Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study

Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study

Asterisk indicates P=.03; # indicates P MAPE (SD) for Garmin and Fitbit was 27.0% (SD 21.8) and 25.1% (SD 17.3), respectively. The lowest mean error in measuring EE was observed during ADL (−8.8% [SD 29.2]) for Fitbit and MAX-C (−4.5% [SD 25.3]) and HIIT-T (−4.7% [SD 29.3]) for Garmin. The highest error was observed during MAX-C (−39.1% [SD 30.6]) and HIIT-C (−41.9% [SD 31.3]) for Fitbit and resistance (21.0% [SD 35.7]) for Garmin.

Ravi Kondama G Reddy, Rubin Pooni, Dessi P Zaharieva, Brian Senf, Joseph El Youssef, Eyal Dassau, Francis J Doyle III, Mark A Clements, Michael R Rickels, Susana R Patton, Jessica R Castle, Michael C Riddell, Peter G Jacobs

JMIR Mhealth Uhealth 2018;6(12):e10338