Published on in Vol 6, No 1 (2018): January

Hearing Tests Based on Biologically Calibrated Mobile Devices: Comparison With Pure-Tone Audiometry

Hearing Tests Based on Biologically Calibrated Mobile Devices: Comparison With Pure-Tone Audiometry

Hearing Tests Based on Biologically Calibrated Mobile Devices: Comparison With Pure-Tone Audiometry

Original Paper

1Department and Clinic of Otolaryngology, Head and Neck Surgery, Faculty of Postgraduate Medical Training, Wroclaw Medical University, Wrocław, Poland

2Department of Biomedical Engineering, Faculty of Fundamental Problems of Technology, Wroclaw University of Science and Technology, Wrocław, Poland

Corresponding Author:

Marcin Masalski, MD, PhD (Eng)

Department and Clinic of Otolaryngology, Head and Neck Surgery

Faculty of Postgraduate Medical Training

Wroclaw Medical University

Borowska 213

Wrocław, 50-556

Poland

Phone: 48 515086252

Email: marcin.masalski@pwr.edu.pl


Background: Hearing screening tests based on pure-tone audiometry may be conducted on mobile devices, provided that the devices are specially calibrated for the purpose. Calibration consists of determining the reference sound level and can be performed in relation to the hearing threshold of normal-hearing persons. In the case of devices provided by the manufacturer, together with bundled headphones, the reference sound level can be calculated once for all devices of the same model.

Objective: This study aimed to compare the hearing threshold measured by a mobile device that was calibrated using a model-specific, biologically determined reference sound level with the hearing threshold obtained in pure-tone audiometry.

Methods: Trial participants were recruited offline using face-to-face prompting from among Otolaryngology Clinic patients, who own Android-based mobile devices with bundled headphones. The hearing threshold was obtained on a mobile device by means of an open access app, Hearing Test, with incorporated model-specific reference sound levels. These reference sound levels were previously determined in uncontrolled conditions in relation to the hearing threshold of normal-hearing persons. An audiologist-assisted self-measurement was conducted by the participants in a sound booth, and it involved determining the lowest audible sound generated by the device within the frequency range of 250 Hz to 8 kHz. The results were compared with pure-tone audiometry.

Results: A total of 70 subjects, 34 men and 36 women, aged 18-71 years (mean 36, standard deviation [SD] 11) participated in the trial. The hearing threshold obtained on mobile devices was significantly different from the one determined by pure-tone audiometry with a mean difference of 2.6 dB (95% CI 2.0-3.1) and SD of 8.3 dB (95% CI 7.9-8.7). The number of differences not greater than 10 dB reached 89% (95% CI 88-91), whereas the mean absolute difference was obtained at 6.5 dB (95% CI 6.2-6.9). Sensitivity and specificity for a mobile-based screening method were calculated at 98% (95% CI 93-100.0) and 79% (95% CI 71-87), respectively.

Conclusions: The method of hearing self-test carried out on mobile devices with bundled headphones demonstrates high compatibility with pure-tone audiometry, which confirms its potential application in hearing monitoring, screening tests, or epidemiological examinations on a large scale.

JMIR Mhealth Uhealth 2018;6(1):e10

doi:10.2196/mhealth.7800

Keywords



This study investigated the accuracy of the hearing tests conducted on mobile devices calibrated by means of the biological method, that is, in relation to the hearing threshold of a normal-hearing person. This study constitutes the second part of the planned research. The calibration method was adopted from a previous study [1], and it involved semiautomated determination of predefined, model-specific reference sound level.

Hearing loss is a disorder widely encountered among the world population. It is estimated that it affects 5.3% of population, which totals 360 million patients suffering from hearing loss around the world [2]. Monitoring and diagnosis of hearing are especially important in preventing and treatment of hearing loss. One of the basic hearing tests is pure-tone audiometry, which determines the hearing threshold in relation to the sound frequency. However, pure-tone audiometry requires access to specialized medical equipment and staff.

Automated audiometry involves self-determining the hearing threshold and is a valuable diagnostic tool in the case of limited access to medical personnel [3-8]. The methods of automated audiometry have been developed over many years. At present, common methods involve the assessment of the reaction to an automatically generated test signal [4,8-21] or determining the lowest audible sound via self-adjustment of test signal intensity [6,7,9,22]. The hearing thresholds obtained by means of both methods on calibrated devices are comparable with conventional pure-tone audiometry [4-7,9,11-14,19,20,22].

