106</pub-id>
          <pub-id pub-id-type="medline">27900173</pub-id>
          <pub-id pub-id-type="pii">bmjsem-2015-000106</pub-id>
          <pub-id pub-id-type="pmcid">PMC5117066</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref72">
        <label>72</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Burton</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Hill</surname>
              <given-names>KD</given-names>
            </name>
            <name name-style="western">
              <surname>Lautenschlager</surname>
              <given-names>NT</given-names>
            </name>
            <name name-style="western">
              <surname>Thøgersen-Ntoumani</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Lewin</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Boyle</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Howie</surname>
              <given-names>E</given-names>
            </name>
          </person-group>
          <article-title>Reliability and validity of two fitness tracker devices in the laboratory and home environment for older community-dwelling people</article-title>
          <source>BMC Geriatr</source>
          <year>2018</year>
          <month>05</month>
          <day>03</day>
          <volume>18</volume>
          <issue>1</issue>
          <fpage>103</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://bmcgeriatr.biomedcentral.com/articles/10.1186/s12877-018-0793-4"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/s12877-018-0793-4</pub-id>
          <pub-id pub-id-type="medline">29724191</pub-id>
          <pub-id pub-id-type="pii">10.1186/s12877-018-0793-4</pub-id>
          <pub-id pub-id-type="pmcid">PMC5934836</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref73">
        <label>73</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Morris</surname>
              <given-names>CE</given-names>
            </name>
            <name name-style="western">
              <surname>Wessel</surname>
              <given-names>PA</given-names>
            </name>
            <name name-style="western">
              <surname>Tinius</surname>
              <given-names>RA</given-names>
            </name>
            <name name-style="western">
              <surname>Schafer</surname>
              <given-names>MA</given-names>
            </name>
            <name name-style="western">
              <surname>Maples</surname>
              <given-names>JM</given-names>
            </name>
          </person-group>
          <article-title>Validity of Activity Trackers in Estimating Energy Expenditure During High-Intensity Functional Training</article-title>
          <source>Res Q Exerc Sport</source>
          <year>2019</year>
          <month>09</month>
          <volume>90</volume>
          <issue>3</issue>
          <fpage>377</fpage>
          <lpage>384</lpage>
          <pub-id pub-id-type="doi">10.1080/02701367.2019.1603989</pub-id>
          <pub-id pub-id-type="medline">31046630</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref74">
        <label>74</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Farina</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Lowry</surname>
              <given-names>RG</given-names>
            </name>
          </person-group>
          <article-title>The Validity of Consumer-Level Activity Monitors in Healthy Older Adults in Free-Living Conditions</article-title>
          <source>J Aging Phys Act</source>
          <year>2018</year>
          <month>01</month>
          <day>01</day>
          <volume>26</volume>
          <issue>1</issue>
          <fpage>128</fpage>
          <lpage>135</lpage>
          <pub-id pub-id-type="doi">10.1123/japa.2016-0344</pub-id>
          <pub-id pub-id-type="medline">28595019</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref75">
        <label>75</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Rozanski</surname>
              <given-names>GM</given-names>
            </name>
            <name name-style="western">
              <surname>Aqui</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Sivakumaran</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Mansfield</surname>
              <given-names>A</given-names>
            </name>
          </person-group>
          <article-title>Consumer Wearable Devices for Activity Monitoring Among Individuals After a Stroke: A Prospective Comparison</article-title>
          <source>JMIR Cardio</source>
          <year>2018</year>
          <month>01</month>
          <day>04</day>
          <volume>2</volume>
          <issue>1</issue>
          <fpage>e1</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://cardio.jmir.org/2018/1/e1/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/cardio.8199</pub-id>
          <pub-id pub-id-type="medline">31758760</pub-id>
          <pub-id pub-id-type="pii">v2i1e1</pub-id>
          <pub-id pub-id-type="pmcid">PMC6834221</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref76">
        <label>76</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Voss</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Gardner</surname>
              <given-names>RF</given-names>
            </name>
            <name name-style="western">
              <surname>Dean</surname>
              <given-names>PH</given-names>
            </name>
            <name name-style="western">
              <surname>Harris</surname>
              <given-names>KC</given-names>
            </name>
          </person-group>
          <article-title>Validity of Commercial Activity Trackers in Children With Congenital Heart Disease</article-title>
          <source>Can J Cardiol</source>
          <year>2017</year>
          <month>06</month>
          <volume>33</volume>
          <issue>6</issue>
          <fpage>799</fpage>
          <lpage>805</lpage>
          <pub-id pub-id-type="doi">10.1016/j.cjca.2016.11.024</pub-id>
          <pub-id pub-id-type="medline">28347581</pub-id>
          <pub-id pub-id-type="pii">S0828-282X(16)31147-3</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref77">
        <label>77</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Brazendale</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Decker</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Hunt</surname>
              <given-names>ET</given-names>
            </name>
            <name name-style="western">
              <surname>Perry</surname>
              <given-names>MW</given-names>
            </name>
            <name name-style="western">
              <surname>Brazendale</surname>
              <given-names>AB</given-names>
            </name>
            <name name-style="western">
              <surname>Weaver</surname>
              <given-names>RG</given-names>
            </name>
            <name name-style="western">
              <surname>Beets</surname>
              <given-names>MW</given-names>
            </name>
          </person-group>
          <article-title>Validity and Wearability of Consumer-based Fitness Trackers in Free-living Children</article-title>
          <source>Int J Exerc Sci</source>
          <year>2019</year>
          <volume>12</volume>
          <issue>5</issue>
          <fpage>471</fpage>
          <lpage>482</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/30899354"/>
          </comment>
          <pub-id pub-id-type="medline">30899354</pub-id>
          <pub-id pub-id-type="pmcid">PMC6413843</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref78">
        <label>78</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Gorny</surname>
              <given-names>AW</given-names>
            </name>
            <name name-style="western">
              <surname>Liew</surname>
              <given-names>SJ</given-names>
            </name>
            <name name-style="western">
              <surname>Tan</surname>
              <given-names>CS</given-names>
            </name>
            <name name-style="western">
              <surname>Müller-Riemenschneider</surname>
              <given-names>F</given-names>
            </name>
          </person-group>
          <article-title>Fitbit Charge HR Wireless Heart Rate Monitor: Validation Study Conducted Under Free-Living Conditions</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2017</year>
          <month>10</month>
          <day>20</day>
          <volume>5</volume>
          <issue>10</issue>
          <fpage>e157</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://mhealth.jmir.org/2017/10/e157/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/mhealth.8233</pub-id>
          <pub-id pub-id-type="medline">29055881</pub-id>
          <pub-id pub-id-type="pii">v5i10e157</pub-id>
          <pub-id pub-id-type="pmcid">PMC5670315</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref79">
        <label>79</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Brooke</surname>
              <given-names>SM</given-names>
            </name>
            <name name-style="western">
              <surname>An</surname>
              <given-names>H</given-names>
            </name>
            <name name-style="western">
              <surname>Kang</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Noble</surname>
              <given-names>JM</given-names>
            </name>
            <name name-style="western">
              <surname>Berg</surname>
              <given-names>KE</given-names>
            </name>
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>Concurrent Validity of Wearable Activity Trackers Under Free-Living Conditions</article-title>
          <source>Journal of Strength and Conditioning Research</source>
          <year>2017</year>
          <volume>31</volume>
          <issue>4</issue>
          <fpage>1097</fpage>
          <lpage>1106</lpage>
          <pub-id pub-id-type="doi">10.1519/jsc.0000000000001571</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref80">
        <label>80</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Chow</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Thom</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Wewege</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Ward</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Parmenter</surname>
              <given-names>B</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of step count measured by physical activity monitors: The effect of gait speed and anatomical placement site</article-title>
          <source>Gait Posture</source>
          <year>2017</year>
          <month>09</month>
          <volume>57</volume>
          <fpage>199</fpage>
          <lpage>203</lpage>
          <pub-id pub-id-type="doi">10.1016/j.gaitpost.2017.06.012</pub-id>
          <pub-id pub-id-type="medline">28666177</pub-id>
          <pub-id pub-id-type="pii">S0966-6362(17)30235-7</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref81">
        <label>81</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Tam</surname>
              <given-names>KM</given-names>
            </name>
            <name name-style="western">
              <surname>Cheung</surname>
              <given-names>SY</given-names>
            </name>
          </person-group>
          <article-title>Validation of Electronic Activity Monitor Devices During Treadmill Walking</article-title>
          <source>Telemed J E Health</source>
          <year>2018</year>
          <month>10</month>
          <volume>24</volume>
          <issue>10</issue>
          <fpage>782</fpage>
          <lpage>789</lpage>
          <pub-id pub-id-type="doi">10.1089/tmj.2017.0263</pub-id>
          <pub-id pub-id-type="medline">29364065</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref82">
        <label>82</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Tophøj</surname>
              <given-names>KH</given-names>
            </name>
            <name name-style="western">
              <surname>Petersen</surname>
              <given-names>MG</given-names>
            </name>
            <name name-style="western">
              <surname>Sæbye</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Baad-Hansen</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Wagner</surname>
              <given-names>S</given-names>
            </name>
          </person-group>
          <article-title>Validity and Reliability Evaluation of Four Commercial Activity Trackers' Step Counting Performance</article-title>
          <source>Telemed J E Health</source>
          <year>2018</year>
          <month>09</month>
          <volume>24</volume>
          <issue>9</issue>
          <fpage>669</fpage>
          <lpage>677</lpage>
          <pub-id pub-id-type="doi">10.1089/tmj.2017.0264</pub-id>
          <pub-id pub-id-type="medline">29303680</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref83">
        <label>83</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Fortune</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Lugade</surname>
              <given-names>V</given-names>
            </name>
            <name name-style="western">
              <surname>Morrow</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Kaufman</surname>
              <given-names>K</given-names>
            </name>
          </person-group>
          <article-title>Validity of using tri-axial accelerometers to measure human movement - Part II: Step counts at a wide range of gait velocities</article-title>
          <source>Med Eng Phys</source>
          <year>2014</year>
          <month>06</month>
          <volume>36</volume>
          <issue>6</issue>
          <fpage>659</fpage>
          <lpage>69</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/24656871"/>
          </comment>
          <pub-id pub-id-type="doi">10.1016/j.medengphy.2014.02.006</pub-id>
          <pub-id pub-id-type="medline">24656871</pub-id>
          <pub-id pub-id-type="pii">S1350-4533(14)00028-9</pub-id>
          <pub-id pub-id-type="pmcid">PMC4030415</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref84">
        <label>84</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Phillips</surname>
              <given-names>LJ</given-names>
            </name>
            <name name-style="western">
              <surname>Petroski</surname>
              <given-names>GF</given-names>
            </name>
            <name name-style="western">
              <surname>Markis</surname>
              <given-names>NE</given-names>
            </name>
          </person-group>
          <article-title>A Comparison of Accelerometer Accuracy in Older Adults</article-title>
          <source>Res Gerontol Nurs</source>
          <year>2015</year>
          <volume>8</volume>
          <issue>5</issue>
          <fpage>213</fpage>
          <lpage>9</lpage>
          <pub-id pub-id-type="doi">10.3928/19404921-20150429-03</pub-id>
          <pub-id pub-id-type="medline">25942386</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref85">
        <label>85</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Adam Noah</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Spierer</surname>
              <given-names>DK</given-names>
            </name>
            <name name-style="western">
              <surname>Gu</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Bronner</surname>
              <given-names>S</given-names>
            </name>
          </person-group>
          <article-title>Comparison of steps and energy expenditure assessment in adults of Fitbit Tracker and Ultra to the Actical and indirect calorimetry</article-title>
          <source>J Med Eng Technol</source>
          <year>2013</year>
          <month>10</month>
          <volume>37</volume>
          <issue>7</issue>
          <fpage>456</fpage>
          <lpage>62</lpage>
          <pub-id pub-id-type="doi">10.3109/03091902.2013.831135</pub-id>
          <pub-id pub-id-type="medline">24007317</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref86">
        <label>86</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Dannecker</surname>
              <given-names>KL</given-names>
            </name>
            <name name-style="western">
              <surname>Sazonova</surname>
              <given-names>NA</given-names>
            </name>
            <name name-style="western">
              <surname>Melanson</surname>
              <given-names>EL</given-names>
            </name>
            <name name-style="western">
              <surname>Sazonov</surname>
              <given-names>ES</given-names>
            </name>
            <name name-style="western">
              <surname>Browning</surname>
              <given-names>RC</given-names>
            </name>
          </person-group>
          <article-title>A comparison of energy expenditure estimation of several physical activity monitors</article-title>
