Original Paper
Abstract
Background: Sleep is essential for one’s health and quality of life. Wearable technologies that use motion and temperature sensors have made it possible to self-monitor sleep. Although there is a growing body of research on sleep monitoring using wearable devices for healthy young-to-middle-aged adults, few studies have focused on older adults, including those living with dementia.
Objective: This study aims to investigate the impact of age and dementia on sleep detection through movement and wrist temperature.
Methods: A total of 10 younger adults, 10 healthy older adults, and 8 older adults living with dementia (OAWD) were recruited. Each participant wore a Mi Band 2 (accemetry-based sleep detection) and our custom-built wristband (actigraphy and wrist temperature) 24 hours a day for 2 weeks and was asked to keep a daily sleep journal. Sleep parameters detected by the Mi Band 2 were compared with sleep journals, and visual analysis of actigraphy and temperature data was performed.
Results: The absolute differences in sleep onset and offset between the sleep journals and Mi Band 2 were 39 (SD 51) minutes and 31 (SD 52) minutes for younger adults, 49 (SD 58) minutes and 33 (SD 58) minutes for older adults, and 253 (SD 104) minutes and 161 (SD 94) minutes for OAWD. The Mi Band 2 was unable to accurately detect sleep in 3 healthy older adults and all OAWDs. The average sleep and wake temperature difference of OAWD (1.26 °C, SD 0.82 °C) was significantly lower than that of healthy older adults (2.04 °C, SD 0.70 °C) and healthy younger adults (2.48 °C, SD 0.88 °C). Actigraphy data showed that older adults had more movement during sleep compared with younger adults and that this trend appears to increase for those with dementia.
Conclusions: The Mi Band 2 did not accurately detect sleep in older adults who had greater levels of nighttime movement. As more nighttime movement appears to be a phenomenon that increases in prevalence with age and even more so with dementia, further research needs to be conducted with a larger sample size and greater diversity of commercially available wearable devices to explore these trends more conclusively. All participants, including older adults and OAWD, had a distinct sleep and wake wrist temperature contrast, which suggests that wrist temperature could be leveraged to create more robust and broadly applicable sleep detection algorithms.
doi:10.2196/26462
Keywords
Introduction
On average, we spend one-third of our lives asleep [
]. Quality and quantity of sleep regulate daytime behaviors and functions as well as significantly impact health and well-being. Sleep has been shown to affect day-to-day memory and concentration [ - ]. Lifetime sleep habits appear to correlate with one’s likelihood of having conditions, such as Alzheimer disease [ - ] and cardiovascular diseases [ , ]. Although the benefits of getting a good night’s rest have received extensive research, there is still much we do not know about the differences between how people sleep and what can be done to support better sleep.Understanding sleep patterns is a complex undertaking, as sleep is impacted by conscious and subconscious control by the individual as well as environmental factors. Standard clinical sleep evaluation typically uses polysomnography (PSG), which is considered the gold standard in sleep studies [
]. PSG requires the person being assessed to sleep in a laboratory with devices attached to the body in single or infrequent sessions. Although the data collected by PSG can be used to diagnose many medical conditions, its intrusive, unnatural nature might not reflect people’s usual sleep quality and is not appropriate for long-term or frequent monitoring of sleep [ ].Actigraphy (ie, accelerometer sensors and gyroscopes) is being increasingly used to measure people’s sleep and activities by estimating related parameters, such as sleep onset and offset [
, ]. Many clinical studies have adopted actigraphy to measure circadian rhythm, which is a major factor in regulating people’s sleep and wake rhythms [ , ]. Driven by circadian rhythms, many biological processes, including core body temperature (CBT), have 24-hour diurnal variations. During sleep, people’s CBT drops about 1 °C [ ]. Thermal regulation inside the human body causes wrist temperature to exhibit an opposite pattern to CBT [ ]; wrist temperature increases before people fall asleep, remains at a relatively high level when people are asleep, and then drops in the morning when people wake up ( ). CBT has been used extensively to study circadian rhythms; however, it usually requires invasive gut or rectal temperature measurements [ ]. As an alternative, wrist temperature can be measured in daily life and is found to be more correlated with sleeping status than CBT [ ]. Recent studies have provided increasing evidence that wrist temperature can be used for sleep monitoring [ , - ].Low sleep quality and irregular sleep patterns become more prevalent with increasing age and dementia [
- ]. For example, poor nighttime sleep has been shown to be a key factor in older adults’ tendency to have more daytime sleepiness [ ], which has been associated with changes in circadian rhythms in older adults [ ]. Sleep can also be used as an indicator of comorbidities, such as sleep apnea and restless leg syndrome, both of which become more prevalent with increasing age [ , ]. Disrupted sleep patterns have been found to be even more prevalent in people with dementia [ ]. Poor sleep has been found to be predictive of more severely impacted cognition in older adults, including people with dementia [ ] and older women without dementia [ ]. Therefore, the ability to monitor older adults’ sleep and circadian rhythms on an ongoing basis is increasingly being used to understand and support their sleep, which, in turn, supports health and well-being [ ].The proliferation of smart wearable technologies has contributed to the substantial growth in technology adoption among older adults; older adults are the most rapidly increasing group of users of new technologies [
, ]. One of the most commonly adopted technologies is smart wearable devices, which usually have built-in accelerometer and gyroscope sensors to detect people’s sleep (eg, Fitbit, Garmin, and Mi Band) [ - ]. Commercial smart wristbands have been designed to output several sleep parameters, including sleep onset, offset, duration, and wake after sleep onset, and some wristbands also give a sleep quality score. Sleep parameters can be synchronized to one’s smartphone via Bluetooth (and then to the cloud, should they wish) so that users can easily access their sleep data. Relative affordability and unobtrusiveness contribute to the growing popularity of commercial smart wristbands. In addition, smart wearable devices are being increasingly and widely used in health research [ - ]. For example, Gibson et al [ ] evaluated the reliability of actigraphy to measure the sleep of people living with dementia.Although most smart wristbands can provide sleep monitoring and analysis, they fail to monitor people’s sleep from an internal perspective; they cannot measure circadian rhythm directly. As discussed earlier, wrist temperature is a reflection of CBT, which is indicative of a person’s biological circadian rhythm. As such, wrist temperature could have value as an addition to accemetry-based wearable sleep monitoring systems, which could then be used to support better understanding and management of sleep. To be successful, these systems should work for different stakeholder groups, including older adults and older adults living with dementia (OAWD). Therefore, this research was guided by the question, “How do sleep and wrist temperature patterns compare for younger adults and older adults, including OAWD?”