Audiometric tests can be conducted on specialized equipment as well as on generally accessible electronic devices such as personal computers [10,22,23] or mobile devices [4-9,15-21,24,25]. However, reliable results may only be obtained if they are correctly calibrated and conducted in silence [1,15,21,26,27]. Assuming that mobile devices sold with bundled headphones have similar frequency characteristics of the sound system within the devices of the same model, it is possible to share calibration coefficients within these groups. Examinations on iOS-based devices confirmed the validity of automated audiometry based on common calibration coefficients determined in laboratory conditions for all devices of this type [4].

Besides applications typical for automated audiometry such as preliminary evaluation of audiological patients and screening tests [3-8], tests on common electronic devices may be useful for self-monitoring of hearing, especially during recovery from sudden sensorineural hearing loss; in Ménière’s disease, tinnitus or other fluctuating hearing loss, or age-related hearing loss; or during ototoxic therapy [5,6,8,10,16,17,22]. The development of audiometric apps is stimulated by a wide range of potential applications. There is a large number of mobile apps on the market, but only few of them have been investigated in research, which is crucial before they can be clinically applied [28,29].

The objective of this trial was to determine the accuracy of automated audiometry conducted on Android-based mobile sets including the device and the bundled headphones that were calibrated semiautomatically via a biological method. Android-based devices constitute 80% of all mobile devices around the world, and their number is estimated to be 1.4 billion [30]. The prevalence of these devices may contribute to improving accessibility of audiological examinations, especially in the parts of the world with limited access to specialized equipment [5,7,24,31,32]. Android-based mobile devices, contrary to iOS-based ones, are produced by many manufacturers, which means that they are not unified in terms of hardware solutions. This is reflected by statistically significant differences that were found in frequency characteristic of the sound system [1]. Additionally, the number of Android-based models or their variations in 2015 exceeded 24,000, with an increase of 5000 compared with the previous year [33]. Due to the diversity of this group, its size, and dynamics of its substantial changes, laboratory calibration appears to be an inefficient solution.

Calibration of mobile devices conducted by means of a biological method involves determining reference sound level in relation to the hearing threshold of normal-hearing persons [1,34]. Calibration conducted several times on different mobile sets of the same model allows for determining a reliable, model-specific reference sound level [1]. The adopted solution provides support without additional workload not only for the current but also for the future models of mobile sets.


This study involved a comparison of hearing thresholds determined by pure-tone audiometry with the hearing thresholds determined on Android-based mobile devices calibrated biologically in uncontrolled conditions. It was a single-center, crossover trial carried out on patients of Otolaryngology Clinic. The consent to conduct the trial has been granted by the Bioethics Committee of Wroclaw Medical University.

Recruitment

Study participants were recruited offline, using face-to-face prompting. Eligibility assessment was conducted among subjects who owned an Android-based device with bundled headphones. Eligibility criterion, apart from the willingness to participate in the study, included owning the device, for which model-specific calibration coefficient has been previously determined by means of biological method [1]. All the study participants conducted examinations on their mobile devices.

Measurements

Measurements involved conducting pure-tone audiometry and the test on a mobile device. The order was arranged in a counterbalanced manner. The measurements for both methods were conducted in a sound booth for the frequencies 250 Hz, 500 Hz, 1 kHz, 2 kHz, 4 kHz, 6 kHz, and 8 kHz.

Pure-tone audiometry was conducted by means of conventional 10 dB down and 5 dB up bracketing method (modified Hughson-Westlake method) in accordance with the recommendations of the British Society of Audiology [35]. The measurements were conducted by an audiologist on a clinical audiometer Interacoustic AD229e using TDH-39 headphones (Interacoustics, Denmark, Audiometer Allé, 5500 Middelfart) previously calibrated according to ISO 389-1:1998.

Hearing examinations on mobile devices were conducted using a free app Hearing Test [36,37] available on Google Play. The examinations involved self-determining of the lowest audible sound generated by the device and were audiologist-assisted. The participant changed intensity of the test sound using the buttons “I can hear” and “I cannot hear” and then confirmed the lowest audible sound using the button “Barely audible.” During the adjustment of the sound intensity, temporary test result was continuously presented on the device screen (Figure 1). The role of the supervising audiologist was to prevent obvious mistakes, such as switching the sides of headphones or omitting the frequency by accidental button pressing. The audiologist also provided assistance in cases of doubts concerning proper execution of the test. Test tone was amplitude modulated with the depth of 100% and the frequency of 2 Hz. Sound intensity was changed in 5 dB steps. When the intensity of the test sound exceeded 40 dB HL, a masking sound was generated contralaterally. The tests on mobile devices were conducted twice: test and retest. Before the retest, the headphones were removed and put on again.