          <source>Med Sci Sports Exerc</source>
          <year>2013</year>
          <month>11</month>
          <volume>45</volume>
          <issue>11</issue>
          <fpage>2105</fpage>
          <lpage>12</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/23669877"/>
          </comment>
          <pub-id pub-id-type="doi">10.1249/MSS.0b013e318299d2eb</pub-id>
          <pub-id pub-id-type="medline">23669877</pub-id>
          <pub-id pub-id-type="pmcid">PMC3800491</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref87">
        <label>87</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Sasaki</surname>
              <given-names>JE</given-names>
            </name>
            <name name-style="western">
              <surname>Hickey</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Mavilia</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Tedesco</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>John</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Kozey Keadle</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Freedson</surname>
              <given-names>PS</given-names>
            </name>
          </person-group>
          <article-title>Validation of the Fitbit wireless activity tracker for prediction of energy expenditure</article-title>
          <source>J Phys Act Health</source>
          <year>2015</year>
          <month>02</month>
          <volume>12</volume>
          <issue>2</issue>
          <fpage>149</fpage>
          <lpage>54</lpage>
          <pub-id pub-id-type="doi">10.1123/jpah.2012-0495</pub-id>
          <pub-id pub-id-type="medline">24770438</pub-id>
          <pub-id pub-id-type="pii">2012-0495</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref88">
        <label>88</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Alinia</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Cain</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Fallahzadeh</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Shahrokni</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Cook</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Ghasemzadeh</surname>
              <given-names>H</given-names>
            </name>
          </person-group>
          <article-title>How Accurate Is Your Activity Tracker? A Comparative Study of Step Counts in Low-Intensity Physical Activities</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2017</year>
          <month>08</month>
          <day>11</day>
          <volume>5</volume>
          <issue>8</issue>
          <fpage>e106</fpage>
          <pub-id pub-id-type="doi">10.2196/mhealth.6321</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref89">
        <label>89</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>An</surname>
              <given-names>H</given-names>
            </name>
            <name name-style="western">
              <surname>Jones</surname>
              <given-names>GC</given-names>
            </name>
            <name name-style="western">
              <surname>Kang</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Welk</surname>
              <given-names>GJ</given-names>
            </name>
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>How valid are wearable physical activity trackers for measuring steps?</article-title>
          <source>Eur J Sport Sci</source>
          <year>2017</year>
          <month>04</month>
          <volume>17</volume>
          <issue>3</issue>
          <fpage>360</fpage>
          <lpage>368</lpage>
          <pub-id pub-id-type="doi">10.1080/17461391.2016.1255261</pub-id>
          <pub-id pub-id-type="medline">27912681</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref90">
        <label>90</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Floegel</surname>
              <given-names>TA</given-names>
            </name>
            <name name-style="western">
              <surname>Florez-Pregonero</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Hekler</surname>
              <given-names>EB</given-names>
            </name>
            <name name-style="western">
              <surname>Buman</surname>
              <given-names>MP</given-names>
            </name>
          </person-group>
          <article-title>Validation of Consumer-Based Hip and Wrist Activity Monitors in Older Adults With Varied Ambulatory Abilities</article-title>
          <source>J Gerontol A Biol Sci Med Sci</source>
          <year>2017</year>
          <month>02</month>
          <volume>72</volume>
          <issue>2</issue>
          <fpage>229</fpage>
          <lpage>236</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/27257217"/>
          </comment>
          <pub-id pub-id-type="doi">10.1093/gerona/glw098</pub-id>
          <pub-id pub-id-type="medline">27257217</pub-id>
          <pub-id pub-id-type="pii">glw098</pub-id>
          <pub-id pub-id-type="pmcid">PMC6082588</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref91">
        <label>91</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Gilmore</surname>
              <given-names>SJ</given-names>
            </name>
            <name name-style="western">
              <surname>Davidson</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Hahne</surname>
              <given-names>AJ</given-names>
            </name>
            <name name-style="western">
              <surname>McClelland</surname>
              <given-names>JA</given-names>
            </name>
          </person-group>
          <article-title>The validity of using activity monitors to detect step count after lumbar fusion surgery</article-title>
          <source>Disabil Rehabil</source>
          <year>2020</year>
          <month>03</month>
          <volume>42</volume>
          <issue>6</issue>
          <fpage>863</fpage>
          <lpage>868</lpage>
          <pub-id pub-id-type="doi">10.1080/09638288.2018.1509140</pub-id>
          <pub-id pub-id-type="medline">30326754</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref92">
        <label>92</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Huang</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Xu</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Yu</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Shull</surname>
              <given-names>PB</given-names>
            </name>
          </person-group>
          <article-title>Validity of FitBit, Jawbone UP, Nike+ and other wearable devices for level and stair walking</article-title>
          <source>Gait Posture</source>
          <year>2016</year>
          <month>07</month>
          <volume>48</volume>
          <fpage>36</fpage>
          <lpage>41</lpage>
          <pub-id pub-id-type="doi">10.1016/j.gaitpost.2016.04.025</pub-id>
          <pub-id pub-id-type="medline">27477705</pub-id>
          <pub-id pub-id-type="pii">S0966-6362(16)30042-X</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref93">
        <label>93</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Imboden</surname>
              <given-names>MT</given-names>
            </name>
            <name name-style="western">
              <surname>Nelson</surname>
              <given-names>MB</given-names>
            </name>
            <name name-style="western">
              <surname>Kaminsky</surname>
              <given-names>LA</given-names>
            </name>
            <name name-style="western">
              <surname>Montoye</surname>
              <given-names>AH</given-names>
            </name>
          </person-group>
          <article-title>Comparison of four Fitbit and Jawbone activity monitors with a research-grade ActiGraph accelerometer for estimating physical activity and energy expenditure</article-title>
          <source>Br J Sports Med</source>
          <year>2018</year>
          <month>07</month>
          <volume>52</volume>
          <issue>13</issue>
          <fpage>844</fpage>
          <lpage>850</lpage>
          <pub-id pub-id-type="doi">10.1136/bjsports-2016-096990</pub-id>
          <pub-id pub-id-type="medline">28483930</pub-id>
          <pub-id pub-id-type="pii">bjsports-2016-096990</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref94">
        <label>94</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Jones</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Crossley</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Dascombe</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Hart</surname>
              <given-names>HF</given-names>
            </name>
            <name name-style="western">
              <surname>Kemp</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>Validity and Reliability of the Fitbit FlexTM and Actigraph Gt3X+ At Jogging and Running Speeds</article-title>
          <source>Intl J Sports Phys Ther</source>
          <year>2018</year>
          <month>8</month>
          <volume>13</volume>
          <issue>5</issue>
          <fpage>860</fpage>
          <lpage>870</lpage>
          <pub-id pub-id-type="doi">10.26603/ijspt20180860</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref95">
        <label>95</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Kendall</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Bellovary</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Gothe</surname>
              <given-names>NP</given-names>
            </name>
          </person-group>
          <article-title>Validity of wearable activity monitors for tracking steps and estimating energy expenditure during a graded maximal treadmill test</article-title>
          <source>J Sports Sci</source>
          <year>2019</year>
          <month>01</month>
          <volume>37</volume>
          <issue>1</issue>
          <fpage>42</fpage>
          <lpage>49</lpage>
          <pub-id pub-id-type="doi">10.1080/02640414.2018.1481723</pub-id>
          <pub-id pub-id-type="medline">29863968</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref96">
        <label>96</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Kooiman</surname>
              <given-names>TJ</given-names>
            </name>
            <name name-style="western">
              <surname>Dontje</surname>
              <given-names>ML</given-names>
            </name>
            <name name-style="western">
              <surname>Sprenger</surname>
              <given-names>SR</given-names>
            </name>
            <name name-style="western">
              <surname>Krijnen</surname>
              <given-names>WP</given-names>
            </name>
            <name name-style="western">
              <surname>van der Schans</surname>
              <given-names>CP</given-names>
            </name>
            <name name-style="western">
              <surname>de Groot</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <article-title>Reliability and validity of ten consumer activity trackers</article-title>
          <source>BMC Sports Sci Med Rehabil</source>
          <year>2015</year>
          <volume>7</volume>
          <fpage>24</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://bmcsportsscimedrehabil.biomedcentral.com/articles/10.1186/s13102-015-0018-5"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/s13102-015-0018-5</pub-id>
          <pub-id pub-id-type="medline">26464801</pub-id>
          <pub-id pub-id-type="pii">18</pub-id>
          <pub-id pub-id-type="pmcid">PMC4603296</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref97">
        <label>97</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Montes</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Young</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Tandy</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Navalta</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>Fitbit Flex Energy Expenditure and Step Count Evaluation</article-title>
          <source>Journal Exercise Physiology Online</source>
          <year>2017</year>
          <volume>20</volume>
          <issue>5</issue>
          <fpage>134</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://pdfs.semanticscholar.org/ae63/86ac2f17b3fd7d0aa235477ece501954829a.pdf"/>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref98">
        <label>98</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Sala</surname>
              <given-names>DA</given-names>
            </name>
            <name name-style="western">
              <surname>Grissom</surname>
              <given-names>HE</given-names>
            </name>
            <name name-style="western">
              <surname>Delsole</surname>
              <given-names>EM</given-names>
            </name>
            <name name-style="western">
              <surname>Chu</surname>
              <given-names>ML</given-names>
            </name>
            <name name-style="western">
              <surname>Godfried</surname>
              <given-names>DH</given-names>
            </name>
            <name name-style="western">
              <surname>Bhattacharyya</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Karamitopoulos</surname>
              <given-names>MS</given-names>
            </name>
            <name name-style="western">
              <surname>Chu</surname>
              <given-names>A</given-names>
            </name>
          </person-group>
          <article-title>Measuring ambulation with wrist-based and hip-based activity trackers for children with cerebral palsy</article-title>
          <source>Dev Med Child Neurol</source>
          <year>2019</year>
          <month>11</month>
          <volume>61</volume>
          <issue>11</issue>
          <fpage>1309</fpage>
          <lpage>1313</lpage>
          <pub-id pub-id-type="doi">10.1111/dmcn.14207</pub-id>
          <pub-id pub-id-type="medline">30883727</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref99">
        <label>99</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Schmal</surname>
              <given-names>H</given-names>
            </name>
            <name name-style="western">
              <surname>Holsgaard-Larsen</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Izadpanah</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Brønd</surname>
              <given-names>JC</given-names>
            </name>
            <name name-style="western">
              <surname>Madsen</surname>
              <given-names>CF</given-names>
            </name>
            <name name-style="western">
              <surname>Lauritsen</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>Validation of Activity Tracking Procedures in Elderly Patients after Operative Treatment of Proximal Femur Fractures</article-title>
          <source>Rehabil Res Pract</source>
          <year>2018</year>
          <volume>2018</volume>
          <fpage>3521271</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://doi.org/10.1155/2018/3521271"/>
          </comment>
          <pub-id pub-id-type="doi">10.1155/2018/3521271</pub-id>
          <pub-id pub-id-type="medline">30018824</pub-id>
          <pub-id pub-id-type="pmcid">PMC6029464</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref100">
        <label>100</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Balto</surname>
              <given-names>JM</given-names>
            </name>
            <name name-style="western">
              <surname>Kinnett-Hopkins</surname>
              <given-names>DL</given-names>
            </name>
            <name name-style="western">
              <surname>Motl</surname>
              <given-names>RW</given-names>
            </name>
          </person-group>
          <article-title>Accuracy and precision of smartphone applications and commercially available motion sensors in multiple sclerosis</article-title>
          <source>Mult Scler J Exp Transl Clin</source>
          <year>2016</year>
          <volume>2</volume>
          <fpage>2055217316634754</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://journals.sagepub.com/doi/10.1177/2055217316634754?url_ver=Z39.88-2003&#38;rfr_id=ori:rid:crossref.org&#38;rfr_dat=cr_pub%3dpubmed"/>
          </comment>
          <pub-id pub-id-type="doi">10.1177/2055217316634754</pub-id>
          <pub-id pub-id-type="medline">28607720</pub-id>
          <pub-id pub-id-type="pii">10.1177_2055217316634754</pub-id>
          <pub-id pub-id-type="pmcid">PMC5433404</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref101">
        <label>101</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Sushames</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Edwards</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Thompson</surname>
              <given-names>F</given-names>
            </name>
            <name name-style="western">
              <surname>McDermott</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Gebel</surname>