Methods
Participants
This study measured sleep patterns and wrist temperature patterns of 3 groups of participants: (1) healthy younger adults (aged 20-30 years), (2) healthy older adults (aged≥65 years), and (3) OAWD. Inclusion criteria were any gender, any race, age in the targeted populations, and ability to understand English. An additional criterion for OAWD was early- to middle-stage dementia. Individuals diagnosed with severe sleep disorders were excluded from the study.
We recruited 10 participants for each group based on the convenience sampling method. Healthy adults were recruited through university-wide poster advertisements and word of mouth. Healthy older adults were recruited through phone calls and email invitations. OAWD participants were recruited from a local long-term care (LTC) residence. The staff at the LTC setting requested that they coordinate the initial identification of potential participants and support the recruitment process. Each potential participant was then contacted and introduced to the study by the researcher. Assent to continue participating in the study was obtained from each OAWD participant every time they were contacted by the researchers. This study was approved by the Human Research Ethics Committee of the University of Waterloo (ORE #31860 and ORE #40459).
Equipment
Each participant was asked to wear a commercially available smart wristband (Mi Band 2, XiaoMi;
) and a custom-built wristband developed by the authors ( ) on their left arm for 14 days. The Mi Band 2 was chosen because of its popularity, low cost, long battery life, and accessibility of actigraphy data. The custom-built band consisted of a temperature sensor (iButton DS1922L, Maxim) that maintained contact with the anterior side of the wrist and a 3-axis accelerometer sensor (AX3, Axivity). Both sensors operated offline by storing data locally; data were extracted after the 2-week study period using USB adaptors. All participants were offered to keep the Mi Band 2 if they wished on completion of the study.Study Protocol
The study protocol was the same for healthy younger adults and healthy older adults, with an altered OAWD protocol. Before the start of the study, all healthy participants were asked to complete 4 questionnaires: (1) a demographic form; (2) morningness-eveningness questionnaire, which is a measure of circadian rhythm type [
]; (3) Pittsburgh Sleep Quality Index (PSQI), which is a measure of subjective sleep quality [ ]; and (4) Epworth Sleepiness Scale (ESS), which is a measure of daytime sleepiness [ ]. Participants were then given the 2 wristbands; instructed on how to wear them; and asked to wear both on their left wrist for 14 days, except when showering or bathing. Participants were also asked to fill out a daily sleep journal that was adapted from a study by Monk et al [ ] to capture sleep onset, offset, and subjective sleep quality.As OAWD were not able to fill out the questionnaires, only the demographic form was completed and was done with the help of staff from their LTC residence. In addition, the Mini-Mental State Examination (MMSE) [
] was conducted to approximate the level of cognition for each OAWD. After giving each OAWD the wristbands, their personal support workers (PSWs) were asked to help with getting wristbands off and on when they assisted OAWD shower or bathe as well as to routinely check that the person was not adversely affected by wearing the wristbands. In addition, PSWs were asked to observe participants’ sleep status and check off either awake or sleeping on a sheet every half an hour throughout the day and night as part of their routine in with the OAWD. Sleep status from this sheet was then used to generate an observed sleep journal.Ground Truth of Sleep Parameters
Sleep onset and offset, the number of nighttime wake-ups, and sleep quality scores were obtained from the sleep journals and were used to help interpret actigraphy and temperature data. Although we thought it was unlikely that people could estimate their sleep at a minute level, we considered sleep journals of younger adults and healthy older adults reliable at the half-hour interval level. Occasionally, participants would miss reporting sleep onset and offset. In these cases, the sleep onset and offset were manually identified from the data from the custom-built wristband AX3 actigraphy data.
For OAWD, the intention was to use the sleep status sheet filled out by PSWs to extract sleep parameters and use these as ground truths for sleep. However, at the end of the study, it was found that, on average, 40% of the sheet was filled (ie, the PSWs did not report sleep status of OAWD 60% of the time). Owing to the large amount of data missing from the OAWD sleep journals, we chose to manually identify sleep periods from the AX3 data and combined those identified sleep periods with PSW sheets to estimate sleep for all OAWD.
Feature Extractions
Sleep Parameter Extraction
Sleep onset, sleep offset, and sleep duration from the sleep journals were compared with the Mi Band 2 for daily sleep monitoring; the mean values across the 14 days were calculated and are presented in
. The absolute differences between the Mi Band 2 and journal parameters were used to indicate the agreement of sleep detection between Mi Band 2 and self-reported sleep.Sleep parameters | Younger adults | Older adults | Older adults living with dementia |
Onset, mean (SD) | 12:32 AM (53 min) | 11:49 PM (61 min) | 9:37 PM (105 min) |
Offset, mean (SD) | 7:55 AM (54 min) | 7:31 AM (78 min) | 7:05 AM (40 min) |
Duration (min), mean (SD) | 442 (26) | 462 (88) | 562 (123) |
Sleep Periods Identification of OAWD
As people generally have minimal movements during sleep, sleep periods can be manually identified by segments of raw AX3 data that have minimal variations. For one night’s sleep, multiple sleep periods could be identified if the person had multiple wake-ups. Once sleep periods were estimated, the sleep onset, offset, and duration were calculated for the longest least varied data period. If participants woke up during the night and remained awake for a long period (ie, more than 2 hours), this period was not counted as sleep duration. AX3-estimated sleep was combined with the partially completed observational sheet by PSWs to create a recreated sleep journal for OAWD.