Calibration of mobile devices was described in detail in the previous paper [1]. The objective of calibration was to determine the intensity of the signal generated by the device, which would generate a sound in bundled headphones at the reference level of 0 dB HL. Calibration was performed by means of a biological method, which led to elaboration of a new mobile hearing level (mHL) scale. The level of 0 dB mHL was defined as the sound level generated by the mobile set equivalent to the hearing threshold of normal-hearing persons, reduced by their estimated hearing threshold. The advantage of this approach is the ability to calibrate new device models based on measurements conducted by users themselves, which is essential due to high rotation of mobile models on the market. The calibration is required only for users who are the first to become the owners of new devices, whereas the others can reuse their results.

Calibration measurements involved self-determining the hearing thresholds by means of Bekesy’s method [1,10,11,22,34] by users aged 18-35 years who considered themselves normal-hearing persons. There were no other requirements for calibration. However, if these requirements were not met, a user could ask another person for help. These calibration thresholds were then decreased by the population median to obtain the final reference level. More precisely, the estimator of central tendency of the determined thresholds were approximated by 37th percentile and then decreased by the median of the hearing threshold estimated based on literature data [38] for the corresponding population, that is, aged 18-35 years [1]. Calibration measurements were conducted in uncontrolled conditions, irrespective of this research, before its commencement and when it was in progress, by application users who were not trial participants. To calibrate the model, at least 15 measurements were required to be conducted on various devices belonging to the same model [1]. After each subsequent measurement, the calibration results were automatically updated.

Statistical Analysis

Sample size was determined on the basis of the standard deviation (SD) between the hearing thresholds in pure-tone audiometry and thresholds determined on mobile devices estimated at 8.42 dB [1]. With the statistical significance of 0.05, statistical power of 0.8, and the effect size of 2.0 dB, the sample volume was estimated on 140 ears (70 subjects). Hearing thresholds obtained on mobile devices have been compared with pure-tone audiometry and analyzed in test-retest examination. Differences in hearing thresholds, intraclass correlation coefficients, and Cronbach alpha with respective confidence intervals have been determined. The results were presented in dependency plots as well as in Bland-Altman plots.

Figure 1. Screenshot of the Hearing Test app during examination.
View this figure

In the period between November 11, 2015, and November 2, 2016, 89 subjects were assessed for eligibility from among patients of Otolaryngology Department who owned a mobile device with bundled headphones. Out of 89 subjects, 7 (8%) were not interested in the examination and did not give their consent; 12 out of the remaining 82 subjects (15%) owned devices that were not yet calibrated (Figure 2).

A total of 70 subjects, including 34 men and 36 women, aged 18-71 years (mean 36, SD 11) participated in the trial. Among the participants were both normal-hearing subjects and hearing-impaired patients. Table 1 presents the pure-tone audiometry thresholds for tested frequencies, Table 2 shows distribution of the hearing loss by type, and Table 3 summarizes the models of mobile devices applied.

The research was arranged in a counterbalanced manner. Out of 70 subjects, 35 (50%) were first tested by pure-tone audiometry, whereas the other half by the mobile device. Out of the range measurements were discarded from further analysis: 13 of 980 (1.3%), 14 of 966 (1.4%), and 10 of 980 (1.0%) in the case of test, retest, and pure-tone audiometry, respectively. Hearing thresholds determined through pure-tone audiometry were compared with thresholds obtained on mobile devices separately for both groups (Table 4). At the level of statistical significance of P=.05, no differences were found between groups, and thus, further analyses were conducted jointly (Tables 5 and 6, and Figures 3 and 4). The mean difference between the hearing threshold determined by pure-tone audiometry and on mobile devices was significantly different from 0 and reached 2.6 dB (95% CI 2.0-3.1), with SD of 8.3 dB (95% CI 7.9-8.7). Intraclass correlation for consistency of single measurement was obtained at the level of 0.85 (95% CI 0.83-0.87), whereas the Cronbach alpha for reliability was calculated at 0.92 (95% CI 0.91-0.93). The number of within-subject differences not greater than 10 dB reached 89% (95% CI 88-91), whereas the mean absolute difference was 6.5 dB (95% CI 6.2-6.9). The largest differences were noted for the frequencies of 6 and 8 kHz at the level of 4.0 dB (95% CI 2.5-5.4) and 7.0 dB (95% CI 5.4-8.6), respectively, whereas the lowest, statistically insignificant at P=.05, for the frequency of 500 Hz and 1 kHz.

Figure 2. Flow diagram.
View this figure
Table 1. Pure-tone audiometry thresholds among study participants.
Hearing loss (dB HL)Number of ears

250 Hz500 Hz1 kHz2 kHz4 kHz6 kHz8 kHz
15 and less1081081181181108793
16-2514161112122215
26-4011118871513
41-606420478
60-800001555
Over 801111246
Table 2. Type of hearing loss among study participants.
Type of hearing lossNumber of participants
Normal hearinga36
Sensorineural hearing loss

Sudden deafness6

Presbyaccusis3

Neuronitis vestibularis3

Postviral complication1

Other and indefinite sensorineural hearing loss8
Conductive and mixed hearing loss

Chronic otitis9

Cholesteatoma2

Exudative otitis1

Other and indefinite conductive or mixed hearing loss1

aHearing threshold in pure-tone audiometry no higher than 25 dB HL for all tested frequencies.