              <given-names>K</given-names>
            </name>
          </person-group>
          <article-title>Validity and Reliability of Fitbit Flex for Step Count, Moderate to Vigorous Physical Activity and Activity Energy Expenditure</article-title>
          <source>PLoS One</source>
          <year>2016</year>
          <volume>11</volume>
          <issue>9</issue>
          <fpage>e0161224</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0161224"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0161224</pub-id>
          <pub-id pub-id-type="medline">27589592</pub-id>
          <pub-id pub-id-type="pii">PONE-D-15-42907</pub-id>
          <pub-id pub-id-type="pmcid">PMC5010194</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref102">
        <label>102</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Ummels</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Beekman</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Theunissen</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Braun</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Beurskens</surname>
              <given-names>AJ</given-names>
            </name>
          </person-group>
          <article-title>Counting Steps in Activities of Daily Living in People With a Chronic Disease Using Nine Commercially Available Fitness Trackers: Cross-Sectional Validity Study</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2018</year>
          <month>04</month>
          <day>02</day>
          <volume>6</volume>
          <issue>4</issue>
          <fpage>e70</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://mhealth.jmir.org/2018/4/e70/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/mhealth.8524</pub-id>
          <pub-id pub-id-type="medline">29610110</pub-id>
          <pub-id pub-id-type="pii">v6i4e70</pub-id>
          <pub-id pub-id-type="pmcid">PMC5902695</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref103">
        <label>103</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Nelson</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Kaminsky</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Dickin</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Montoye</surname>
              <given-names>A</given-names>
            </name>
          </person-group>
          <article-title>Validity of Consumer-Based Physical Activity Monitors for Specific Activity Types</article-title>
          <source>Med Sci Sports Exerc</source>
          <year>2016</year>
          <month>08</month>
          <volume>48</volume>
          <issue>8</issue>
          <fpage>1619</fpage>
          <lpage>28</lpage>
          <pub-id pub-id-type="doi">10.1249/MSS.0000000000000933</pub-id>
          <pub-id pub-id-type="medline">27015387</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref104">
        <label>104</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Murakami</surname>
              <given-names>H</given-names>
            </name>
            <name name-style="western">
              <surname>Kawakami</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Nakae</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Nakata</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Ishikawa-Takata</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Tanaka</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Miyachi</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of Wearable Devices for Estimating Total Energy Expenditure: Comparison With Metabolic Chamber and Doubly Labeled Water Method</article-title>
          <source>JAMA Intern Med</source>
          <year>2016</year>
          <month>05</month>
          <day>01</day>
          <volume>176</volume>
          <issue>5</issue>
          <fpage>702</fpage>
          <lpage>3</lpage>
          <pub-id pub-id-type="doi">10.1001/jamainternmed.2016.0152</pub-id>
          <pub-id pub-id-type="medline">26999758</pub-id>
          <pub-id pub-id-type="pii">2500062</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref105">
        <label>105</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Bai</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Welk</surname>
              <given-names>GJ</given-names>
            </name>
            <name name-style="western">
              <surname>Nam</surname>
              <given-names>YH</given-names>
            </name>
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>JA</given-names>
            </name>
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Kim</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Meier</surname>
              <given-names>NF</given-names>
            </name>
            <name name-style="western">
              <surname>Dixon</surname>
              <given-names>PM</given-names>
            </name>
          </person-group>
          <article-title>Comparison of Consumer and Research Monitors under Semistructured Settings</article-title>
          <source>Medicine &#38; Science in Sports &#38; Exercise</source>
          <year>2016</year>
          <volume>48</volume>
          <issue>1</issue>
          <fpage>151</fpage>
          <lpage>158</lpage>
          <pub-id pub-id-type="doi">10.1249/mss.0000000000000727</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref106">
        <label>106</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Alharbi</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Bauman</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Neubeck</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Gallagher</surname>
              <given-names>R</given-names>
            </name>
          </person-group>
          <article-title>Validation of Fitbit-Flex as a measure of free-living physical activity in a community-based phase III cardiac rehabilitation population</article-title>
          <source>Eur J Prev Cardiol</source>
          <year>2016</year>
          <month>09</month>
          <volume>23</volume>
          <issue>14</issue>
          <fpage>1476</fpage>
          <lpage>85</lpage>
          <pub-id pub-id-type="doi">10.1177/2047487316634883</pub-id>
          <pub-id pub-id-type="medline">26907794</pub-id>
          <pub-id pub-id-type="pii">2047487316634883</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref107">
        <label>107</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Chu</surname>
              <given-names>AH</given-names>
            </name>
            <name name-style="western">
              <surname>Ng</surname>
              <given-names>SH</given-names>
            </name>
            <name name-style="western">
              <surname>Paknezhad</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Gauterin</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Koh</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Brown</surname>
              <given-names>MS</given-names>
            </name>
            <name name-style="western">
              <surname>Müller-Riemenschneider</surname>
              <given-names>F</given-names>
            </name>
          </person-group>
          <article-title>Comparison of wrist-worn Fitbit Flex and waist-worn ActiGraph for measuring steps in free-living adults</article-title>
          <source>PLoS One</source>
          <year>2017</year>
          <volume>12</volume>
          <issue>2</issue>
          <fpage>e0172535</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0172535"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0172535</pub-id>
          <pub-id pub-id-type="medline">28234953</pub-id>
          <pub-id pub-id-type="pii">PONE-D-16-33057</pub-id>
          <pub-id pub-id-type="pmcid">PMC5325470</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref108">
        <label>108</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Dominick</surname>
              <given-names>GM</given-names>
            </name>
            <name name-style="western">
              <surname>Winfree</surname>
              <given-names>KN</given-names>
            </name>
            <name name-style="western">
              <surname>Pohlig</surname>
              <given-names>RT</given-names>
            </name>
            <name name-style="western">
              <surname>Papas</surname>
              <given-names>MA</given-names>
            </name>
          </person-group>
          <article-title>Physical Activity Assessment Between Consumer- and Research-Grade Accelerometers: A Comparative Study in Free-Living Conditions</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2016</year>
          <month>09</month>
          <day>19</day>
          <volume>4</volume>
          <issue>3</issue>
          <fpage>e110</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://mhealth.jmir.org/2016/3/e110/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/mhealth.6281</pub-id>
          <pub-id pub-id-type="medline">27644334</pub-id>
          <pub-id pub-id-type="pii">v4i3e110</pub-id>
          <pub-id pub-id-type="pmcid">PMC5048058</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref109">
        <label>109</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Block</surname>
              <given-names>VJ</given-names>
            </name>
            <name name-style="western">
              <surname>Lizée</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Crabtree-Hartman</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Bevan</surname>
              <given-names>CJ</given-names>
            </name>
            <name name-style="western">
              <surname>Graves</surname>
              <given-names>JS</given-names>
            </name>
            <name name-style="western">
              <surname>Bove</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Green</surname>
              <given-names>AJ</given-names>
            </name>
            <name name-style="western">
              <surname>Nourbakhsh</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Tremblay</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Gourraud</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Ng</surname>
              <given-names>MY</given-names>
            </name>
            <name name-style="western">
              <surname>Pletcher</surname>
              <given-names>MJ</given-names>
            </name>
            <name name-style="western">
              <surname>Olgin</surname>
              <given-names>JE</given-names>
            </name>
            <name name-style="western">
              <surname>Marcus</surname>
              <given-names>GM</given-names>
            </name>
            <name name-style="western">
              <surname>Allen</surname>
              <given-names>DD</given-names>
            </name>
            <name name-style="western">
              <surname>Cree</surname>
              <given-names>BA</given-names>
            </name>
            <name name-style="western">
              <surname>Gelfand</surname>
              <given-names>JM</given-names>
            </name>
          </person-group>
          <article-title>Continuous daily assessment of multiple sclerosis disability using remote step count monitoring</article-title>
          <source>J Neurol</source>
          <year>2017</year>
          <month>02</month>
          <volume>264</volume>
          <issue>2</issue>
          <fpage>316</fpage>
          <lpage>326</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/27896433"/>
          </comment>
          <pub-id pub-id-type="doi">10.1007/s00415-016-8334-6</pub-id>
          <pub-id pub-id-type="medline">27896433</pub-id>
          <pub-id pub-id-type="pii">10.1007/s00415-016-8334-6</pub-id>
          <pub-id pub-id-type="pmcid">PMC5292081</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref110">
        <label>110</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Reid</surname>
              <given-names>RE</given-names>
            </name>
            <name name-style="western">
              <surname>Insogna</surname>
              <given-names>JA</given-names>
            </name>
            <name name-style="western">
              <surname>Carver</surname>
              <given-names>TE</given-names>
            </name>
            <name name-style="western">
              <surname>Comptour</surname>
              <given-names>AM</given-names>
            </name>
            <name name-style="western">
              <surname>Bewski</surname>
              <given-names>NA</given-names>
            </name>
            <name name-style="western">
              <surname>Sciortino</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Andersen</surname>
              <given-names>RE</given-names>
            </name>
          </person-group>
          <article-title>Validity and reliability of Fitbit activity monitors compared to ActiGraph GT3X+ with female adults in a free-living environment</article-title>
          <source>J Sci Med Sport</source>
          <year>2017</year>
          <month>06</month>
          <volume>20</volume>
          <issue>6</issue>
          <fpage>578</fpage>
          <lpage>582</lpage>
          <pub-id pub-id-type="doi">10.1016/j.jsams.2016.10.015</pub-id>
          <pub-id pub-id-type="medline">27887786</pub-id>
          <pub-id pub-id-type="pii">S1440-2440(16)30231-6</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref111">
        <label>111</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Rowe</surname>
              <given-names>VT</given-names>
            </name>
            <name name-style="western">
              <surname>Neville</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <article-title>Measuring Reliability of Movement With Accelerometry: Fitbit Versus ActiGraph</article-title>
          <source>Am J Occup Ther</source>
          <year>2019</year>
          <volume>73</volume>
          <issue>2</issue>
          <fpage>7302205150p1</fpage>
          <lpage>7302205150p6</lpage>
          <pub-id pub-id-type="doi">10.5014/ajot.2019.030692</pub-id>
          <pub-id pub-id-type="medline">30915976</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref112">
        <label>112</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>St-Laurent</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Mony</surname>
              <given-names>MM</given-names>
            </name>
            <name name-style="western">
              <surname>Mathieu</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Ruchat</surname>
              <given-names>SM</given-names>
            </name>
          </person-group>
          <article-title>Validation of the Fitbit Zip and Fitbit Flex with pregnant women in free-living conditions</article-title>
          <source>J Med Eng Technol</source>
          <year>2018</year>
          <month>05</month>
          <volume>42</volume>
          <issue>4</issue>
          <fpage>259</fpage>
          <lpage>264</lpage>
          <pub-id pub-id-type="doi">10.1080/03091902.2018.1472822</pub-id>
          <pub-id pub-id-type="medline">30198806</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref113">
        <label>113</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Clevenger</surname>
              <given-names>KA</given-names>
            </name>
            <name name-style="western">
              <surname>Molesky</surname>
              <given-names>MJ</given-names>
            </name>
            <name name-style="western">
              <surname>Vusich</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Montoye</surname>
              <given-names>AH</given-names>
            </name>
          </person-group>
          <article-title>Free-Living Comparison of Physical Activity and Sleep Data from Fitbit Activity Trackers Worn on the Dominant and Nondominant Wrists</article-title>
          <source>Measurement in Physical Education and Exercise Science</source>
          <year>2019</year>
          <month>02</month>
          <day>12</day>
          <volume>23</volume>
          <issue>2</issue>
          <fpage>194</fpage>
          <lpage>204</lpage>
          <pub-id pub-id-type="doi">10.1080/1091367x.2019.1577737</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref114">
        <label>114</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Chen</surname>
              <given-names>MD</given-names>
            </name>
            <name name-style="western">
              <surname>Kuo</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Pellegrini</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Hsu</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of Wristband Activity Monitors during Ambulation and Activities</article-title>
          <source>Med Sci Sports Exerc</source>
          <year>2016</year>
          <month>10</month>
          <volume>48</volume>
          <issue>10</issue>
