Impact of Remote Blood Pressure Monitoring Device Connectivity on Engagement Among Pregnant Individuals Enrolled in the Delfina Care Platform: Observational Study

Impact of Remote Blood Pressure Monitoring Device Connectivity on Engagement Among Pregnant Individuals Enrolled in the Delfina Care Platform: Observational Study

Impact of Remote Blood Pressure Monitoring Device Connectivity on Engagement Among Pregnant Individuals Enrolled in the Delfina Care Platform: Observational Study

1Delfina Care Inc, , San Francisco, CA, , United States

2Department of Population Health, New York University, , New York, NY, , United States

3Department of Obstetrics, Gynecology, and Reproductive Sciences, University of California San Francisco, , San Francisco, CA, , United States

4Center for Women's Health, , Richmond, TX, , United States

Corresponding Author:

Mia Charifson, BA, MA


User engagement with remote blood pressure monitoring during pregnancy is critical to optimize the associated benefits of blood pressure control and early detection of hypertensive disorders of pregnancy. In our study population of pregnant individuals, we found that connected blood pressure cuffs, which automatically sync measures to a monitoring platform or health record, increase engagement (2.13 [95% CI 1.36‐3.35] times more measures per day) with remote blood pressure monitoring compared to unconnected cuffs that require manual entry of measures.

JMIR Mhealth Uhealth 2024;12:e55617

doi:10.2196/55617

Keywords



Hypertensive disorders of pregnancy affect 5%‐10% of pregnancies and increase the risk of adverse pregnancy and postpartum outcomes [1,2]. Typically, management of these disorders involves blood pressure (BP) monitoring and initiation of antihypertensive therapy. Remote BP monitoring (RBPM) enables at-home BP monitoring to inform clinical decision-making in a timely manner [2-4]. Previous studies have shown that RBPM facilitates earlier detection of elevated BP and reduces prenatal hospitalizations and clinic visits [2,3]. Despite studies establishing feasibility and patient acceptability of RBPM [4-9], best practices for integrating RBPM within prenatal care have not yet been established. One open question is whether connected BP cuffs, which automatically sync measures to a health record, improve data quality and frequency over standard BP cuffs, which require users to manually enter measures [3-8]. The goal of this study was to compare user engagement with RBPM between connected and unconnected BP devices among users of a pregnancy care platform.


Study Population

The study population consisted of pregnant individuals enrolled in Delfina Care [10], a comprehensive pregnancy care platform, at a community practice in Texas, USA, between January and July 2023. Initially, these users were provided with unconnected BP devices at their provider’s discretion, with a recommendation to record 2 BP measures a day per internal expert clinical consensus. Connected devices were introduced in April 2023 as part of a quality improvement initiative. Differences in user experiences between the connected and unconnected device groups are further described in Multimedia Appendix 1.

The outcome of interest was user engagement, defined as the number of daily BP measures taken, and as a binary indicator of completing the daily recommended BP measures (ie, 2 distinct BP measures per day). Analogous engagement outcomes at the weekly level were also reported. To avoid inflated differences due to repeated measures from connected device users, we considered multiple entries within 1 hour as a single measurement (Multimedia Appendix 2). The exposure of interest was receiving a connected versus unconnected device. Clinical and demographic characteristics were collated from user-reported questionnaire data and electronic health records.

Statistical Analysis

Poisson and logistic regressions were fit for the daily number of BP measures and ≥2 daily BP measures, respectively. Both models controlled for relevant confounders (ie, age, parity, weeks since enrollment, primary clinic, and preferred language) and included a random effect for users across enrollment days.

Ethical Considerations

The study team received an institutional review board exemption waiver of HIPAA (Health Insurance Portability and Accountability Act) authorization on August 22, 2022, from WIRB-Copernicus Group (WCG) Institutional Review Board (protocol number 202208‐001).