Table 3. Mobile devices by models.
Manufacturer and modelNumber of devices
Samsung SM-G900F7
Huawei ALE-L214
Samsung SM-G3504
Samsung GT-I93003
Samsung GT-I95053
Samsung SM-A500FU3
Samsung SM-G386F3
Samsung SM-G530FZ3
Samsung GT-I8190N2
Samsung GT-I9060I2
Samsung GT-I95152
Samsung SM-G800F2
Samsung SM-A300FU2
Sony D66032
Sony D23032
Sony E23032
Other24
Table 4. Differences in hearing threshold determined by pure-tone audiometry and on mobile devices depending on the order of examinations.
Frequency in HzFirst pure-tone audiometryFirst examination on a mobile device
nMean difference in dB (95% CI)nMean difference in dB (95% CI)
250703.6 (1.3-5.8)691.8 (−0.2 to 3.8)
50070−0.8 (−2.6 to 1.0)69−0.4 (−2.5 to 1.6)
100070−0.4 (−2.1 to 1.2)69−1.7 (−3.1 to −0.2)
2000702.6 (1.2-3.9)693.3 (1.9-4.7)
4000702.4 (0.4-4.3)694.1 (2.4-5.9)
6000704.0 (2.2-5.8)693.9 (1.6-6.2)
8000696.5 (4.2-8.8)647.5 (5.3-9.7)
Total4892.5 (1.8-3.3)4782.6 (1.9-3.4)
Table 5. Comparison of the hearing threshold determined by pure-tone audiometry and on mobile devices.

Frequency in Hz

n

Mean difference
in dB (95% CI)

Standard deviation
in dB (95% CI)

Mean absolute difference
in dB (95% CI)
Intraclass correlationa
(95% CI)
2501392.7 (1.2-4.2)9.0 (8.1-10.2)6.9 (6.0-7.9)0.75 (0.67-0.82)
500139−0.6 (−2.0 to 0.7)8.0 (7.2-9.1)5.9 (5.2-6.8)0.79 (0.72-0.85)
1000139−1.0 (−2.1 to 0.1)6.6 (5.9-7.4)4.8 (4.1-5.5)0.80 (0.73-0.85)
20001392.9 ( 2.0-3.9)5.7 (5.1-6.4)4.9 (4.3-5.5)0.86 (0.82-0.90)
40001393.2 (1.9-4.5)7.8 (7.0-8.8)6.7 (5.9-7.6)0.90 (0.86-0.92)
60001394.0 (2.5-5.4)8.4 (7.5-9.6)7.3 (6.4-8.2)0.90 (0.87-0.93)
80001337.0 (5.4-8.6)9.3 (8.3-10.6)9.3 (8.2-10.5)0.85 (0.80-0.89)
Total9672.6 (2.0-3.1)8.3 (7.9-8.7)6.5 (6.2-6.9)0.85 (0.83-0.87)

aTwo-way random for consistency of single measurement.

Table 6. Percentage distribution of within-subject differences in the hearing threshold determined by pure-tone audiometry and on mobile devices.
Frequency (Hz)n0-5 dB (%)0-10 dB (%)0-15 dB (%)0-20 dB (%)0-25 dB (%)0-30 dB (%)
2501397487929699100
500139879498100100100
100013993979899100100
20001398196100100100100
400013969919799100100
600013966869499100100
80001335074899798100
Total96774899599100100
Figure 3. Hearing threshold determined on mobile devices (mHT) in relation to the hearing threshold in pure-tone audiometry (HT).
View this figure
Figure 4. Difference plots (Bland-Altman) between the hearing threshold determined by pure-tone audiometry (HT) and on mobile devices (mHT) (continuous line indicates mean difference; dashed line indicates ±1.96 standard deviation range).
View this figure

Retest was conducted on a mobile device in 69 of 70 subjects (99%). At the level of statistical significance of P=.05, no differences in hearing threshold were found at any frequency (Table 7). The mean difference was −0.1 dB (95% CI −0.4 to 0.2) with SD of 4.4 dB (95% CI 4.2-4.6); 99% (95% CI 99-100) of within-subject differences were within the range of 10 dB, and the mean absolute difference reached 2.8 dB (95% CI 2.6-3.0) (Tables 7 and 8).