          <fpage>1942</fpage>
          <lpage>9</lpage>
          <pub-id pub-id-type="doi">10.1249/MSS.0000000000000984</pub-id>
          <pub-id pub-id-type="medline">27183123</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref115">
        <label>115</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Daligadu</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Pollock</surname>
              <given-names>CL</given-names>
            </name>
            <name name-style="western">
              <surname>Carlaw</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Chin</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Haynes</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Thevaraajah Kopal</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Tahsinul</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Walters</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Colella</surname>
              <given-names>TJ</given-names>
            </name>
          </person-group>
          <article-title>Validation of the Fitbit Flex in an Acute Post-Cardiac Surgery Patient Population</article-title>
          <source>Physiother Can</source>
          <year>2018</year>
          <volume>70</volume>
          <issue>4</issue>
          <fpage>314</fpage>
          <lpage>320</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/30745716"/>
          </comment>
          <pub-id pub-id-type="doi">10.3138/ptc.2017-34</pub-id>
          <pub-id pub-id-type="medline">30745716</pub-id>
          <pub-id pub-id-type="pii">ptc.2017-34</pub-id>
          <pub-id pub-id-type="pmcid">PMC6361401</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref116">
        <label>116</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Diaz</surname>
              <given-names>KM</given-names>
            </name>
            <name name-style="western">
              <surname>Krupka</surname>
              <given-names>DJ</given-names>
            </name>
            <name name-style="western">
              <surname>Chang</surname>
              <given-names>MJ</given-names>
            </name>
            <name name-style="western">
              <surname>Peacock</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Ma</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Goldsmith</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Schwartz</surname>
              <given-names>JE</given-names>
            </name>
            <name name-style="western">
              <surname>Davidson</surname>
              <given-names>KW</given-names>
            </name>
          </person-group>
          <article-title>Fitbit®: An accurate and reliable device for wireless physical activity tracking</article-title>
          <source>Int J Cardiol</source>
          <year>2015</year>
          <month>04</month>
          <day>15</day>
          <volume>185</volume>
          <fpage>138</fpage>
          <lpage>40</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/25795203"/>
          </comment>
          <pub-id pub-id-type="doi">10.1016/j.ijcard.2015.03.038</pub-id>
          <pub-id pub-id-type="medline">25795203</pub-id>
          <pub-id pub-id-type="pii">S0167-5273(15)00276-4</pub-id>
          <pub-id pub-id-type="pmcid">PMC4406840</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref117">
        <label>117</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Diaz</surname>
              <given-names>KM</given-names>
            </name>
            <name name-style="western">
              <surname>Krupka</surname>
              <given-names>DJ</given-names>
            </name>
            <name name-style="western">
              <surname>Chang</surname>
              <given-names>MJ</given-names>
            </name>
            <name name-style="western">
              <surname>Shaffer</surname>
              <given-names>JA</given-names>
            </name>
            <name name-style="western">
              <surname>Ma</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Goldsmith</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Schwartz</surname>
              <given-names>JE</given-names>
            </name>
            <name name-style="western">
              <surname>Davidson</surname>
              <given-names>KW</given-names>
            </name>
          </person-group>
          <article-title>Validation of the Fitbit One® for physical activity measurement at an upper torso attachment site</article-title>
          <source>BMC Res Notes</source>
          <year>2016</year>
          <month>04</month>
          <day>12</day>
          <volume>9</volume>
          <fpage>213</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://bmcresnotes.biomedcentral.com/articles/10.1186/s13104-016-2020-8"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/s13104-016-2020-8</pub-id>
          <pub-id pub-id-type="medline">27068022</pub-id>
          <pub-id pub-id-type="pii">10.1186/s13104-016-2020-8</pub-id>
          <pub-id pub-id-type="pmcid">PMC4828816</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref118">
        <label>118</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Battenberg</surname>
              <given-names>AK</given-names>
            </name>
            <name name-style="western">
              <surname>Donohoe</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Robertson</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Schmalzried</surname>
              <given-names>TP</given-names>
            </name>
          </person-group>
          <article-title>The accuracy of personal activity monitoring devices</article-title>
          <source>Seminars in Arthroplasty</source>
          <year>2017</year>
          <month>06</month>
          <volume>28</volume>
          <issue>2</issue>
          <fpage>71</fpage>
          <lpage>75</lpage>
          <pub-id pub-id-type="doi">10.1053/j.sart.2017.07.006</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref119">
        <label>119</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Prieto-Centurion</surname>
              <given-names>V</given-names>
            </name>
            <name name-style="western">
              <surname>Bracken</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Norwick</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Zaidi</surname>
              <given-names>F</given-names>
            </name>
            <name name-style="western">
              <surname>Mutso</surname>
              <given-names>AA</given-names>
            </name>
            <name name-style="western">
              <surname>Morken</surname>
              <given-names>V</given-names>
            </name>
            <name name-style="western">
              <surname>Coultas</surname>
              <given-names>DB</given-names>
            </name>
            <name name-style="western">
              <surname>Rand</surname>
              <given-names>CS</given-names>
            </name>
            <name name-style="western">
              <surname>Marquez</surname>
              <given-names>DX</given-names>
            </name>
            <name name-style="western">
              <surname>Krishnan</surname>
              <given-names>JA</given-names>
            </name>
          </person-group>
          <article-title>Can Commercially Available Pedometers Be Used For Physical Activity Monitoring In Patients With COPD Following Exacerbations?</article-title>
          <source>Chronic Obstr Pulm Dis</source>
          <year>2016</year>
          <volume>3</volume>
          <issue>3</issue>
          <fpage>636</fpage>
          <lpage>642</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://doi.org/10.15326/jcopdf.3.3.2015.0164"/>
          </comment>
          <pub-id pub-id-type="doi">10.15326/jcopdf.3.3.2015.0164</pub-id>
          <pub-id pub-id-type="medline">27547817</pub-id>
          <pub-id pub-id-type="pmcid">PMC4989920</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref120">
        <label>120</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Arch</surname>
              <given-names>ES</given-names>
            </name>
            <name name-style="western">
              <surname>Sions</surname>
              <given-names>JM</given-names>
            </name>
            <name name-style="western">
              <surname>Horne</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Bodt</surname>
              <given-names>BA</given-names>
            </name>
          </person-group>
          <article-title>Step count accuracy of StepWatch and FitBit One™ among individuals with a unilateral transtibial amputation</article-title>
          <source>Prosthet Orthot Int</source>
          <year>2018</year>
          <month>10</month>
          <volume>42</volume>
          <issue>5</issue>
          <fpage>518</fpage>
          <lpage>526</lpage>
          <pub-id pub-id-type="doi">10.1177/0309364618767138</pub-id>
          <pub-id pub-id-type="medline">29623810</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref121">
        <label>121</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Klassen</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Simpson</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Lim</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Louie</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Parappilly</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Sakakibara</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Zbogar</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Eng</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>"Stepping Up" Activity Poststroke: Ankle-Positioned Accelerometer Can Accurately Record Steps During Slow Walking</article-title>
          <source>Phys Ther</source>
          <year>2016</year>
          <month>03</month>
          <volume>96</volume>
          <issue>3</issue>
          <fpage>355</fpage>
          <lpage>60</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/26251478"/>
          </comment>
          <pub-id pub-id-type="doi">10.2522/ptj.20140611</pub-id>
          <pub-id pub-id-type="medline">26251478</pub-id>
          <pub-id pub-id-type="pii">ptj.20140611</pub-id>
          <pub-id pub-id-type="pmcid">PMC4774387</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref122">
        <label>122</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>O'Connell</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>ÓLaighin</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Kelly</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Murphy</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Beirne</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Burke</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Kilgannon</surname>
              <given-names>O</given-names>
            </name>
            <name name-style="western">
              <surname>Quinlan</surname>
              <given-names>LR</given-names>
            </name>
          </person-group>
          <article-title>These Shoes Are Made for Walking: Sensitivity Performance Evaluation of Commercial Activity Monitors under the Expected Conditions and Circumstances Required to Achieve the International Daily Step Goal of 10,000 Steps</article-title>
          <source>PLoS One</source>
          <year>2016</year>
          <volume>11</volume>
          <issue>5</issue>
          <fpage>e0154956</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0154956"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0154956</pub-id>
          <pub-id pub-id-type="medline">27167121</pub-id>
          <pub-id pub-id-type="pii">PONE-D-15-48144</pub-id>
          <pub-id pub-id-type="pmcid">PMC4864247</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref123">
        <label>123</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>O'Connell</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>ÓLaighin</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Quinlan</surname>
              <given-names>LR</given-names>
            </name>
          </person-group>
          <article-title>When a Step Is Not a Step! Specificity Analysis of Five Physical Activity Monitors</article-title>
          <source>PLoS One</source>
          <year>2017</year>
          <volume>12</volume>
          <issue>1</issue>
          <fpage>e0169616</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0169616"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0169616</pub-id>
          <pub-id pub-id-type="medline">28085918</pub-id>
          <pub-id pub-id-type="pii">PONE-D-16-29880</pub-id>
          <pub-id pub-id-type="pmcid">PMC5234787</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref124">
        <label>124</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Simpson</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Eng</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Klassen</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Lim</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Louie</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Parappilly</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Sakakibara</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Zbogar</surname>
              <given-names>D</given-names>
            </name>
          </person-group>
          <article-title>Capturing step counts at slow walking speeds in older adults: comparison of ankle and waist placement of measuring device</article-title>
          <source>J Rehabil Med</source>
          <year>2015</year>
          <month>10</month>
          <day>05</day>
          <volume>47</volume>
          <issue>9</issue>
          <fpage>830</fpage>
          <lpage>5</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://www.medicaljournals.se/jrm/content/abstract/10.2340/16501977-1993"/>
          </comment>
          <pub-id pub-id-type="doi">10.2340/16501977-1993</pub-id>
          <pub-id pub-id-type="medline">26181670</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref125">
        <label>125</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Takacs</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Pollock</surname>
              <given-names>CL</given-names>
            </name>
            <name name-style="western">
              <surname>Guenther</surname>
              <given-names>JR</given-names>
            </name>
            <name name-style="western">
              <surname>Bahar</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Napier</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Hunt</surname>
              <given-names>MA</given-names>
            </name>
          </person-group>
          <article-title>Validation of the Fitbit One activity monitor device during treadmill walking</article-title>
          <source>J Sci Med Sport</source>
          <year>2014</year>
          <month>09</month>
          <volume>17</volume>
          <issue>5</issue>
          <fpage>496</fpage>
          <lpage>500</lpage>
          <pub-id pub-id-type="doi">10.1016/j.jsams.2013.10.241</pub-id>
          <pub-id pub-id-type="medline">24268570</pub-id>
          <pub-id pub-id-type="pii">S1440-2440(13)00472-6</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref126">
        <label>126</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Klassen</surname>
              <given-names>TD</given-names>
            </name>
            <name name-style="western">
              <surname>Semrau</surname>
              <given-names>JA</given-names>
            </name>
            <name name-style="western">
              <surname>Dukelow</surname>
              <given-names>SP</given-names>
            </name>
            <name name-style="western">
              <surname>Bayley</surname>
              <given-names>MT</given-names>
            </name>
            <name name-style="western">
              <surname>Hill</surname>
              <given-names>MD</given-names>
            </name>
            <name name-style="western">
              <surname>Eng</surname>
              <given-names>JJ</given-names>
            </name>
          </person-group>
          <article-title>Consumer-Based Physical Activity Monitor as a Practical Way to Measure Walking Intensity During Inpatient Stroke Rehabilitation</article-title>
          <source>Stroke</source>
          <year>2017</year>
          <month>09</month>
          <volume>48</volume>
          <issue>9</issue>
          <fpage>2614</fpage>
          <lpage>2617</lpage>
          <pub-id pub-id-type="doi">10.1161/strokeaha.117.018175</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref127">
        <label>127</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Leth</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Hansen</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Nielsen</surname>
              <given-names>O</given-names>