During the study period, 164 users with BP cuffs were enrolled in the Delfina Care platform. Restricting to those with covariate data, the analytic sample consisted of 163 users (97 unconnected device users and 66 connected device users). Users with connected devices had more mean BP measure entries per day (0.51 vs 0.32) and a higher proportion completed the recommended ≥2 daily BP measures (12% vs 7%) compared to users with unconnected devices (Table 1). At the weekly level, the mean entries per week (3.37 vs 2.07) and the proportion of users who completed the recommended ≥1 weekly BP measure (63% vs 47%) were higher for connected users than for unconnected device users (Table 1).

Adjusting for confounders, users with connected devices had 2.13 (95% CI 1.36‐3.35) times more measures per day and 5.62 (95% CI 2.28‐13.83) times the odds of meeting the recommendation of ≥2 daily BP measures than unconnected device users (Table 2).

Table 1. Study population characteristics.
CharacteristicsOverall (n=163)Unconnected device users (n=97)Connected device users (n=66)P value
Age in years, mean (SD)28.73 (6.02)28.71 (5.61)28.76 (6.61).96
Language, n (%).32
English139 (85.3)80 (82.5)59 (89.4)
Spanish24 (14.7)17 (17.5)7 (10.6)
Parity, mean (SD)1.37 (1.44)1.36 (1.32)1.38 (1.60).99
Primary clinic, n (%).07
A35 (21.5)20 (20.6)15 (22.7)
B71 (43.6)49 (50.5)22 (33.3)
C57 (35.0)28 (28.9)29 (43.9)
User engagementa, mean (SD)
Weekly entries2.59 (3.02)2.07 (2.78)3.37 (3.20).007
Daily entries0.40 (0.46)0.32 (0.42)0.51 (0.49).008
Proportion ≥1 weekly entries0.53 (0.35)0.47 (0.35)0.63 (0.33).003
Proportion ≥2 daily entries0.09 (0.17)0.07 (0.15)0.12 (0.19).048

aUser engagement metrics were first averaged within a user and then averaged across users for each device type and overall.

Table 2. Effect estimates and 95% CIs by outcome model.
Number of BP measures per dayCompleted ≥2 BP measures per day
PredictorsIncidence rate ratio95% CIOdds ratio95% CI
Connected versus unconnected device2.131.36‐3.355.622.28‐13.83
Age (years)1.061.02‐1.111.111.03‐1.20
Weeks since enrollment0.970.97‐0.981.000.98‐1.01
English vs Spanish language1.250.66‐2.350.720.22‐2.38
Parity (number of live births)0.880.74‐1.050.790.57‐1.10
Clinic B vs A1.440.79‐2.614.181.25‐14.03
Clinic C vs A1.540.83‐2.864.061.16‐14.20

We observed user engagement with RBPM was significantly higher among those with connected devices than those with unconnected devices. Previous studies among pregnant individuals have shown that the recommended frequency of BP measures ranges from several times daily to once weekly [4,7,8,11]. In our cohort, the proportion of users meeting the twice-daily recommendation was low, but the majority of users completed the once-weekly entry at least, and connected device users still had higher utilization than unconnected device users (63% vs 47%; P=.003). Compared to in-clinic BP measures, all users had a higher average number of weekly readings than the in-clinic average (2.59 vs 0.50 BP readings/week). Our study corroborated the feasibility of at-home RBPM during pregnancy and highlights the potential advantages of device connectivity on user engagement.

Limitations

Our findings are limited by the lack of true randomization to device types, which we addressed by controlling for potential confounders. We also addressed potential time-related confounding via a sensitivity analysis with no change in findings (Multimedia Appendix 2).

Conclusion

This study highlights how connected BP devices can improve patient engagement to RBPM during the prenatal period. Other aspects of RBPM, such as recommended frequency and patient education, should be further investigated to ensure users are able to successfully engage with monitoring technologies.

Acknowledgments

This study was supported by grant funding from the Eunice Kennedy Shriver National Institute of Child Health and Human Development (NICHD) under grant number R43HD114360.

Conflicts of Interest

MC reports serving as a data science consultant for Delfina, Inc. TW reports serving as the associate chief medical officer for Delfina, Inc. BZ reports serving as the chief medical officer for Delfina, Inc. PV reports serving as the chief product officer for Delfina, Inc. IF reports serving as the chief scientific officer for Delfina, Inc

Multimedia Appendix 1

Additional information on Delfina Care.