The assessment of specificity and sensitivity of hearing loss detection was conducted based on the criterion [39] adopted for the purposes of this study. Hearing loss was diagnosed when the threshold exceeded 30 dB at one of the following frequencies: 500 Hz, 1 kHz, 2 kHz, or 25 dB at more than one, or when the hearing threshold exceeded 50 dB at 4 kHz. The obtained sensitivity was at 98% (95% CI 93-100) with specificity of 79% (95% CI 71-87).

Table 7. Comparison of hearing thresholds determined on mobile devices in test-retest examination.

Frequency in Hz

n

Mean difference
in dB (95% CI)

Standard deviation
in dB (95% CI)

Mean absolute difference
in dB (95% CI)
Intraclass correlationa
(95% CI)
2501370.0 (−0.7 to 0.7)4.4 (3.9-5.0)2.9 (2.5-3.4)0.95 (0.93-0.96)
500137−0.3 (−1.0 to 0.5)4.3 (3.8-4.9)2.7 (2.3-3.2)0.94 (0.92-0.96)
10001370.1 (−0.6 to 0.8)4.3 (3.9-4.9)2.6 (2.2-3.0)0.92 (0.89-0.94)
2000137−0.4 (−1.0 to 0.2)3.6 (3.3-4.1)2.4 (2.0-2.8)0.94 (0.92-0.96)
40001370.0 (−0.7 to 0.7)4.2 (3.8-4.8)2.7 (2.3-3.1)0.97 (0.95-0.98)
6000137−0.1 (−0.9 to 0.7)4.5 (4.1-5.2)3.2 (2.8-3.7)0.97 (0.95-0.98)
8000130−0.2 (−1.1 to 0.7)5.1 (4.6-5.9)3.2 (2.6-3.7)0.95 (0.93-0.96)
Total952−0.1 (−0.4 to 0.2)4.4 (4.2-4.6)2.8 (2.6-3.0)0.96 (0.95-0.96)

aTwo-way random for consistency of single measurement.

Table 8. Percentage distribution of within-subject differences in the hearing threshold determined on mobile devices in test-retest examination.
Frequency (Hz)n0-5 dB (%)0-10 dB (%)0-15 dB (%)
2501379799100
5001379899100
10001379699100
200013799100100
400013797100100
600013797100100
80001309798100
Total9529799100

Principal Findings

In this paper, hearing thresholds determined by pure-tone audiometry were compared with hearing thresholds determined on Android-based mobile devices previously calibrated biologically in uncontrolled conditions. At the level of statistical significance P=.05, no differences in relation to the test order were observed. The mean hearing threshold difference of 2.6 dB (SD 8.3 dB) confirms the reliability of the method. Additionally, the mean test-retest difference of −0.1 dB (SD 4.4) indicates its high repeatability.

Comparison With Prior Study

The results are consistent with the research presented in previous works. The SD of the difference at 8.3 dB (95% CI 7.9-8.7) corresponds to the value estimated for this calibration method of 8.42 dB [1] and is close to the value of 7.8 dB obtained in another study conducted on the same app Hearing Test [7]. The mean difference of 2.6 dB (95% CI 2.0-3.1) and the mean absolute difference of 6.5 dB (95% CI 6.2-6.9) are also comparable with those presented in [7], that is, 0.7 dB and 7.8 dB, respectively. Contrary to study [7], in which all the tests were carried out on one device, each test in this trial was performed on a different device, thus validating this calibration method.

SD of the difference for the proposed method was 8.3 dB (95% CI 7.9-8.7) with the number of differences no higher than 10 dB at 89% (95% CI 88-91). These values are comparable with the results determined on a set calibrated by a normal-hearing person with verified hearing threshold (SD 6.9 and 8.29 dB [10] and SD 7.88 dB [22]) and are significantly smaller when based on a single biological calibration conducted in uncontrolled conditions (SD 10.66 dB [22]). The highest calibration accuracy is obtained in laboratory conditions, which, however, is not scalable for the Android-based devices. In laboratory conditions, SD varies from 6.4 dB to 9.9 dB [5,6,25], whereas the number of differences not higher than 10 dB changes from 81% up to 96% [4,5,6,19], depending on other settings.

SD of test-retest difference was obtained at 4.4 dB (95% CI 4.2-4.6) with the number of differences no greater than 10 dB at 99% (95% CI 99-100). The results are consistent with other works (SD 4.97 [34], number of differences 97% [6]) using the method of determining the lowest audible sound through self-adjustment of the intensity of the test signal. Moreover, these values do not differ much from the test-retest examinations obtained for automated 10 dB down and 5 dB up bracketing method (SD 6.05 dB and 5.00 dB [34], the number of differences 98% [13]). This confirms the comparable accuracy of both methods [9], at least in the cases of mobile device owners who are familiar with the use of touch screens. Additionally, the results of test-retest examinations confirm the efficiency of the method, especially in self-monitoring of hearing.