            </name>
            <name name-style="western">
              <surname>Dinesen</surname>
              <given-names>B</given-names>
            </name>
          </person-group>
          <article-title>Evaluation of Commercial Self-Monitoring Devices for Clinical Purposes: Results from the Future Patient Trial, Phase I</article-title>
          <source>Sensors (Basel)</source>
          <year>2017</year>
          <month>01</month>
          <day>22</day>
          <volume>17</volume>
          <issue>1</issue>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://www.mdpi.com/resolver?pii=s17010211"/>
          </comment>
          <pub-id pub-id-type="doi">10.3390/s17010211</pub-id>
          <pub-id pub-id-type="medline">28117736</pub-id>
          <pub-id pub-id-type="pii">s17010211</pub-id>
          <pub-id pub-id-type="pmcid">PMC5298782</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref128">
        <label>128</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Storm</surname>
              <given-names>FA</given-names>
            </name>
            <name name-style="western">
              <surname>Heller</surname>
              <given-names>BW</given-names>
            </name>
            <name name-style="western">
              <surname>Mazzà</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
          <article-title>Step detection and activity recognition accuracy of seven physical activity monitors</article-title>
          <source>PLoS One</source>
          <year>2015</year>
          <volume>10</volume>
          <issue>3</issue>
          <fpage>e0118723</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0118723"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0118723</pub-id>
          <pub-id pub-id-type="medline">25789630</pub-id>
          <pub-id pub-id-type="pii">PONE-D-14-23091</pub-id>
          <pub-id pub-id-type="pmcid">PMC4366111</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref129">
        <label>129</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Kim</surname>
              <given-names>Y</given-names>
            </name>
            <name name-style="western">
              <surname>Welk</surname>
              <given-names>GJ</given-names>
            </name>
          </person-group>
          <article-title>Validity of consumer-based physical activity monitors</article-title>
          <source>Med Sci Sports Exerc</source>
          <year>2014</year>
          <month>09</month>
          <volume>46</volume>
          <issue>9</issue>
          <fpage>1840</fpage>
          <lpage>8</lpage>
          <pub-id pub-id-type="doi">10.1249/MSS.0000000000000287</pub-id>
          <pub-id pub-id-type="medline">24777201</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref130">
        <label>130</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Price</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Bird</surname>
              <given-names>SR</given-names>
            </name>
            <name name-style="western">
              <surname>Lythgo</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Raj</surname>
              <given-names>IS</given-names>
            </name>
            <name name-style="western">
              <surname>Wong</surname>
              <given-names>JY</given-names>
            </name>
            <name name-style="western">
              <surname>Lynch</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
          <article-title>Validation of the Fitbit One, Garmin Vivofit and Jawbone UP activity tracker in estimation of energy expenditure during treadmill walking and running</article-title>
          <source>J Med Eng Technol</source>
          <year>2017</year>
          <month>04</month>
          <volume>41</volume>
          <issue>3</issue>
          <fpage>208</fpage>
          <lpage>215</lpage>
          <pub-id pub-id-type="doi">10.1080/03091902.2016.1253795</pub-id>
          <pub-id pub-id-type="medline">27919170</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref131">
        <label>131</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Ferguson</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Rowlands</surname>
              <given-names>AV</given-names>
            </name>
            <name name-style="western">
              <surname>Olds</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Maher</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
          <article-title>The validity of consumer-level, activity monitors in healthy adults worn in free-living conditions: a cross-sectional study</article-title>
          <source>Int J Behav Nutr Phys Act</source>
          <year>2015</year>
          <month>03</month>
          <day>27</day>
          <volume>12</volume>
          <fpage>42</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://ijbnpa.biomedcentral.com/articles/10.1186/s12966-015-0201-9"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/s12966-015-0201-9</pub-id>
          <pub-id pub-id-type="medline">25890168</pub-id>
          <pub-id pub-id-type="pii">10.1186/s12966-015-0201-9</pub-id>
          <pub-id pub-id-type="pmcid">PMC4416251</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref132">
        <label>132</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Gomersall</surname>
              <given-names>SR</given-names>
            </name>
            <name name-style="western">
              <surname>Ng</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Burton</surname>
              <given-names>NW</given-names>
            </name>
            <name name-style="western">
              <surname>Pavey</surname>
              <given-names>TG</given-names>
            </name>
            <name name-style="western">
              <surname>Gilson</surname>
              <given-names>ND</given-names>
            </name>
            <name name-style="western">
              <surname>Brown</surname>
              <given-names>WJ</given-names>
            </name>
          </person-group>
          <article-title>Estimating Physical Activity and Sedentary Behavior in a Free-Living Context: A Pragmatic Comparison of Consumer-Based Activity Trackers and ActiGraph Accelerometry</article-title>
          <source>J Med Internet Res</source>
          <year>2016</year>
          <month>09</month>
          <day>07</day>
          <volume>18</volume>
          <issue>9</issue>
          <fpage>e239</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://www.jmir.org/2016/9/e239/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/jmir.5531</pub-id>
          <pub-id pub-id-type="medline">27604226</pub-id>
          <pub-id pub-id-type="pii">v18i9e239</pub-id>
          <pub-id pub-id-type="pmcid">PMC5031913</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref133">
        <label>133</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Hamari</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Kullberg</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Ruohonen</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Heinonen</surname>
              <given-names>O</given-names>
            </name>
            <name name-style="western">
              <surname>Díaz-Rodríguez</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Lilius</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Pakarinen</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Myllymäki</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Leppänen</surname>
              <given-names>V</given-names>
            </name>
            <name name-style="western">
              <surname>Salanterä</surname>
              <given-names>S</given-names>
            </name>
          </person-group>
          <article-title>Physical activity among children: objective measurements using Fitbit One and ActiGraph</article-title>
          <source>BMC Res Notes</source>
          <year>2017</year>
          <month>04</month>
          <day>20</day>
          <volume>10</volume>
          <issue>1</issue>
          <fpage>161</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://bmcresnotes.biomedcentral.com/articles/10.1186/s13104-017-2476-1"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/s13104-017-2476-1</pub-id>
          <pub-id pub-id-type="medline">28427441</pub-id>
          <pub-id pub-id-type="pii">10.1186/s13104-017-2476-1</pub-id>
          <pub-id pub-id-type="pmcid">PMC5397828</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref134">
        <label>134</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Hui</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Heyden</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Bao</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Accettone</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>McBay</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Richardson</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Tang</surname>
              <given-names>A</given-names>
            </name>
          </person-group>
          <article-title>Validity of the Fitbit One for Measuring Activity in Community-Dwelling Stroke Survivors</article-title>
          <source>Physiother Can</source>
          <year>2018</year>
          <volume>70</volume>
          <issue>1</issue>
          <fpage>81</fpage>
          <lpage>89</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/29434422"/>
          </comment>
          <pub-id pub-id-type="doi">10.3138/ptc.2016-40.ep</pub-id>
          <pub-id pub-id-type="medline">29434422</pub-id>
          <pub-id pub-id-type="pii">ptc.2016-40.ep</pub-id>
          <pub-id pub-id-type="pmcid">PMC5802948</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref135">
        <label>135</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Middelweerd</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Van Der Ploeg</surname>
              <given-names>HP</given-names>
            </name>
            <name name-style="western">
              <surname>Van Halteren</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Twisk</surname>
              <given-names>JW</given-names>
            </name>
            <name name-style="western">
              <surname>Brug</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Te Velde</surname>
              <given-names>SJ</given-names>
            </name>
          </person-group>
          <article-title>A Validation Study of the Fitbit One in Daily Life Using Different Time Intervals</article-title>
          <source>Med Sci Sports Exerc</source>
          <year>2017</year>
          <month>06</month>
          <volume>49</volume>
          <issue>6</issue>
          <fpage>1270</fpage>
          <lpage>1279</lpage>
          <pub-id pub-id-type="doi">10.1249/MSS.0000000000001225</pub-id>
          <pub-id pub-id-type="medline">28511193</pub-id>
          <pub-id pub-id-type="pii">00005768-201706000-00025</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref136">
        <label>136</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Van Blarigan</surname>
              <given-names>EL</given-names>
            </name>
            <name name-style="western">
              <surname>Kenfield</surname>
              <given-names>SA</given-names>
            </name>
            <name name-style="western">
              <surname>Tantum</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Cadmus-Bertram</surname>
              <given-names>LA</given-names>
            </name>
            <name name-style="western">
              <surname>Carroll</surname>
              <given-names>PR</given-names>
            </name>
            <name name-style="western">
              <surname>Chan</surname>
              <given-names>JM</given-names>
            </name>
          </person-group>
          <article-title>The Fitbit One Physical Activity Tracker in Men With Prostate Cancer: Validation Study</article-title>
          <source>JMIR Cancer</source>
          <year>2017</year>
          <month>04</month>
          <day>18</day>
          <volume>3</volume>
          <issue>1</issue>
          <fpage>e5</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://cancer.jmir.org/2017/1/e5/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/cancer.6935</pub-id>
          <pub-id pub-id-type="medline">28420602</pub-id>
          <pub-id pub-id-type="pii">v3i1e5</pub-id>
          <pub-id pub-id-type="pmcid">PMC5413797</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref137">
        <label>137</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Rosenberger</surname>
              <given-names>ME</given-names>
            </name>
            <name name-style="western">
              <surname>Buman</surname>
              <given-names>MP</given-names>
            </name>
            <name name-style="western">
              <surname>Haskell</surname>
              <given-names>WL</given-names>
            </name>
            <name name-style="western">
              <surname>McConnell</surname>
              <given-names>MV</given-names>
            </name>
            <name name-style="western">
              <surname>Carstensen</surname>
              <given-names>LL</given-names>
            </name>
          </person-group>
          <article-title>Twenty-four Hours of Sleep, Sedentary Behavior, and Physical Activity with Nine Wearable Devices</article-title>
          <source>Med Sci Sports Exerc</source>
          <year>2016</year>
          <month>03</month>
          <volume>48</volume>
          <issue>3</issue>
          <fpage>457</fpage>
          <lpage>65</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://europepmc.org/abstract/MED/26484953"/>
          </comment>
          <pub-id pub-id-type="doi">10.1249/MSS.0000000000000778</pub-id>
          <pub-id pub-id-type="medline">26484953</pub-id>
          <pub-id pub-id-type="pmcid">PMC4760880</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref138">
        <label>138</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Duclos</surname>
              <given-names>NC</given-names>
            </name>
            <name name-style="western">
              <surname>Aguiar</surname>
              <given-names>LT</given-names>
            </name>
            <name name-style="western">
              <surname>Aissaoui</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Faria</surname>
              <given-names>CD</given-names>
            </name>
            <name name-style="western">
              <surname>Nadeau</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Duclos</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
          <article-title>Activity Monitor Placed at the Nonparetic Ankle Is Accurate in Measuring Step Counts During Community Walking in Poststroke Individuals: A Validation Study</article-title>
          <source>PM R</source>
          <year>2019</year>
          <month>09</month>
          <volume>11</volume>
          <issue>9</issue>
          <fpage>963</fpage>
          <lpage>971</lpage>
          <pub-id pub-id-type="doi">10.1002/pmrj.12080</pub-id>
          <pub-id pub-id-type="medline">30688026</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref139">
        <label>139</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Bunn</surname>
              <given-names>JA</given-names>
            </name>
            <name name-style="western">
              <surname>Jones</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Oliviera</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Webster</surname>
              <given-names>MJ</given-names>
            </name>
          </person-group>
          <article-title>Assessment of step accuracy using the Consumer Technology Association standard</article-title>
          <source>J Sports Sci</source>
          <year>2019</year>
          <month>02</month>
          <day>29</day>
          <volume>37</volume>
          <issue>3</issue>
          <fpage>244</fpage>
          <lpage>248</lpage>
          <pub-id pub-id-type="doi">10.1080/02640414.2018.1491941</pub-id>
          <pub-id pub-id-type="medline">29958058</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref140">
        <label>140</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Navalta</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Montes</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Bodell</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Aguilar</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Lujan</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Guzman</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Kam</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Manning</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>DeBeliso</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <article-title>Wearable Device Validity in Determining Step Count During Hiking and Trail Running</article-title>