Wrist Temperature Feature Extraction
As participants took the customized wristband off to shower, all data points lower than 28 °C were removed and interpolated using the nearest values. The temperature data were then smoothed by applying a median filter with a window size of 3. Descriptive statistics were calculated to provide a profile of the general characteristics of the participants’ wrist temperature patterns and rhythm indices related to circadian rhythms. The mean and SD were calculated for the wrist temperature rhythm of every 24-hour period that began at 3 PM because this segmentation included full nighttime sleep periods and daytime naps. On the basis of sleep onset and sleep offsets obtained from sleep journals, the mean temperatures were calculated for sleep time and wake time. The mean wake time temperature was subtracted from the mean sleep time temperature to calculate the mean sleep and wake temperature difference, which was used as a measure of the extent of temperature changes between the sleep and wake states in this study.
Interday stability (IS) and intraday variability (IV) were calculated to examine the regularity and rhythmicity of wrist temperature rhythms for each participant [
]. IS has been used to measure the stability of rhythms across several consecutive days, whereas IV reflects the fragmentation of the wrist temperature rhythm of every 24-hour period. IS and IV were calculated using Equations (1) and (2), respectively:where N is the total number of wrist temperature data points, p is the number of wrist temperature data points per day (in this study, P=288 as the iButton is sampled every 5 min), X is the mean of all temperature data, Xh is the hourly mean of wrist temperature, and Xi is the data point at time i. IS and IV were calculated for each participant for the total study period.
Statistical Analysis
One-tailed t test was used to compare the sleep parameters and rhythm indices between different groups (young vs old and old vs OAWD). The Pearson correlation coefficient was used to assess correlations between different circadian rhythms and sleep parameters calculated; P<.05 was considered to be significant.
Case Study Analysis
When analyzing the collected data, it was found that the sleep of some participants was not correctly reported by the Mi Band 2. For example, most PSWs reported that OAWD had 8 hours of sleep at night, but the Mi Band 2 detected as few as 40 minutes of sleep. The AX3 data from the custom-built band were inspected for the periods Mi Band 2 had classified as wake periods during the night as well as wrist temperature measurements for the same time. A healthy younger participant who had good sleep quality; a healthy older participant with poor sleep quality; and 2 OAWD who had poor, interrupted sleep were selected as case studies to illustrate this phenomenon and are presented later in this paper.
Results
Demographics and Sleep Patterns
In total, we recruited 10 healthy younger adults, 10 healthy older adults, and 8 OAWD. A total of 7 OAWD were able to give their own consent and sign the consent form; for the 1 participant who could not self-consent, his power of attorney signed the consent form, and assent was obtained from the participant. Participants’ demographic data are presented in
. The average MMSE score of the 8 OAWD participants was 20. Moreover, 7 OAWD were in the mild cognitive impairment category (ie, a score between 18 and 23), and 1 OAWD had a score of 16, which indicates severe cognitive impairment [ ].Demographic | Healthy younger adult group (n=10) | Healthy older adult group (n=10) | Older adults living with dementia group (n=8) | |
Age (years), mean (SD) | 24.1 (2.23) | 75 (7.36) | 83.25 (10.38) | |
Sex, n (%) | ||||
Male | 4 (40) | 4 (40) | 4 (50) | |
Female | 6 (60) | 6 (60) | 4 (50) | |
BMI (kg/m2), mean (SD) | 23.69 (6.30) | 25.94 (3.11) | 29.90 (8.91) | |
Pittsburgh Sleep Quality Index, mean (SD)a | 5.2 (2.39) | 7.8 (3.56) | —b | |
Mini-Mental State Examination, mean (SD)c | — | — | 20 (2.39) |
aThe Pittsburgh Sleep Quality Index was not measured for the older adults living with dementia group as they cannot reliably complete this measure.
bTest was not administered as the test was not appropriate for the participant group.
cThe Mini-Mental State Examination of the healthy younger adult group and the healthy older adult group as they were cognitively intact.
The sleep parameters obtained from the sleep journals for all participants are presented in
. Compared with older adult groups, healthy younger participants tended to sleep later and wake up later, with the shortest average sleep duration among the 3 groups. Between the 2 older adult groups, OAWD slept earlier and had a longer sleep duration. Although we did not require participants to record daytime naps, we asked them to report whether they had the habit of napping. Only 1 younger adult and 1 healthy older adult reported that they took regular noontime naps. All OAWD napped during the day, sometimes more than once. As the recreated sleep journals of OAWD only summarized nighttime sleep and did not capture daytime naps, the total sleep duration of OAWD might be longer than reported. Healthy older adults had better compliance in reporting and provided more detailed sleep journals (eg, a few participants consistently recorded their nap time during the day).Comparison Between Mi Band 2 and Sleep Journal
The absolute differences in sleep parameters between the Mi Band 2 and sleep journals are summarized in
, and significant differences were found in sleep duration between the younger adult group and the healthy older adult group. Between the 2 older adult groups, the absolute differences in sleep onset, offset, and duration of OAWD were significantly higher than those in the healthy older adult group. As shown in , of the 106 days of data measured from the 8 OAWD, for 28 of the days (28/106, 26% of days monitored), the Mi Band 2 did not detect any sleep, and we confirmed that all OAWDs were wearing the Mi Band 2 correctly throughout the study. Discrepancies between sleep parameters reported in the recreated sleep journal and detected by Mi Band 2 for OAWD were large, namely, sleep onsets detected by Mi Band 2 were often noticeably delayed (ie, on the order of hours), and the offset was earlier than what was reported. The average absolute difference in sleep duration between the Mi Band 2 data and the recreated sleep journal for OAWD was more than 6 hours. A discussion of this detection error is presented later in the Case Studies section.Sleep parameters | Healthy younger adults | Healthy older adults | Older adults living with dementia | |
Onset | ||||
Difference (min), mean (SD) | 39 (51) | 49 (58) | 253 (104) | |
P value | —a | — | <.001 | |
Offset | ||||
Difference (min), mean (SD) | 31 (52) | 33 (58) | 161 (94) | |
P value | — | — | <.001 | |
Duration | ||||
Difference (min), mean (SD) | 49 (59) | 64 (77) | 379 (163) | |
P value | — | .04 | <.001 |
aNo significance.