Figure 5. Mean hearing thresholds in a population of young normal-hearing adults estimated on the basis of literature data.
View this figure

Limitations

Tests were audiologist-assisted and conducted in a sound booth; therefore, the results are devoid of user mistakes such as omitting the frequency by accidental button pressing or switching the sides of headphones as well as performing the examination in noise. When tested in uncontrolled conditions, care should be taken to minimize the risk of such mistakes, for example, by monitoring the test duration at single frequency and tracking the background noise using a built-in microphone.

The factors limiting the accessibility of the test in relation to the number of Android-based mobile devices are bundled headphones and calibration coefficients. Not all the devices are offered with bundled headphones. Moreover, calibration coefficients for less popular devices may not be determined if the required number of calibrations is not achieved.

During tests on mobile devices, temporary result was continuously displayed on the screen. Therefore, during the retest, subjects could be biased by previous results, consequently reducing the test-retest difference.

Mean difference in the hearing threshold determined through pure-tone audiometry and on mobile devices differed significantly from 0 dB and reached 7.0 dB (95% CI 5.5-8.5) at 8 kHz. The reason for such discrepancies may be related to the literature-based median values of the hearing threshold of normal-hearing subjects, which were adopted to determine 0 dB HL. The mean hearing threshold estimated on the basis of literature data for people aged between 18 and 35 years displays considerable differences (Figure 5) [38,40-44]. It may be related to differing definitions of the normal-hearing person, standards of the measurement method used, or other factors such as genetic composition or occupational distribution of the population. Therefore, the determination of the values for adoption is complex. This research applies values presented in study [38]. In the future, however, it seems rational to correct them by taking into consideration the identified discrepancies (Table 5).

Conclusions

The method of hearing self-test carried out on mobile devices with bundled headphones calibrated by model-specific coefficients determined in relation to the hearing thresholds of normal-hearing persons demonstrates high compatibility with pure-tone audiometry, which confirms its potential application in hearing monitoring, screening tests, or epidemiological examinations on a large scale. However, one must acknowledge its limitations resulting mainly from the calibration method. Further evaluation of the efficiency of the method, especially in particular applications, is justified and required.

Acknowledgments

The trial was cofinanced by Wroclaw Medical University (Research Grant for Young Scientists, Grant No 170/Pbmn and the statutory funds) as well as Wrocław University of Science and Technology (the statutory funds).

Conflicts of Interest

MM is the author and owner of the Hearing Test app, by means of which the measurements described in this paper were conducted.