          <source>Journal for the Measurement of Physical Behaviour</source>
          <year>2018</year>
          <volume>1</volume>
          <issue>2</issue>
          <fpage>86</fpage>
          <lpage>93</lpage>
          <pub-id pub-id-type="doi">10.1123/jmpb.2018-0010</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref141">
        <label>141</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Wendel</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Macpherson</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Webber</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Hendron</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>DeAngelis</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Colon-Semenza</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Ellis</surname>
              <given-names>T</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of Activity Trackers in Parkinson Disease: Should We Prescribe Them?</article-title>
          <source>Phys Ther</source>
          <year>2018</year>
          <month>08</month>
          <day>01</day>
          <volume>98</volume>
          <issue>8</issue>
          <fpage>705</fpage>
          <lpage>714</lpage>
          <pub-id pub-id-type="doi">10.1093/ptj/pzy054</pub-id>
          <pub-id pub-id-type="medline">29718452</pub-id>
          <pub-id pub-id-type="pii">4990561</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref142">
        <label>142</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Thiebaud</surname>
              <given-names>RS</given-names>
            </name>
            <name name-style="western">
              <surname>Funk</surname>
              <given-names>MD</given-names>
            </name>
            <name name-style="western">
              <surname>Patton</surname>
              <given-names>JC</given-names>
            </name>
            <name name-style="western">
              <surname>Massey</surname>
              <given-names>BL</given-names>
            </name>
            <name name-style="western">
              <surname>Shay</surname>
              <given-names>TE</given-names>
            </name>
            <name name-style="western">
              <surname>Schmidt</surname>
              <given-names>MG</given-names>
            </name>
            <name name-style="western">
              <surname>Giovannitti</surname>
              <given-names>N</given-names>
            </name>
          </person-group>
          <article-title>Validity of wrist-worn consumer products to measure heart rate and energy expenditure</article-title>
          <source>Digit Health</source>
          <year>2018</year>
          <month>04</month>
          <day>13</day>
          <volume>4</volume>
          <fpage>2055207618770322</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://journals.sagepub.com/doi/10.1177/2055207618770322?url_ver=Z39.88-2003&#38;rfr_id=ori:rid:crossref.org&#38;rfr_dat=cr_pub%3dpubmed"/>
          </comment>
          <pub-id pub-id-type="doi">10.1177/2055207618770322</pub-id>
          <pub-id pub-id-type="medline">29942628</pub-id>
          <pub-id pub-id-type="pii">10.1177_2055207618770322</pub-id>
          <pub-id pub-id-type="pmcid">PMC6001222</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref143">
        <label>143</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Pribyslavska</surname>
              <given-names>V</given-names>
            </name>
            <name name-style="western">
              <surname>Caputo</surname>
              <given-names>JL</given-names>
            </name>
            <name name-style="western">
              <surname>Coons</surname>
              <given-names>JM</given-names>
            </name>
            <name name-style="western">
              <surname>Barry</surname>
              <given-names>VW</given-names>
            </name>
          </person-group>
          <article-title>Impact of EPOC adjustment on estimation of energy expenditure using activity monitors</article-title>
          <source>J Med Eng Technol</source>
          <year>2018</year>
          <month>05</month>
          <day>18</day>
          <volume>42</volume>
          <issue>4</issue>
          <fpage>265</fpage>
          <lpage>273</lpage>
          <pub-id pub-id-type="doi">10.1080/03091902.2018.1472823</pub-id>
          <pub-id pub-id-type="medline">29911930</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref144">
        <label>144</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Fulk</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Combs</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Danks</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Nirider</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Raja</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Reisman</surname>
              <given-names>D</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of 2 activity monitors in detecting steps in people with stroke and traumatic brain injury</article-title>
          <source>Phys Ther</source>
          <year>2014</year>
          <month>02</month>
          <volume>94</volume>
          <issue>2</issue>
          <fpage>222</fpage>
          <lpage>9</lpage>
          <pub-id pub-id-type="doi">10.2522/ptj.20120525</pub-id>
          <pub-id pub-id-type="medline">24052577</pub-id>
          <pub-id pub-id-type="pii">ptj.20120525</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref145">
        <label>145</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Park</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>VJ</given-names>
            </name>
            <name name-style="western">
              <surname>Ku</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Tanaka</surname>
              <given-names>H</given-names>
            </name>
          </person-group>
          <article-title>Effect of walking speed and placement position interactions in determining the accuracy of various newer pedometers</article-title>
          <source>Journal of Exercise Science &#38; Fitness</source>
          <year>2014</year>
          <month>06</month>
          <volume>12</volume>
          <issue>1</issue>
          <fpage>31</fpage>
          <lpage>37</lpage>
          <pub-id pub-id-type="doi">10.1016/j.jesf.2014.01.003</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref146">
        <label>146</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Stackpool</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Porcari</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Mikat</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Gillette</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Foster</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
          <article-title>The Accuracy of Various Activity Trackers in Estimating Steps Taken and Energy Expenditure</article-title>
          <source>J Fit Res</source>
          <year>2014</year>
          <volume>3</volume>
          <issue>3</issue>
          <fpage>32</fpage>
          <lpage>48</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://minds.wisconsin.edu/bitstream/handle/1793/66292/Stackpool_Caitlin_Thesis.pdf?sequence=1"/>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref147">
        <label>147</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Wong</surname>
              <given-names>CK</given-names>
            </name>
            <name name-style="western">
              <surname>Mentis</surname>
              <given-names>HM</given-names>
            </name>
            <name name-style="western">
              <surname>Kuber</surname>
              <given-names>R</given-names>
            </name>
          </person-group>
          <article-title>The bit doesn't fit: Evaluation of a commercial activity-tracker at slower walking speeds</article-title>
          <source>Gait Posture</source>
          <year>2018</year>
          <month>01</month>
          <volume>59</volume>
          <fpage>177</fpage>
          <lpage>181</lpage>
          <pub-id pub-id-type="doi">10.1016/j.gaitpost.2017.10.010</pub-id>
          <pub-id pub-id-type="medline">29049964</pub-id>
          <pub-id pub-id-type="pii">S0966-6362(17)30966-9</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref148">
        <label>148</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Gusmer</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Bosch</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Watkins</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Ostrem</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Dengel</surname>
              <given-names>D</given-names>
            </name>
          </person-group>
          <article-title>Comparison of FitBit® Ultra to ActiGraph™ GT1M for Assessment of Physical Activity in Young Adults During Treadmill Walking</article-title>
          <source>TOSMJ</source>
          <year>2014</year>
          <month>04</month>
          <day>04</day>
          <volume>8</volume>
          <issue>1</issue>
          <fpage>11</fpage>
          <lpage>15</lpage>
          <pub-id pub-id-type="doi">10.2174/1874387001408010011</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref149">
        <label>149</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Paul</surname>
              <given-names>SS</given-names>
            </name>
            <name name-style="western">
              <surname>Tiedemann</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Hassett</surname>
              <given-names>LM</given-names>
            </name>
            <name name-style="western">
              <surname>Ramsay</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Kirkham</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Chagpar</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Sherrington</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
          <article-title>Validity of the  activity tracker for measuring steps in community-dwelling older adults</article-title>
          <source>BMJ Open Sport Exerc Med</source>
          <year>2015</year>
          <volume>1</volume>
          <issue>1</issue>
          <fpage>e000013</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://doi.org/10.1136/bmjsem-2015-000013"/>
          </comment>
          <pub-id pub-id-type="doi">10.1136/bmjsem-2015-000013</pub-id>
          <pub-id pub-id-type="medline">27900119</pub-id>
          <pub-id pub-id-type="pii">bmjsem-2015-000013</pub-id>
          <pub-id pub-id-type="pmcid">PMC5117050</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref150">
        <label>150</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Schaffer</surname>
              <given-names>SD</given-names>
            </name>
            <name name-style="western">
              <surname>Holzapfel</surname>
              <given-names>SD</given-names>
            </name>
            <name name-style="western">
              <surname>Fulk</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Bosch</surname>
              <given-names>PR</given-names>
            </name>
          </person-group>
          <article-title>Step count accuracy and reliability of two activity tracking devices in people after stroke</article-title>
          <source>Physiother Theory Pract</source>
          <year>2017</year>
          <month>10</month>
          <day>04</day>
          <volume>33</volume>
          <issue>10</issue>
          <fpage>788</fpage>
          <lpage>796</lpage>
          <pub-id pub-id-type="doi">10.1080/09593985.2017.1354412</pub-id>
          <pub-id pub-id-type="medline">28777710</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref151">
        <label>151</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Sharp</surname>
              <given-names>CA</given-names>
            </name>
            <name name-style="western">
              <surname>Mackintosh</surname>
              <given-names>KA</given-names>
            </name>
            <name name-style="western">
              <surname>Erjavec</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Pascoe</surname>
              <given-names>DM</given-names>
            </name>
            <name name-style="western">
              <surname>Horne</surname>
              <given-names>PJ</given-names>
            </name>
          </person-group>
          <article-title>Validity and reliability of the Fitbit Zip as a measure of preschool children's step count</article-title>
          <source>BMJ Open Sport Exerc Med</source>
          <year>2017</year>
          <volume>3</volume>
          <issue>1</issue>
          <fpage>e000272</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://doi.org/10.1136/bmjsem-2017-000272"/>
          </comment>
          <pub-id pub-id-type="doi">10.1136/bmjsem-2017-000272</pub-id>
          <pub-id pub-id-type="medline">29081984</pub-id>
          <pub-id pub-id-type="pii">bmjsem-2017-000272</pub-id>
          <pub-id pub-id-type="pmcid">PMC5652475</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref152">
        <label>152</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Singh</surname>
              <given-names>AK</given-names>
            </name>
            <name name-style="western">
              <surname>Farmer</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Van Den Berg</surname>
              <given-names>ML</given-names>
            </name>
            <name name-style="western">
              <surname>Killington</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Barr</surname>
              <given-names>CJ</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of the FitBit at walking speeds and cadences relevant to clinical rehabilitation populations</article-title>
          <source>Disabil Health J</source>
          <year>2016</year>
          <month>04</month>
          <volume>9</volume>
          <issue>2</issue>
          <fpage>320</fpage>
          <lpage>3</lpage>
          <pub-id pub-id-type="doi">10.1016/j.dhjo.2015.10.011</pub-id>
          <pub-id pub-id-type="medline">26905972</pub-id>
          <pub-id pub-id-type="pii">S1936-6574(15)00174-0</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref153">
        <label>153</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Appelboom</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Taylor</surname>
              <given-names>BE</given-names>
            </name>
            <name name-style="western">
              <surname>Bruce</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Bassile</surname>
              <given-names>CC</given-names>
            </name>
            <name name-style="western">
              <surname>Malakidis</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Yang</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Youngerman</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>D'Amico</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Bruce</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Bruyère</surname>
              <given-names>O</given-names>
            </name>
            <name name-style="western">
              <surname>Reginster</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Dumont</surname>
              <given-names>EP</given-names>
            </name>
            <name name-style="western">
              <surname>Connolly</surname>
              <given-names>ES</given-names>
            </name>
          </person-group>
          <article-title>Mobile Phone-Connected Wearable Motion Sensors to Assess Postoperative Mobilization</article-title>
          <source>JMIR Mhealth Uhealth</source>
          <year>2015</year>
          <month>07</month>
          <day>28</day>
          <volume>3</volume>
          <issue>3</issue>
          <fpage>e78</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://mhealth.jmir.org/2015/3/e78/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/mhealth.3785</pub-id>
          <pub-id pub-id-type="medline">26220691</pub-id>
          <pub-id pub-id-type="pii">v3i3e78</pub-id>
          <pub-id pub-id-type="pmcid">PMC4705357</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref154">
        <label>154</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Thorup</surname>
              <given-names>CB</given-names>
            </name>
            <name name-style="western">
              <surname>Andreasen</surname>
              <given-names>JJ</given-names>
            </name>
            <name name-style="western">
              <surname>Sørensen</surname>
              <given-names>EE</given-names>
            </name>
            <name name-style="western">
              <surname>Grønkjær</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Dinesen</surname>
              <given-names>BI</given-names>
            </name>