ID | Total days (n) | Valid days, n (%) |
OAWDa 1 | 14 | 14 (100) |
OAWD 2 | 13 | 2 (15) |
OAWD 3 | 10 | 10 (100) |
OAWD 4 | 14 | 7 (50) |
OAWD 5 | 14 | 12 (86) |
OAWD 6 | 13 | 8 (62) |
OAWD 7 | 14 | 11 (79) |
OAWD 8 | 14 | 14 (100) |
Total | 106 | 78 (74) |
aOAWD: older adults living with dementia.
Wrist Temperature Rhythm Comparison
The average wrist temperature data of the 3 groups are shown in
. Although all curves have higher daytime (awake) temperatures, the OAWD group has the flattest temperature curve with the lowest nighttime wrist temperature and the highest daytime wrist temperature compared with the other groups. Although the average nighttime wrist temperature was similar for healthy older and younger adults, older adults tended to have a higher daytime wrist temperature, causing a flatter average curve for healthy older adults compared with younger adults. The wrist temperature indices are presented in .Parameters | Healthy younger adults (n=10) | Healthy older adults (n=10) | Older adults living with dementia (n=8) | ||||
Average wrist temperature parameters (°C), mean (SD) | |||||||
Sleep temperature | 34.98 (0.55) | 34.97 (0.34) | 34.66 (0.53) | ||||
Wake temperature | 32.49 (0.83) | 32.94 (0.67) | 33.40 (1.03) | ||||
Sleep and wake temperature difference | 2.48 (0.88) | 2.04 (0.70) | 1.26 (0.82; P=.02) | ||||
Cosinor analysis | |||||||
MESORa (°C), mean (SD) | 33.34 (0.61) | 33.65 (0.46) | 33.90 (0.84) | ||||
Amplitude (°C), mean (SD) | 1.72 (0.57) | 1.45 (0.50) | 0.93 (0.59; P=.03) | ||||
Acrophase (min), mean (SD) | 246 (59) | 183 (91; P=.04) | 44 (145; P<.001) | ||||
Nonparametric analyses (°C), mean (SD) | |||||||
Interday stability | 0.53 (0.10) | 0.52 (0.16) | 0.32 (0. 19; P=.02) |
aMESOR: midline-estimating statistic of rhythm.
Case Studies
Data from 4 participants are presented below as illustrative case studies. One of the case studies is of a younger healthy sleeper, and the other 3 cases are representative data from irregular sleepers, 1 of which is a healthy older adult and the other 2 are OAWD.
Case I: Healthy Younger Adult With Regular Sleep
The healthy younger adult (YA1) was a 25-year-old man (PSQI=5 and ESS=3) who did not report any sleep disorders. One day of YA1 data, including wrist temperature, AX3 data, and sleep onset or offset detected by Mi Band 2, is shown in
. The sleep period and wake period can be distinguished using AX3 data, which shows that YA1 was mostly static with a relatively high and stable wrist temperature during sleep. The sleep onset and offset detected by the Mi Band 2 align well with changes in AX3 and wrist temperature data.Case II: Healthy Older Adult With Irregular Sleep
The healthy older adult (OA1) was an 83-year-old man (PSQI=6 and ESS=7) who did not report any sleep disorders. OA1 was considered to be an irregular sleeper based on his data; one day of representative data is shown in
. Sleep onsets detected by the Mi Band 2 were, on average, 2 hours later than OA1’s self-reported sleep onset.For the day represented in
, OA1 self-reported falling asleep at around 11:45 PM, whereas the Mi Band 2 reported that his sleep started at 2:43 AM. Significant movements were captured by AX3 between 12:00 AM and 2:30 AM; a magnified image, except for these data, is shown on the right side of , where multiple peaks can be observed in all 3 axes. After 2:30 AM, the AX3 data became more static and similar to the AX3 data collected from regular sleepers. The movements between 12:00 AM and 2:30 AM were very periodic and occurred approximately every 30 seconds for most of the 2.5-hour time frame; during this time, the wrist temperature remained relatively high, and the participant reported being asleep. As it seems unlikely that these movements were made by the participant consciously and people cannot fake body temperature changes, the participant was likely asleep but experiencing irregular body movements.Case III: OAWD1 With Parkinson Disease and Sleep Apnea
OAWD1 was a 68-year-old man (MMSE score=21, diagnosed with Parkinson disease and sleep apnea); 2 days of data are shown in
. Compared with OA1 ( ), OAWD1’s data (both wrist temperature and AX3 data) have a more evident contrast between day and night. Even so, only parts of his sleep were detected by Mi Band 2. As shown in the left portion of , there was little movement between 9:00 PM and 3:00 AM, and the wrist temperature was relatively high, but this time frame was not identified as sleep by Mi Band 2.Case IV: OAWD2 With Sleep Apnea and Insomnia
OAWD2 was a 68-year-old woman (MMSE score=19) who was diagnosed with sleep apnea and insomnia and was an irregular sleeper. Before the start of the study, OAWD2’s PSW reported that she had very poor sleep (ie, would wake up at night frequently and sometimes cannot fall asleep at all). During her 13 days in our study, the Mi Band 2 indicated no sleep for 11 days; PSWs confirmed that she wore the wristbands the entire time.