  1. Masalski M, Kipiński L, Grysiński T, Kręcicki T. Hearing tests on mobile devices: evaluation of the reference sound level by means of biological calibration. J Med Internet Res 2016 May 30;18(5):e130 [FREE Full text] [CrossRef] [Medline]
  2. World Health Organization. Geneva, Switzerland; 2017 Feb. Deafness and hearing loss   URL: http://www.who.int/mediacentre/factsheets/fs300/en/ [accessed 2017-12-14] [WebCite Cache]
  3. Margolis RH, Morgan DE. Automated pure-tone audiometry: an analysis of capacity, need, and benefit. Am J Audiol 2008 Dec;17(2):109-113. [CrossRef] [Medline]
  4. Foulad A, Bui P, Djalilian H. Automated audiometry using apple iOS-based application technology. Otolaryngol Head Neck Surg 2013 Nov;149(5):700-706. [CrossRef] [Medline]
  5. Sandström J, Swanepoel DW, Carel Myburgh H, Laurent C. Smartphone threshold audiometry in underserved primary health-care contexts. Int J Audiol 2016;55(4):232-238. [CrossRef] [Medline]
  6. Kam AC, Sung JK, Lee T, Wong TK, van Hasselt A. Clinical evaluation of a computerized self-administered hearing test. Int J Audiol 2012 Aug;51(8):606-610. [CrossRef] [Medline]
  7. Renda L, Selçuk Ö, Eyigör H, Osma Ü, Yılmaz MD. Smartphone based audiometric test for confirming the level of hearing; is it useable in underserved areas? J Int Adv Otol 2016 Apr;12(1):61-66 [FREE Full text] [CrossRef] [Medline]
  8. Peer S, Fagan JJ. Hearing loss in the developing world: evaluating the iPhone mobile device as a screening tool. S Afr Med J 2015 Jan;105(1):35-39. [Medline]
  9. Van Tasell DJ, Folkeard P. Reliability and accuracy of a method of adjustment for self-measurement of auditory thresholds. Otol Neurotol 2013 Jan;34(1):9-15. [CrossRef] [Medline]
  10. Honeth L, Bexelius C, Eriksson M, Sandin S, Litton J, Rosenhall U, et al. An internet-based hearing test for simple audiometry in nonclinical settings: preliminary validation and proof of principle. Otol Neurotol 2010 Jul;31(5):708-714. [CrossRef] [Medline]
  11. McPherson B, Law MM, Wong MS. Hearing screening for school children: comparison of low-cost, computer-based and conventional audiometry. Child Care Health Dev 2010 May;36(3):323-331. [CrossRef] [Medline]
  12. Margolis RH, Glasberg BR, Creeke S, Moore BCJ. AMTAS: automated method for testing auditory sensitivity: validation studies. Int J Audiol 2010 Mar;49(3):185-194. [CrossRef] [Medline]
  13. Swanepoel DW, Mngemane S, Molemong S, Mkwanazi H, Tutshini S. Hearing assessment-reliability, accuracy, and efficiency of automated audiometry. Telemed J E Health 2010 Jun;16(5):557-563. [CrossRef] [Medline]
  14. Ho ATP, Hildreth AJ, Lindsey L. Computer-assisted audiometry versus manual audiometry. Otol Neurotol 2009 Oct;30(7):876-883. [CrossRef] [Medline]
  15. Khoza-Shangase K, Kassner L. Automated screening audiometry in the digital age: exploring uhear™ and its use in a resource-stricken developing country. Int J Technol Assess Health Care 2013 Jan;29(1):42-47. [CrossRef] [Medline]
  16. Handzel O, Ben-Ari O, Damian D, Priel MM, Cohen J, Himmelfarb M. Smartphone-based hearing test as an aid in the initial evaluation of unilateral sudden sensorineural hearing loss. Audiol Neurootol 2013;18(4):201-207. [CrossRef] [Medline]
  17. Abu-Ghanem S, Handzel O, Ness L, Ben-Artzi-Blima M, Fait-Ghelbendorf K, Himmelfarb M. Smartphone-based audiometric test for screening hearing loss in the elderly. Eur Arch Otorhinolaryngol 2015 Feb 6;273(2):333-339. [CrossRef] [Medline]
  18. Szudek J, Ostevik A, Dziegielewski P, Robinson-Anagor J, Gomaa N, Hodgetts B, et al. Can Uhear me now? validation of an iPod-based hearing loss screening test. J Otolaryngol Head Neck Surg 2012 Apr;41 Suppl 1:S78-S84. [Medline]
  19. Saliba J, Al-Reefi M, Carriere JS, Verma N, Provencal C, Rappaport JM. Accuracy of mobile-based audiometry in the evaluation of hearing loss in quiet and noisy environments. Otolaryngol Head Neck Surg 2017 Apr;156(4):706-711. [CrossRef] [Medline]
  20. Yeung J, Javidnia H, Heley S, Beauregard Y, Champagne S, Bromwich M. The new age of play audiometry: prospective validation testing of an iPad-based play audiometer. J Otolaryngol Head Neck Surg 2013;42:21 [FREE Full text] [CrossRef] [Medline]
  21. Yeung JC, Heley S, Beauregard Y, Champagne S, Bromwich MA. Self-administered hearing loss screening using an interactive, tablet play audiometer with ear bud headphones. Int J Pediatr Otorhinolaryngol 2015 Aug;79(8):1248-1252. [CrossRef] [Medline]
  22. Masalski M, Kręcicki T. Self-test web-based pure-tone audiometry: validity evaluation and measurement error analysis. J Med Internet Res 2013;15(4):e71 [FREE Full text] [CrossRef] [Medline]
  23. Choi JM, Lee HB, Park CS, Oh SH, Park KS. PC-based tele-audiometry. Telemed J E Health 2007 Oct;13(5):501-508. [CrossRef] [Medline]