            <name name-style="western">
              <surname>Hansen</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of a step counter during treadmill and daily life walking by healthy adults and patients with cardiac disease</article-title>
          <source>BMJ Open</source>
          <year>2017</year>
          <month>03</month>
          <day>31</day>
          <volume>7</volume>
          <issue>3</issue>
          <fpage>e011742</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://bmjopen.bmj.com/cgi/pmidlookup?view=long&#38;pmid=28363918"/>
          </comment>
          <pub-id pub-id-type="doi">10.1136/bmjopen-2016-011742</pub-id>
          <pub-id pub-id-type="medline">28363918</pub-id>
          <pub-id pub-id-type="pii">bmjopen-2016-011742</pub-id>
          <pub-id pub-id-type="pmcid">PMC5387973</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref155">
        <label>155</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Mooses</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Oja</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Reisberg</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Vilo</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Kull</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <article-title>Validating Fitbit Zip for monitoring physical activity of children in school: a cross-sectional study</article-title>
          <source>BMC Public Health</source>
          <year>2018</year>
          <month>07</month>
          <day>11</day>
          <volume>18</volume>
          <issue>1</issue>
          <fpage>858</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://bmcpublichealth.biomedcentral.com/articles/10.1186/s12889-018-5752-7"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/s12889-018-5752-7</pub-id>
          <pub-id pub-id-type="medline">29996797</pub-id>
          <pub-id pub-id-type="pii">10.1186/s12889-018-5752-7</pub-id>
          <pub-id pub-id-type="pmcid">PMC6042315</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref156">
        <label>156</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Tully</surname>
              <given-names>MA</given-names>
            </name>
            <name name-style="western">
              <surname>McBride</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Heron</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Hunter</surname>
              <given-names>RF</given-names>
            </name>
          </person-group>
          <article-title>The validation of Fibit Zip™ physical activity monitor as a measure of free-living physical activity</article-title>
          <source>BMC Res Notes</source>
          <year>2014</year>
          <month>12</month>
          <day>23</day>
          <volume>7</volume>
          <fpage>952</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://bmcresnotes.biomedcentral.com/articles/10.1186/1756-0500-7-952"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/1756-0500-7-952</pub-id>
          <pub-id pub-id-type="medline">25539733</pub-id>
          <pub-id pub-id-type="pii">1756-0500-7-952</pub-id>
          <pub-id pub-id-type="pmcid">PMC4307145</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref157">
        <label>157</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Schneider</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Chau</surname>
              <given-names>L</given-names>
            </name>
          </person-group>
          <article-title>Validation of the Fitbit Zip for monitoring physical activity among free-living adolescents</article-title>
          <source>BMC Res Notes</source>
          <year>2016</year>
          <month>09</month>
          <day>21</day>
          <volume>9</volume>
          <issue>1</issue>
          <fpage>448</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://bmcresnotes.biomedcentral.com/articles/10.1186/s13104-016-2253-6"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/s13104-016-2253-6</pub-id>
          <pub-id pub-id-type="medline">27655477</pub-id>
          <pub-id pub-id-type="pii">10.1186/s13104-016-2253-6</pub-id>
          <pub-id pub-id-type="pmcid">PMC5031304</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref158">
        <label>158</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Beevi</surname>
              <given-names>FH</given-names>
            </name>
            <name name-style="western">
              <surname>Miranda</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Pedersen</surname>
              <given-names>CF</given-names>
            </name>
            <name name-style="western">
              <surname>Wagner</surname>
              <given-names>S</given-names>
            </name>
          </person-group>
          <article-title>An Evaluation of Commercial Pedometers for Monitoring Slow Walking Speed Populations</article-title>
          <source>Telemed J E Health</source>
          <year>2016</year>
          <month>05</month>
          <volume>22</volume>
          <issue>5</issue>
          <fpage>441</fpage>
          <lpage>9</lpage>
          <pub-id pub-id-type="doi">10.1089/tmj.2015.0120</pub-id>
          <pub-id pub-id-type="medline">26451900</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref159">
        <label>159</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Boolani</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Towler</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>LeCours</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Blank</surname>
              <given-names>H</given-names>
            </name>
            <name name-style="western">
              <surname>Larue</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Fulk</surname>
              <given-names>G</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of 6 Commercially Available Activity Monitors in Measuring Heart Rate, Caloric Expenditure, Steps Walked, and Distance Traveled</article-title>
          <source>Cardiopulmonary Physical Therapy Journal</source>
          <year>2019</year>
          <volume>30</volume>
          <issue>4</issue>
          <fpage>153</fpage>
          <lpage>161</lpage>
          <pub-id pub-id-type="doi">10.1097/cpt.0000000000000107</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref160">
        <label>160</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Chandrasekar</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Hensor</surname>
              <given-names>EM</given-names>
            </name>
            <name name-style="western">
              <surname>Mackie</surname>
              <given-names>SL</given-names>
            </name>
            <name name-style="western">
              <surname>Backhouse</surname>
              <given-names>MR</given-names>
            </name>
            <name name-style="western">
              <surname>Harris</surname>
              <given-names>E</given-names>
            </name>
          </person-group>
          <article-title>Preliminary concurrent validity of the Fitbit-Zip and ActiGraph activity monitors for measuring steps in people with polymyalgia rheumatica</article-title>
          <source>Gait Posture</source>
          <year>2018</year>
          <month>03</month>
          <volume>61</volume>
          <fpage>339</fpage>
          <lpage>345</lpage>
          <pub-id pub-id-type="doi">10.1016/j.gaitpost.2018.01.035</pub-id>
          <pub-id pub-id-type="medline">29427859</pub-id>
          <pub-id pub-id-type="pii">S0966-6362(18)30046-8</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref161">
        <label>161</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Haegele</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Brian</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Wolf</surname>
              <given-names>D</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of the Fitbit Zip for Measuring Steps for Adolescents With Visual Impairments</article-title>
          <source>Adapt Phys Activ Q</source>
          <year>2017</year>
          <month>04</month>
          <volume>34</volume>
          <issue>2</issue>
          <fpage>195</fpage>
          <lpage>200</lpage>
          <pub-id pub-id-type="doi">10.1123/apaq.2016-0055</pub-id>
          <pub-id pub-id-type="medline">28556761</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref162">
        <label>162</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Claes</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Buys</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Avila</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Finlay</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Kennedy</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Guldenring</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Budts</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Cornelissen</surname>
              <given-names>V</given-names>
            </name>
          </person-group>
          <article-title>Validity of heart rate measurements by the Garmin Forerunner 225 at different walking intensities</article-title>
          <source>J Med Eng Technol</source>
          <year>2017</year>
          <month>08</month>
          <day>04</day>
          <volume>41</volume>
          <issue>6</issue>
          <fpage>480</fpage>
          <lpage>485</lpage>
          <pub-id pub-id-type="doi">10.1080/03091902.2017.1333166</pub-id>
          <pub-id pub-id-type="medline">28675070</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref163">
        <label>163</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Støve</surname>
              <given-names>MP</given-names>
            </name>
            <name name-style="western">
              <surname>Haucke</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Nymann</surname>
              <given-names>ML</given-names>
            </name>
            <name name-style="western">
              <surname>Sigurdsson</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Larsen</surname>
              <given-names>BT</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of the wearable activity tracker Garmin Forerunner 235 for the assessment of heart rate during rest and activity</article-title>
          <source>J Sports Sci</source>
          <year>2019</year>
          <month>04</month>
          <day>17</day>
          <volume>37</volume>
          <issue>8</issue>
          <fpage>895</fpage>
          <lpage>901</lpage>
          <pub-id pub-id-type="doi">10.1080/02640414.2018.1535563</pub-id>
          <pub-id pub-id-type="medline">30326780</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref164">
        <label>164</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Mahendran</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Kuys</surname>
              <given-names>SS</given-names>
            </name>
            <name name-style="western">
              <surname>Downie</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Ng</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Brauer</surname>
              <given-names>SG</given-names>
            </name>
          </person-group>
          <article-title>Are Accelerometers and GPS Devices Valid, Reliable and Feasible Tools for Measurement of Community Ambulation After Stroke?</article-title>
          <source>Brain Impairment</source>
          <year>2016</year>
          <month>06</month>
          <day>08</day>
          <volume>17</volume>
          <issue>2</issue>
          <fpage>151</fpage>
          <lpage>161</lpage>
          <pub-id pub-id-type="doi">10.1017/brimp.2016.13</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref165">
        <label>165</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Alsubheen</surname>
              <given-names>SA</given-names>
            </name>
            <name name-style="western">
              <surname>George</surname>
              <given-names>AM</given-names>
            </name>
            <name name-style="western">
              <surname>Baker</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Rohr</surname>
              <given-names>LE</given-names>
            </name>
            <name name-style="western">
              <surname>Basset</surname>
              <given-names>FA</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of the vivofit activity tracker</article-title>
          <source>J Med Eng Technol</source>
          <year>2016</year>
          <month>08</month>
          <volume>40</volume>
          <issue>6</issue>
          <fpage>298</fpage>
          <lpage>306</lpage>
          <pub-id pub-id-type="doi">10.1080/03091902.2016.1193238</pub-id>
          <pub-id pub-id-type="medline">27266422</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref166">
        <label>166</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Woodman</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Crouter</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Bassett</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Fitzhugh</surname>
              <given-names>E</given-names>
            </name>
            <name name-style="western">
              <surname>Boyer</surname>
              <given-names>W</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of Consumer Monitors for Estimating Energy Expenditure and Activity Type</article-title>
          <source>Med Sci Sports Exerc</source>
          <year>2017</year>
          <month>02</month>
          <volume>49</volume>
          <issue>2</issue>
          <fpage>371</fpage>
          <lpage>377</lpage>
          <pub-id pub-id-type="doi">10.1249/MSS.0000000000001090</pub-id>
          <pub-id pub-id-type="medline">27580155</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref167">
        <label>167</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Simunek</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Dygryn</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Gaba</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Jakubec</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Stelzer</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Chmelik</surname>
              <given-names>F</given-names>
            </name>
          </person-group>
          <article-title>Validity of Garmin Vivofit and Polar Loop for measuring daily step counts in free-living conditions in adults</article-title>
          <source>Acta Gymnica</source>
          <year>2016</year>
          <month>9</month>
          <day>30</day>
          <volume>46</volume>
          <issue>3</issue>
          <fpage>129</fpage>
          <lpage>135</lpage>
          <pub-id pub-id-type="doi">10.5507/ag.2016.014</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref168">
        <label>168</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Šimůnek</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Dygrýn</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Jakubec</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Neuls</surname>
              <given-names>F</given-names>
            </name>
            <name name-style="western">
              <surname>Frömel</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Welk</surname>
              <given-names>G</given-names>
            </name>
          </person-group>
          <article-title>Validity of Garmin Vívofit 1 and Garmin Vívofit 3 for School-Based Physical Activity Monitoring</article-title>
          <source>Pediatr Exerc Sci</source>
          <year>2019</year>
          <month>02</month>
          <day>01</day>
          <volume>31</volume>
          <issue>1</issue>
          <fpage>130</fpage>
          <lpage>136</lpage>
          <pub-id pub-id-type="doi">10.1123/pes.2018-0019</pub-id>
          <pub-id pub-id-type="medline">30304974</pub-id>
          <pub-id pub-id-type="pii">pes.2018-0019</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref169">
        <label>169</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Ehrler</surname>
              <given-names>F</given-names>
            </name>
            <name name-style="western">
              <surname>Weber</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Lovis</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
          <article-title>Influence of Pedometer Position on Pedometer Accuracy at Various Walking Speeds: A Comparative Study</article-title>
          <source>J Med Internet Res</source>
          <year>2016</year>
          <month>10</month>
          <day>06</day>
          <volume>18</volume>
          <issue>10</issue>
          <fpage>e268</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://www.jmir.org/2016/10/e268/"/>
          </comment>
          <pub-id pub-id-type="doi">10.2196/jmir.5916</pub-id>
          <pub-id pub-id-type="medline">27713114</pub-id>
          <pub-id pub-id-type="pii">v18i10e268</pub-id>