shows 2 consecutive days of wrist temperature and AX3 data. The left portion of shows data where no sleep was detected by Mi Band 2, and the right portion shows a short period of sleep detected by Mi Band 2. Compared with data from OA1 in , there are no obvious static periods, and the day and night contrast for movement is relatively indistinguishable. Although there are differences in daytime and nighttime wrist temperatures, these are more difficult to distinguish than in other participants.A close-up of an excerpt of AX3 data for day 1 is shown in
, where 8 short episodes of static data are highlighted. Each highlighted episode lasted <30 minutes, and considerable movement was observed between episodes. These data indicate that the participant may have had some sleep, but if so, the sleep was in short, fragmented periods. As the participant was diagnosed with sleep apnea, this could partially account for frequent nighttime wake-ups.Discussion
Principal Findings
In line with other recent studies [
, ], this study found that wrist temperature increased when people were asleep, even when their sleep patterns were irregular. As illustrated in the case studies presented earlier, wrist temperature appears to correspond with sleep status changes regardless of whether body movements occurred; wrist temperature increased when people were asleep and decreased when they woke up. This adds evidence that wrist temperature is correlated with circadian rhythms and associated changes in thermal regulation of the body during sleep. This appears to hold true for OAWD as well, whose circadian rhythms are impacted by both aging and dementia (and often other morbidities); their wrist temperature appears to reflect when they are asleep and awake. A larger sample size is needed to draw more definitive conclusions regarding the impact dementia may have on CBT and wrist temperature.From the results presented in
and , there seems to be a trend of increasing wake temperature and decreasing sleep and wake temperature difference as people age and people become more pronounced dementia in this study. It is not clear from our data whether and how much this increased discrepancy is due to dementia or other factors, such as increased average age of OAWD, greater prevalence of comorbidities, and/or how living in an (often busy) LTC environment may disrupt sleep. The less significant day versus night wrist temperature contrast suggests changes in the circadian rhythms of older adults and OAWD compared with younger adults. This finding is consistent with other studies (eg, [ ]).The IS values suggest that the 2 healthy groups had similar IS for their wrist temperature rhythm, and the dementia group had lower wrist temperature rhythm stability. The lower IV values of the older adult groups indicate that changes in wrist temperature rhythms tend to be less variable intraday as people get older, especially with people who have dementia. As IS reflects the stability and the IV reflects the fragmentation of one’s wrist temperature rhythm, the significantly lower IS and IV of OAWD could be associated with irregular sleep patterns and increased daytime sleepiness. Most OAWD in this study took frequent naps and experienced insomnia at night, which almost certainly increased the average daytime wrist temperature, lowered the average nighttime wrist temperature, and accounted for a greater amount of flux in nighttime wrist temperature. Rhythm indices, such as IS and IV, could be further explored with these and other populations to determine how they might be used to infer circadian rhythm health.
Some studies [
, - ] have reported excessive body or wrist movements of OAWD during sleep. This research supports the hypothesis that advanced sleep monitoring systems should detect nighttime body movements. Although the exact cause of abnormal body movements by the older adults in this study is unknown, body movement during sleep seems to become increasingly common with age and was present in all our OAWD participants. A system that can detect involuntary movements could be valuable in the detection of previously unknown conditions (such as the case with our OA1 participant) and possibly support the diagnosis and management of conditions that impact sleep.There are 2 noteworthy findings regarding sleep detection using the Mi Band 2. First, by comparing the results of the Mi Band 2 with AX3 data from the custom-built wristband and reported sleep from sleep journals, it appears that the Mi Band 2 does not accurately detect people’s sleep when they have pronounced or irregular body movements during sleep. Second, the Mi Band 2 sleep detection algorithm does not appear to work well if the user has fragmented sleep (ie, waking up multiple times during the night). For example, OAWD2 slept in short and sporadic episodes; most of her sleep was undetected by Mi Band 2, which reported sleep for only 2 of her 13 recorded days. From the example data in
, we can observe that each sleep episode of OAWD2 was short and each wake-up accompanied body movements and that the Mi Band 2 did not detect any sleep. As the Mi Band 2 detects sleep based on accemetry data, it is plausible that periodic body movements during sleep, such as those seen in the older adults presented in the case studies, may be the cause of the misclassification of these periods as awake instead of asleep. Moreover, the Mi Band 2 may be trained on data from people with 1 or 2 long nighttime sleeping periods than multiple short ones. The hypothesis that movements and short sleeping episodes cause the Mi Band to miss episodes of sleep is supported by the data from our other participants.Although the sleep detection algorithms for most commercial smart wristbands are proprietary information, as they rely on accelerometers and gyroscopes, their sleep detection is based on movement. For a regular sleeper, frequent and substantial body movements at night usually indicate the person is awake, and fewer and infrequent movements indicate sleep. Although an activity-based rule for sleep detection may be robust for people who are mostly still when they sleep, this study shows that sleep detection based on activity is not accurate for some older adults, especially OAWD. As younger adults are the primary consumers of wearable wristbands, the Mi Band 2 sleep detection algorithms may have been trained predominantly on data from younger adults with healthy sleep patterns, thus biasing the algorithm to detect sleep for people who fit that profile. This aligns with other studies that have found that commercial wristbands performed poorly on populations with sleep disorders [
]. Therefore, other smart wristbands that track sleep based on movement alone may have similar performance as the Mi Band 2 for the same reasons; however, this was not investigated in this study; therefore, it remains a question for future research.This research highlights the need for the older adult population to be included in the development of sleep detection algorithms, including the possibility of developing smarter algorithms that autonomously identify the type of sleeper a person is and self-adjust appropriately to that person’s sleep habits. This requires research into how to categorize what good sleep is for that particular person in a way that helps people, their care partners, and/or caregivers understand. This information could then be used to better manage sleep without missing information that may indicate a change in sleep or sleep patterns that are of possible concern. As wrist temperature appears to be highly correlated with sleep regardless of age or dementia, the inclusion of wrist temperature as a complementary sensor to movement-based ones could enable their strengths to be leveraged and fused to support more accurate sleep detection.