DOCX File, 344 KB

Multimedia Appendix 2

Additional analyses.

DOCX File, 15 KB

  1. Hutcheon JA, Lisonkova S, Joseph KS. Epidemiology of pre-eclampsia and the other hypertensive disorders of pregnancy. Best Pract Res Clin Obstet Gynaecol. Aug 2011;25(4):391-403. [CrossRef] [Medline]
  2. Nathan HL, Duhig K, Hezelgrave NL, Chappell LC, Shennan AH. Blood pressure measurement in pregnancy. Obstet Gynaecol. Apr 2015;17(2):91-98. [CrossRef]
  3. Yeh PT, Rhee DK, Kennedy CE, et al. Self-monitoring of blood pressure among women with hypertensive disorders of pregnancy: a systematic review. BMC Pregnancy Childbirth. May 31, 2022;22(1):454. [CrossRef] [Medline]
  4. Aquino M, Munce S, Griffith J, Pakosh M, Munnery M, Seto E. Exploring the use of telemonitoring for patients at high risk for hypertensive disorders of pregnancy in the antepartum and postpartum periods: scoping review. JMIR Mhealth Uhealth. Apr 17, 2020;8(4):e15095. [CrossRef] [Medline]
  5. van den Heuvel JFM, Kariman SS, van Solinge WW, Franx A, Lely AT, Bekker MN. SAFE@HOME - feasibility study of a telemonitoring platform combining blood pressure and preeclampsia symptoms in pregnancy care. Eur J Obstet Gynecol Reprod Biol. Sep 2019;240:226-231. [CrossRef] [Medline]
  6. Marko KI, Krapf JM, Meltzer AC, et al. Testing the feasibility of remote patient monitoring in prenatal care using a mobile app and connected devices: a prospective observational trial. JMIR Res Protoc. Nov 18, 2016;5(4):e200. [CrossRef] [Medline]
  7. Tucker KL, Taylor KS, Crawford C, et al. Blood pressure self-monitoring in pregnancy: examining feasibility in a prospective cohort study. BMC Pregnancy Childbirth. Dec 28, 2017;17(1):442. [CrossRef] [Medline]
  8. Runkle JD, Sugg MM, McCrory S, Coulson CC. Examining the feasibility of smart blood pressure home monitoring: advancing remote prenatal care in rural Appalachia. Telemed Rep. 2021;2(1):125-134. [CrossRef] [Medline]
  9. Khalil A, Perry H, Lanssens D, Gyselaers W. Telemonitoring for hypertensive disease in pregnancy. Expert Rev Med Devices. Aug 2019;16(8):653-661. [CrossRef] [Medline]
  10. Delfina - intelligent pregnancy care for all. URL: https://www.delfina.com/ [Accessed 2024-06-28]
  11. Chappell LC, Tucker KL, Galal U, et al. Effect of self-monitoring of blood pressure on blood pressure control in pregnant individuals with chronic or gestational hypertension: the BUMP 2 randomized clinical trial. JAMA. May 3, 2022;327(17):1666-1678. [CrossRef] [Medline]


BP: blood pressure
HIPAA: Health Insurance Portability and Accountability Act
NICHD: National Institute of Child Health and Human Development
RBPM: remote blood pressure monitoring
WCG: WIRB-Copernicus Group


Edited by Lorraine Buis; submitted 19.12.23; peer-reviewed by Benjamin Senst, Chu-Pak Lau; final revised version received 03.05.24; accepted 12.06.24; published 12.07.24.

Copyright

© Mia Charifson, Timothy Wen, Bonnie Zell, Priyanka Vaidya, Cynthia I Rios, C Funsho Fagbohun, Isabel Fulcher. Originally published in JMIR mHealth and uHealth (https://mhealth.jmir.org), 12.7.2024.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mHealth and uHealth, is properly cited. The complete bibliographic information, a link to the original publication on https://mhealth.jmir.org/, as well as this copyright and license information must be included.