          <pub-id pub-id-type="pmcid">PMC5073206</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref170">
        <label>170</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Höchsmann</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Knaier</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Eymann</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Hintermann</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Infanger</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Schmidt-Trucksäss</surname>
              <given-names>A</given-names>
            </name>
          </person-group>
          <article-title>Validity of activity trackers, smartphones, and phone applications to measure steps in various walking conditions</article-title>
          <source>Scand J Med Sci Sports</source>
          <year>2018</year>
          <month>07</month>
          <volume>28</volume>
          <issue>7</issue>
          <fpage>1818</fpage>
          <lpage>1827</lpage>
          <pub-id pub-id-type="doi">10.1111/sms.13074</pub-id>
          <pub-id pub-id-type="medline">29460319</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref171">
        <label>171</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>De Ridder</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>De Blaiser</surname>
              <given-names>C</given-names>
            </name>
          </person-group>
          <article-title>Activity trackers are not valid for step count registration when walking with crutches</article-title>
          <source>Gait Posture</source>
          <year>2019</year>
          <month>05</month>
          <volume>70</volume>
          <fpage>30</fpage>
          <lpage>32</lpage>
          <pub-id pub-id-type="doi">10.1016/j.gaitpost.2019.02.009</pub-id>
          <pub-id pub-id-type="medline">30798092</pub-id>
          <pub-id pub-id-type="pii">S0966-6362(18)31554-6</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref172">
        <label>172</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Boeselt</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Spielmanns</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Nell</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Storre</surname>
              <given-names>JH</given-names>
            </name>
            <name name-style="western">
              <surname>Windisch</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Magerhans</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Beutel</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Kenn</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Greulich</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Alter</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Vogelmeier</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Koczulla</surname>
              <given-names>AR</given-names>
            </name>
          </person-group>
          <article-title>Validity and Usability of Physical Activity Monitoring in Patients with Chronic Obstructive Pulmonary Disease (COPD)</article-title>
          <source>PLoS One</source>
          <year>2016</year>
          <volume>11</volume>
          <issue>6</issue>
          <fpage>e0157229</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://dx.plos.org/10.1371/journal.pone.0157229"/>
          </comment>
          <pub-id pub-id-type="doi">10.1371/journal.pone.0157229</pub-id>
          <pub-id pub-id-type="medline">27305105</pub-id>
          <pub-id pub-id-type="pii">PONE-D-16-05572</pub-id>
          <pub-id pub-id-type="pmcid">PMC4909270</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref173">
        <label>173</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Lee</surname>
              <given-names>JA</given-names>
            </name>
            <name name-style="western">
              <surname>Williams</surname>
              <given-names>SM</given-names>
            </name>
            <name name-style="western">
              <surname>Brown</surname>
              <given-names>DD</given-names>
            </name>
            <name name-style="western">
              <surname>Laurson</surname>
              <given-names>KR</given-names>
            </name>
          </person-group>
          <article-title>Concurrent validation of the Actigraph gt3x+, Polar Active accelerometer, Omron HJ-720 and Yamax Digiwalker SW-701 pedometer step counts in lab-based and free-living settings</article-title>
          <source>J Sports Sci</source>
          <year>2015</year>
          <month>12</month>
          <day>17</day>
          <volume>33</volume>
          <issue>10</issue>
          <fpage>991</fpage>
          <lpage>1000</lpage>
          <pub-id pub-id-type="doi">10.1080/02640414.2014.981848</pub-id>
          <pub-id pub-id-type="medline">25517396</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref174">
        <label>174</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Hernández-Vicente</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Santos-Lozano</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>De Cocker</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Garatachea</surname>
              <given-names>N</given-names>
            </name>
          </person-group>
          <article-title>Validation study of Polar V800 accelerometer</article-title>
          <source>Ann Transl Med</source>
          <year>2016</year>
          <month>08</month>
          <volume>4</volume>
          <issue>15</issue>
          <fpage>278</fpage>
          <lpage>278</lpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://doi.org/10.21037/atm.2016.07.16"/>
          </comment>
          <pub-id pub-id-type="doi">10.21037/atm.2016.07.16</pub-id>
          <pub-id pub-id-type="medline">27570772</pub-id>
          <pub-id pub-id-type="pii">atm-04-15-278</pub-id>
          <pub-id pub-id-type="pmcid">PMC4980380</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref175">
        <label>175</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Gruwez</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Libert</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Ameye</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Bruyneel</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <article-title>Reliability of commercially available sleep and activity trackers with manual switch-to-sleep mode activation in free-living healthy individuals</article-title>
          <source>Int J Med Inform</source>
          <year>2017</year>
          <month>06</month>
          <volume>102</volume>
          <fpage>87</fpage>
          <lpage>92</lpage>
          <pub-id pub-id-type="doi">10.1016/j.ijmedinf.2017.03.008</pub-id>
          <pub-id pub-id-type="medline">28495352</pub-id>
          <pub-id pub-id-type="pii">S1386-5056(17)30069-2</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref176">
        <label>176</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Wang</surname>
              <given-names>R</given-names>
            </name>
            <name name-style="western">
              <surname>Blackburn</surname>
              <given-names>G</given-names>
            </name>
            <name name-style="western">
              <surname>Desai</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Phelan</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Gillinov</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Houghtaling</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Gillinov</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <article-title>Accuracy of Wrist-Worn Heart Rate Monitors</article-title>
          <source>JAMA Cardiol</source>
          <year>2017</year>
          <month>01</month>
          <day>01</day>
          <volume>2</volume>
          <issue>1</issue>
          <fpage>104</fpage>
          <lpage>106</lpage>
          <pub-id pub-id-type="doi">10.1001/jamacardio.2016.3340</pub-id>
          <pub-id pub-id-type="medline">27732703</pub-id>
          <pub-id pub-id-type="pii">2566167</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref177">
        <label>177</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Roos</surname>
              <given-names>L</given-names>
            </name>
            <name name-style="western">
              <surname>Taube</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Beeler</surname>
              <given-names>N</given-names>
            </name>
            <name name-style="western">
              <surname>Wyss</surname>
              <given-names>T</given-names>
            </name>
          </person-group>
          <article-title>Validity of sports watches when estimating energy expenditure during running</article-title>
          <source>BMC Sports Sci Med Rehabil</source>
          <year>2017</year>
          <month>12</month>
          <day>20</day>
          <volume>9</volume>
          <issue>1</issue>
          <fpage>22</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://bmcsportsscimedrehabil.biomedcentral.com/articles/10.1186/s13102-017-0089-6"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/s13102-017-0089-6</pub-id>
          <pub-id pub-id-type="medline">29296281</pub-id>
          <pub-id pub-id-type="pii">89</pub-id>
          <pub-id pub-id-type="pmcid">PMC5738849</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref178">
        <label>178</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Evenson</surname>
              <given-names>KR</given-names>
            </name>
            <name name-style="western">
              <surname>Goto</surname>
              <given-names>MM</given-names>
            </name>
            <name name-style="western">
              <surname>Furberg</surname>
              <given-names>RD</given-names>
            </name>
          </person-group>
          <article-title>Systematic review of the validity and reliability of consumer-wearable activity trackers</article-title>
          <source>Int J Behav Nutr Phys Act</source>
          <year>2015</year>
          <month>12</month>
          <day>18</day>
          <volume>12</volume>
          <fpage>159</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://ijbnpa.biomedcentral.com/articles/10.1186/s12966-015-0314-1"/>
          </comment>
          <pub-id pub-id-type="doi">10.1186/s12966-015-0314-1</pub-id>
          <pub-id pub-id-type="medline">26684758</pub-id>
          <pub-id pub-id-type="pii">10.1186/s12966-015-0314-1</pub-id>
          <pub-id pub-id-type="pmcid">PMC4683756</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref179">
        <label>179</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Gerrior</surname>
              <given-names>S</given-names>
            </name>
            <name name-style="western">
              <surname>Juan</surname>
              <given-names>W</given-names>
            </name>
            <name name-style="western">
              <surname>Basiotis</surname>
              <given-names>P</given-names>
            </name>
          </person-group>
          <article-title>An easy approach to calculating estimated energy requirements</article-title>
          <source>Prev Chronic Dis</source>
          <year>2006</year>
          <month>10</month>
          <volume>3</volume>
          <issue>4</issue>
          <fpage>A129</fpage>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="http://www.cdc.gov/pcd/issues/2006/oct/06_0034.htm"/>
          </comment>
          <pub-id pub-id-type="medline">16978504</pub-id>
          <pub-id pub-id-type="pii">A129</pub-id>
          <pub-id pub-id-type="pmcid">PMC1784117</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref180">
        <label>180</label>
        <nlm-citation citation-type="web">
          <source>Open mHealth</source>
          <access-date>2020-08-17</access-date>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://www.openmhealth.org/">https://www.openmhealth.org/</ext-link>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref181">
        <label>181</label>
        <nlm-citation citation-type="web">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Ralls</surname>
              <given-names>J</given-names>
            </name>
          </person-group>
          <article-title>Fitbit to Be Acquired by Google</article-title>
          <source>Fitbit</source>
          <year>2019</year>
          <access-date>2019-11-01</access-date>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://investor.fitbit.com/press/press-releases/press-release-details/2019/Fitbit-to-Be-Acquired-by-Google/default.aspx">https://investor.fitbit.com/press/press-releases/press-release-details/2019/Fitbit-to-Be-Acquired-by-Google/default.aspx</ext-link>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref182">
        <label>182</label>
        <nlm-citation citation-type="web">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Osterloh</surname>
              <given-names>R</given-names>
            </name>
          </person-group>
          <article-title>Helping more people with wearables: Google to acquire Fitbit Internet</article-title>
          <source>Google Blog</source>
          <access-date>2020-08-17</access-date>
          <comment>
            <ext-link ext-link-type="uri" xlink:type="simple" xlink:href="https://blog.google/products/hardware/agreement-with-fitbit?_ga=2.109995341.918473813.1572613323-1996097189.1566566630">https://blog.google/products/hardware/agreement-with-fitbit?_ga=2.109995341.918473813.1572613323-1996097189.1566566630</ext-link>
          </comment>
        </nlm-citation>
      </ref>
      <ref id="ref183">
        <label>183</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Morrison</surname>
              <given-names>A</given-names>
            </name>
            <name name-style="western">
              <surname>Polisena</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Husereau</surname>
              <given-names>D</given-names>
            </name>
            <name name-style="western">
              <surname>Moulton</surname>
              <given-names>K</given-names>
            </name>
            <name name-style="western">
              <surname>Clark</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Fiander</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Mierzwinski-Urban</surname>
              <given-names>M</given-names>
            </name>
            <name name-style="western">
              <surname>Clifford</surname>
              <given-names>T</given-names>
            </name>
            <name name-style="western">
              <surname>Hutton</surname>
              <given-names>B</given-names>
            </name>
            <name name-style="western">
              <surname>Rabb</surname>
              <given-names>D</given-names>
            </name>
          </person-group>
          <article-title>The effect of English-language restriction on systematic review-based meta-analyses: a systematic review of empirical studies</article-title>
          <source>Int J Technol Assess Health Care</source>
          <year>2012</year>
          <month>04</month>
          <volume>28</volume>
          <issue>2</issue>
          <fpage>138</fpage>
          <lpage>44</lpage>
          <pub-id pub-id-type="doi">10.1017/S0266462312000086</pub-id>
          <pub-id pub-id-type="medline">22559755</pub-id>
          <pub-id pub-id-type="pii">S0266462312000086</pub-id>
        </nlm-citation>
      </ref>
      <ref id="ref184">
        <label>184</label>
        <nlm-citation citation-type="journal">
          <person-group person-group-type="author">
            <name name-style="western">
              <surname>Jüni</surname>
              <given-names>P</given-names>
            </name>
            <name name-style="western">
              <surname>Holenstein</surname>
              <given-names>F</given-names>
            </name>
            <name name-style="western">
              <surname>Sterne</surname>
              <given-names>J</given-names>
            </name>
            <name name-style="western">
              <surname>Bartlett</surname>
              <given-names>C</given-names>
            </name>
            <name name-style="western">
              <surname>Egger</surname>
              <given-names>M</given-names>
            </name>
          </person-group>
          <article-title>Direction and impact of language bias in meta-analyses of controlled trials: empirical study</article-title>
          <source>Int J Epidemiol</source>
          <year>2002</year>
          <month>02</month>
          <volume>31</volume>
          <issue>1</issue>
          <fpage>115</fpage>
          <lpage>23</lpage>
          <pub-id pub-id-type="doi">10.1093/ije/31.1.115</pub-id>
          <pub-id pub-id-type="medline">11914306</pub-id>
        </nlm-citation>
      </ref>
    </ref-list>
  </back>
</article>