More accurate sleep detection for older adults using wearable systems could support better short-term and long-term monitoring of sleep. Through daily sleep tracking, irregular sleep quality and patterns can be recognized by machine learning algorithms and then be reported to older adults or their health care providers. Such monitoring systems are especially useful for OAWD because the incidence of poor sleep is quite high in this population, and self-reporting of sleep is often difficult or impossible. Wearable technologies for sleep tracking and sleep problem identification could help to better understand older adults’ sleep using objective measures, but this approach is only possible if the results are accurate. Accurate sleep data would enable the creation of tailored sleep management plans, enabling better sleep support for each individual, thus supporting better health outcomes and quality of life.
Limitations
This study had several limitations. First, only the Mi Band 2 was the only commercially available wearable device that was examined. As other commercially available smart wristbands collect other types of data (eg, PPG) and use different sleep detection algorithms, their performance will likely be different. These alternatives should be examined in healthy older adults and OAWD in future studies. Second, the sample size of this study was small. Future research with larger sample sizes is required to provide a deeper, comprehensive understanding that needs to support greater generalization of the results. Third, all our OAWD participants lived in the same LTC home. Data from other LTC homes as well as OAWD living in the community need to be investigated to determine how the environment plays a role in sleep detection and dementia. Finally, sleep onset and offset were estimated from sleep journals and/or actigraphy data. Using another method, such as computer vision, could estimate onset and offset more accurately as well as provide information about what was happening during sleep (eg, giving insight into repetitive movements).
Conclusions
This research adds evidence that wrist temperature can be used as an indicator of sleep status, including for OAWD and people with irregular sleep patterns. As illustrated by case studies from our data, suboptimal sleep detection performance by a commercial wristband was likely because of broken sleep patterns and body movements. As wearable technologies are increasingly being used by LTC homes and healthy older adults to track sleep and inform sleep management and support, this research suggests that caution should be used when interpreting sleep data when monitoring the sleep of older adults, particularly those living with dementia. This highlights the need for future research and development to create systems that better complement older adult populations. Future sleep monitoring wearable systems could consider adding a temperature sensor to capture the waist temperature as an extra indicator of sleep combined with motion-based data and machine learning algorithms of typical sleep patterns.
Acknowledgments
This study was funded by the principal investigator’s Research Chair from the Schlegel-University of Waterloo Research Institute for Aging. JW was affiliated with the Department of Systems Design Engineering at the University of Waterloo at the time of the study and is currently affiliated with School of Computing and Information System at the University of Melbourne.
Conflicts of Interest
None declared.
References
- Buxton OM, Marcelli E. Short and long sleep are positively associated with obesity, diabetes, hypertension, and cardiovascular disease among adults in the United States. Soc Sci Med 2010 Sep;71(5):1027-1036. [CrossRef]
- Van Dongen HP, Maislin G, Mullington JM, Dinges DF. The cumulative cost of additional wakefulness: dose-response effects on neurobehavioral functions and sleep physiology from chronic sleep restriction and total sleep deprivation. Sleep 2003 Mar 15;26(2):117-126. [Medline]
- Walker MP, Stickgold R. Sleep, memory, and plasticity. Annu Rev Psychol 2006;57:139-166. [CrossRef] [Medline]
- Hudson AN, Van Dongen HP, Honn KA. Sleep deprivation, vigilant attention, and brain function: a review. Neuropsychopharmacol 2019 Jun 8;45(1):21-30. [CrossRef]
- Ju YE, Lucey BP, Holtzman DM. Sleep and Alzheimer disease pathology--a bidirectional relationship. Nat Rev Neurol 2014 Feb;10(2):115-119 [FREE Full text] [CrossRef] [Medline]
- Musiek ES, Xiong DD, Holtzman DM. Sleep, circadian rhythms, and the pathogenesis of Alzheimer disease. Exp Mol Med 2015 Mar 13;47(3):e148. [CrossRef]
- Cordone S, Annarumma L, Rossini PM, De Gennaro L. Sleep and β-Amyloid deposition in Alzheimer disease: insights on mechanisms and possible innovative treatments. Front Pharmacol 2019;10:695 [FREE Full text] [CrossRef] [Medline]
- Drager LF, McEvoy RD, Barbe F, Lorenzi-Filho G, Redline S, INCOSACT Initiative (International Collaboration of Sleep Apnea Cardiovascular Trialists). Sleep apnea and cardiovascular disease: lessons from recent trials and need for team science. Circulation 2017 Nov 07;136(19):1840-1850 [FREE Full text] [CrossRef] [Medline]