  24. Swanepoel DW, Myburgh HC, Howe DM, Mahomed F, Eikelboom RH. Smartphone hearing screening with integrated quality control and data management. Int J Audiol 2014 Dec;53(12):841-849. [CrossRef] [Medline]
  25. Larrosa F, Rama-Lopez J, Benitez J, Morales JM, Martinez A, Alañon MA, et al. Development and evaluation of an audiology app for iPhone/iPad mobile devices. Acta Otolaryngol 2015;135(11):1119-1127. [CrossRef] [Medline]
  26. Al-Abri R, Al-Balushi M, Kolethekkat A, Bhargava D, Al-Alwi A, Al-Bahlani H, et al. The accuracy of IOS device-based uHear as a screening tool for hearing loss: a preliminary study from the Middle East. Oman Med J 2016 Mar;31(2):142-145 [FREE Full text] [CrossRef] [Medline]
  27. Livshitz L, Ghanayim R, Kraus C, Farah R, Even-Tov E, Avraham Y, et al. Application-based hearing screening in the elderly population. Ann Otol Rhinol Laryngol 2017 Jan;126(1):36-41. [CrossRef] [Medline]
  28. Bright T, Pallawela D. Validated smartphone-based apps for ear and hearing assessments: a review. JMIR Rehabil Assist Technol 2016 Dec 23;3(2):e13. [CrossRef]
  29. Paglialonga A, Tognola G, Pinciroli F. Apps for hearing science and care. Am J Audiol 2015 Sep;24(3):293-298. [CrossRef] [Medline]
  30. Statista. Number of smartphone users worldwide from 2014 to 2020   URL: https://www.statista.com/statistics/330695/number-of-smartphone-users-worldwide/ [accessed 2017-02-15] [WebCite Cache]
  31. Swanepoel DW, Clark JL, Koekemoer D, Hall JW, Krumm M, Ferrari DV, et al. Telehealth in audiology: the need and potential to reach underserved communities. Int J Audiol 2010 Mar;49(3):195-202. [CrossRef] [Medline]
  32. Fagan JJ, Jacobs M. Survey of ENT services in Africa: need for a comprehensive intervention. Glob Health Action 2009;2 [FREE Full text] [CrossRef] [Medline]
  33. Open Signal. 2015 Aug. Android fragmentation visualized   URL: https://opensignal.com/reports/2015/08/android-fragmentation/ [accessed 2017-12-14] [WebCite Cache]
  34. Masalski M, Grysiński T, Kręcicki T. Biological calibration for web-based hearing tests: evaluation of the methods. J Med Internet Res 2014;16(1):e11 [FREE Full text] [CrossRef] [Medline]
  35. British Society of Audiology. Recommended procedure: pure-tone air-conduction and bone-conduction threshold audiometry with and without masking   URL: http://www.thebsa.org.uk/wp-content/uploads/2014/04/BSA_RP_PTA_FINAL_24Sept11_MinorAmend06Feb12.pdf [accessed 2017-12-14] [WebCite Cache]
  36. Masalski M. Google Play. Hearing Test   URL: https://play.google.com/store/apps/details?id=mobile.eaudiologia&hl=en [accessed 2017-12-14] [WebCite Cache]
  37. Amlani AM. Apps for the ears: tablet and smartphone users can choose from a growing number of apps that test hearing sensitivity and amplify sound. The ASHA Leader 2014 Jul;19:34-35 [FREE Full text] [CrossRef]
  38. Engdahl B, Tambs K, Borchgrevink HM, Hoffman HJ. Screened and unscreened hearing threshold levels for the adult population: results from the nord-trøndelag hearing loss study. Int J Audiol 2005 Apr;44(4):213-230. [Medline]
  39. Klockhoff I, Drettner B, Hagelin KW, Lindholm L. A method for computerized classification of pure tone screening audiometry results in noise-exposed groups. Acta Otolaryngol 1973 Apr;75(4):339-340. [Medline]
  40. Lutman ME, Davis AC. The distribution of hearing threshold levels in the general population aged 18-30 years. Audiology 1994;33(6):327-350. [Medline]
  41. Smith P, Davis A, Ferguson M, Lutman M. Hearing in young adults: report to ISO/TC43/WG1. Noise Health 1999;1(4):1-10 [FREE Full text] [Medline]
  42. ISO. 2000. Acoustics: statistical distribution of hearing thresholds as a function of age   URL: https://www.iso.org/standard/26314.html [accessed 2017-12-22] [WebCite Cache]
  43. Johansson MSK, Arlinger SD. Hearing threshold levels for an otologically unscreened, non-occupationally noise-exposed population in Sweden. Int J Audiol 2002 Apr;41(3):180-194. [Medline]
  44. de Beer B, Graamans K, Zielhuis G, Ad S. Hearing thresholds of otologically healthy 18-year-olds. J Hear Sci 2011;1(2):27-34. [CrossRef]


HT: hearing threshold determined by pure-tone audiometry
mHL: mobile hearing level
mHT: hearing threshold determined on mobile devices
SD: standard deviation


Edited by G Eysenbach; submitted 05.04.17; peer-reviewed by A Paglialonga, J Rey-Martinez; comments to author 12.07.17; revised version received 07.09.17; accepted 30.10.17; published 10.01.18

Copyright

©Marcin Masalski, Tomasz Grysiński, Tomasz Kręcicki. Originally published in JMIR Mhealth and Uhealth (http://mhealth.jmir.org), 10.01.2018.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mhealth and uhealth, is properly cited. The complete bibliographic information, a link to the original publication on http://mhealth.jmir.org/, as well as this copyright and license information must be included.