- Azarbarzin A, Sands SA, Stone KL, Taranto-Montemurro L, Messineo L, Terrill PI, et al. The hypoxic burden of sleep apnoea predicts cardiovascular disease-related mortality: the Osteoporotic Fractures in Men Study and the Sleep Heart Health Study. Eur Heart J 2019 Apr 07;40(14):1149-1157 [FREE Full text] [CrossRef] [Medline]
- Kushida CA, Chang A, Gadkary C, Guilleminault C, Carrillo O, Dement WC. Comparison of actigraphic, polysomnographic, and subjective assessment of sleep parameters in sleep-disordered patients. Sleep Med 2001 Sep;2(5):389-396. [CrossRef] [Medline]
- Kingshott RN, Douglas NJ. The effect of in-laboratory polysomnography on sleep and objective daytime sleepiness. Sleep 2000 Dec 15;23(8):1109-1113. [Medline]
- Sadeh A, Hauri PJ, Kripke DF, Lavie P. The role of actigraphy in the evaluation of sleep disorders. Sleep 1995 May;18(4):288-302. [CrossRef] [Medline]
- Sadeh A. The role and validity of actigraphy in sleep medicine: an update. Sleep Med Rev 2011 Aug;15(4):259-267. [CrossRef] [Medline]
- Ancoli-Israel S, Cole R, Alessi C, Chambers M, Moorcroft W, Pollak CP. The role of actigraphy in the study of sleep and circadian rhythms. Sleep 2003 May 01;26(3):342-392. [Medline]
- Smith MT, McCrae CS, Cheung J, Martin JL, Harrod CG, Heald JL, et al. Use of actigraphy for the evaluation of sleep disorders and circadian rhythm sleep-wake disorders: an American Academy of Sleep Medicine Clinical Practice Guideline. J Clin Sleep Med 2018 Jul 15;14(7):1231-1237 [FREE Full text] [CrossRef] [Medline]
- Refinetti R, Menaker M. The circadian rhythm of body temperature. Physiol Behav 1992 Mar;51(3):613-637. [CrossRef] [Medline]
- Sarabia JA, Rol MA, Mendiola P, Madrid JA. Circadian rhythm of wrist temperature in normal-living subjects A candidate of new index of the circadian system. Physiol Behav 2008 Nov 28;95(4):570-580. [CrossRef] [Medline]
- Ortiz-Tudela E, Martinez-Nicolas A, Campos M, Rol MA, Madrid JA. A new integrated variable based on thermometry, actimetry and body position (TAP) to evaluate circadian system status in humans. PLoS Comput Biol 2010 Nov 11;6(11):e1000996 [FREE Full text] [CrossRef] [Medline]
- Wei J, Zhang J, Boger J. What wrist temperature tells us when we sleep late: a new perspective of sleep health. In: Proceedings of the IEEE SmartWorld, Ubiquitous Intelligence & Computing, Advanced & Trusted Computing, Scalable Computing & Communications, Cloud & Big Data Computing, Internet of People and Smart City Innovation (SmartWorld/SCALCOM/UIC/ATC/CBDCom/IOP/SCI). 2018 Presented at: IEEE SmartWorld, Ubiquitous Intelligence & Computing, Advanced & Trusted Computing, Scalable Computing & Communications, Cloud & Big Data Computing, Internet of People and Smart City Innovation (SmartWorld/SCALCOM/UIC/ATC/CBDCom/IOP/SCI); Oct 8-12, 2018; Guangzhou, China p. A. [CrossRef]
- Wei J, Boger J. You are how you sleep: personalized sleep monitoring based on wrist temperature and accelerometer data. In: Proceedings of the 13th EAI International Conference on Pervasive Computing Technologies for Healthcare - Demos and Posters. 2019 Presented at: 13th EAI International Conference on Pervasive Computing Technologies for Healthcare - Demos and Posters; June 11, 2019; Trento, Italy. [CrossRef]
- Guillodo E, Lemey C, Simonnet M, Walter M, Baca-García E, Masetti V, et al. Clinical applications of mobile health wearable–based sleep monitoring: systematic review. JMIR Mhealth Uhealth 2020 Apr 1;8(4):e10733. [CrossRef]
- Madrid-Navarro CJ, Cuesta FJ, Escamilla-Sevilla F, Campos M, Ruiz Abellán F, Rol MA, et al. Validation of a device for the ambulatory monitoring of sleep patterns: a pilot study on Parkinson's disease. Front Neurol 2019;10:356 [FREE Full text] [CrossRef] [Medline]
- Landry GJ, Best JR, Liu-Ambrose T. Measuring sleep quality in older adults: a comparison using subjective and objective methods. Front Aging Neurosci 2015;7:166 [FREE Full text] [CrossRef] [Medline]
- Potvin O, Lorrain D, Forget H, Dubé M, Grenier S, Préville M, et al. Sleep quality and 1-year incident cognitive impairment in community-dwelling older adults. Sleep 2012 Apr 01;35(4):491-499 [FREE Full text] [CrossRef] [Medline]
- Nebes RD, Buysse DJ, Halligan EM, Houck PR, Monk TH. Self-reported sleep quality predicts poor cognitive performance in healthy older adults. J Gerontol B Psychol Sci Soc Sci 2009 Mar;64(2):180-187 [FREE Full text] [CrossRef] [Medline]
- Wang H, Zhao Y, Yu L, Liu J, Zwetsloot IM, Cabrera J, et al. A personalized health monitoring system for community-dwelling elderly people in Hong Kong: design, implementation, and evaluation study. J Med Internet Res 2020 Sep 30;22(9):e19223 [FREE Full text] [CrossRef] [Medline]
- Jirong Y, Changquan H, Hongmei W, Bi-Rong D. Association of sleep quality and dementia among long-lived Chinese older adults. Age (Dordr) 2013 Aug;35(4):1423-1432 [FREE Full text] [CrossRef] [Medline]
- Kim JH, Duffy JF. Circadian rhythm sleep-wake disorders in older adults. Sleep Med Clin 2018 Mar;13(1):39-50. [CrossRef] [Medline]
- Gulia KK, Kumar VM. Sleep disorders in the elderly: a growing challenge. Psychogeriatrics 2018 May;18(3):155-165. [CrossRef] [Medline]
- Shi L, Chen S, Ma M, Bao Y, Han Y, Wang Y, et al. Sleep disturbances increase the risk of dementia: a systematic review and meta-analysis. Sleep Med Rev 2018 Aug;40:4-16. [CrossRef]
- Rongve A, Boeve BF, Aarsland D. Frequency and correlates of caregiver-reported sleep disturbances in a sample of persons with early dementia. J Am Geriatr Soc 2010 Mar;58(3):480-486. [CrossRef] [Medline]
- Naismith SL, Glozier N, Burke D, Carter PE, Scott E, Hickie IB. Early intervention for cognitive decline: is there a role for multiple medical or behavioural interventions? Early Interv Psychiatry 2009 Feb;3(1):19-27. [CrossRef] [Medline]
- Yaffe K, Blackwell T, Barnes DE, Ancoli-Israel S, Stone KL, Study of Osteoporotic Fractures Group. Preclinical cognitive decline and subsequent sleep disturbance in older women. Neurology 2007 Jul 17;69(3):237-242. [CrossRef] [Medline]
- Leng Y, Musiek ES, Hu K, Cappuccio FP, Yaffe K. Association between circadian rhythms and neurodegenerative diseases. Lancet Neurol 2019 Mar;18(3):307-318 [FREE Full text] [CrossRef] [Medline]
- Yu RP, Ellison NB, McCammon RJ, Langa KM. Mapping the two levels of digital divide: internet access and social network site adoption among older adults in the USA. Inf Commun Soc 2015 Nov 19;19(10):1445-1464. [CrossRef]
- Berkowsky RW, Sharit J, Czaja SJ. Factors predicting decisions about technology adoption among older adults. Innov Aging 2018 Jan;2(1):igy002 [FREE Full text] [CrossRef] [Medline]
- Fitbit. URL: https://www.fitbit.com [accessed 2021-04-30]
- Mi Band. URL: https://www.mi.com/global/miband/ [accessed 2021-04-30]
- Garmin. URL: https://www.garmin.com [accessed 2021-04-30]
- Izmailova ES, Wagner JA, Perakslis ED. Wearable devices in clinical trials: hype and hypothesis. Clin Pharmacol Ther 2018 Jul;104(1):42-52 [FREE Full text] [CrossRef] [Medline]
- Degroote L, Hamerlinck G, Poels K, Maher C, Crombez G, De Bourdeaudhuij I, et al. Low-cost consumer-based trackers to measure physical activity and sleep duration among adults in free-living conditions: validation study. JMIR Mhealth Uhealth 2020 May 19;8(5):e16674. [CrossRef]
- Liu J, Zhao Y, Lai B, Wang H, Tsui KL. Wearable device heart rate and activity data in an unsupervised approach to personalized sleep monitoring: algorithm validation. JMIR Mhealth Uhealth 2020 Aug 05;8(8):e18370 [FREE Full text] [CrossRef] [Medline]
- Gibson RH, Gander PH. Monitoring the sleep patterns of people with dementia and their family carers in the community. Australas J Ageing 2019 Mar;38(1):47-51. [CrossRef] [Medline]
- Horne JA, Ostberg O. A self-assessment questionnaire to determine morningness-eveningness in human circadian rhythms. Int J Chronobiol 1976;4(2):97-110. [Medline]
- Buysse DJ, Reynolds CF, Monk TH, Berman SR, Kupfer DJ. The Pittsburgh Sleep Quality Index: a new instrument for psychiatric practice and research. Psychiatry Res 1989 May;28(2):193-213. [Medline]
- Johns MW. A new method for measuring daytime sleepiness: the Epworth sleepiness scale. Sleep 1991 Dec;14(6):540-545. [Medline]
- Monk TH, Reynolds CF, Kupfer DJ, Buysse DJ, Coble PA, Hayes AJ, et al. The Pittsburgh sleep diary. J Sleep Res 1994;3:111-120 [FREE Full text] [Medline]
- Tombaugh TN, McIntyre NJ. The mini-mental state examination: a comprehensive review. J Am Geriatr Soc 1992 Sep;40(9):922-935. [Medline]
- Martinez-Nicolas A, Madrid JA, García FJ, Campos M, Moreno-Casbas MT, Almaida-Pagán PF, et al. Circadian monitoring as an aging predictor. Sci Rep 2018 Oct 09;8(1):15027 [FREE Full text] [CrossRef] [Medline]
- Folstein MF, Folstein SE, McHugh PR. "Mini-mental state". A practical method for grading the cognitive state of patients for the clinician. J Psychiatr Res 1975 Nov;12(3):189-198. [Medline]
- Huang Y, Liu R, Wang Q, Van Someren EJ, Xu H, Zhou J. Age-associated difference in circadian sleep-wake and rest-activity rhythms. Physiol Behav 2002 Aug;76(4-5):597-603. [CrossRef] [Medline]
- Hibi S, Yamaguchi Y, Umeda-Kameyama Y, Yamamoto H, Iijima K, Momose T, et al. The high frequency of periodic limb movements in patients with Lewy body dementia. J Psychiatr Res 2012 Dec;46(12):1590-1594. [CrossRef] [Medline]
- Pao WC, Boeve BF, Ferman TJ, Lin S, Smith GE, Knopman DS, et al. Polysomnographic findings in dementia with Lewy bodies. Neurologist 2013 Jan;19(1):1-6. [CrossRef] [Medline]
- McCurry SM, Ancoli-Israel S. Sleep dysfunction in Alzheimer's disease and other Dementias. Curr Treat Options Neurol 2003 May;5(3):261-272. [CrossRef] [Medline]
- Kang SG, Kang JM, Ko K, Park S, Mariani S, Weng J. Validity of a commercial wearable sleep tracker in adult insomnia disorder patients and good sleepers. J Psychosom Res 2017 Jun;97:38-44. [CrossRef] [Medline]
Abbreviations
CBT: core body temperature |
ESS: Epworth Sleepiness Scale |
IS: interday stability |
IV: intraday variability |
LTC: long-term care |
MMSE: Mini-Mental State Examination |
OAWD: older adults living with dementia |
PSG: polysomnography |
PSQI: Pittsburgh Sleep Quality Index |
PSW: personal support worker |
Edited by L Buis; submitted 12.12.20; peer-reviewed by H Kazemi Arpanahi, V Stara; comments to author 24.01.21; revised version received 03.02.21; accepted 12.04.21; published 01.06.21
Copyright©Jing Wei, Jennifer Boger. Originally published in JMIR mHealth and uHealth (https://mhealth.jmir.org), 01.06.2021.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mHealth and uHealth, is properly cited. The complete bibliographic information, a link to the original publication on https://mhealth.jmir.org/, as well as this copyright and license information must be included.